FreeStyle Libre 3 Plus 78768 (UDI 00357599844004)Recalls & reported problems

Abbott Diabetes Care Inc · FreeStyle Libre 3 PLUS Sensor · Class 2

Quick summary

This is an Abbott FreeStyle Libre 3 Plus sensor. Its reports mostly describe incorrect glucose readings, some far below readings from a comparison meter. There are 28,165 reports, including 1 death and 1,156 injuries. No recalls name this device. Reports rose to 26,222 in the last 12 months, from 1,943 before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
33,035
Deaths reported
1
Injuries reported
1,176
Recalls naming it
0

Summary

The FDA's MAUDE database contains 33,035 adverse event reports that list this exact device (DI 00357599844004), received between Nov 11, 2024 and Sep 30, 2026 — classified as 1 death, 1,176 injuries, 31,857 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (27,176 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 4 reportsDec 2024: 12 reportsJan 2025: 34 reportsFeb 2025: 62 reportsMar 2025: 115 reportsApr 2025: 125 reportsMay 2025: 189 reportsJun 2025: 199 reportsJul 2025: 340 reportsAug 2025: 340 reportsSep 2025: 232 reportsOct 2025: 291 reportsNov 2025: 331 reportsDec 2025: 1,565 reportsJan 2026: 1,792 reportsFeb 2026: 703 reportsMar 2026: 1,638 reportsApr 2026: 3,535 reportsMay 2026: 2,738 reportsJun 2026: 2,273 reportsJul 2026: 3,686 reportsAug 2026: 8,608 reportsSep 2026: 4,223 reportsOct 2026: 0 reports2024202520268,608
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings27,176
  • Low Readings4,736
  • High Readings235
  • Device Alarm System233
  • Adverse Event Without Identified Device or Use Problem200
  • Device Displays Incorrect Message158
  • Unable to Obtain Readings153
  • Detachment of Device or Device Component129

Patient problems

  • No Clinical Signs, Symptoms or Conditions31,838
  • Hypoglycemia287
  • Hyperglycemia273
  • Loss of consciousness202
  • Insufficient Information149
  • Dizziness126
  • Skin Infection125
  • Device Embedded In Tissue or Plaque103

Recent adverse event reports

The 25 most recent of 33,035 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Unable to Obtain Readings; Device Displays Incorrect Message

    A "replace sensor" error message was reported with the Abbott Diabetes Care (ADC) device and the customer was unable to obtain glucose readings. As a result, the customer experienced hyperglycemia and received IV treatment at urgent care from a healthcare professional for a diagnosis of hyperglycemia. There was no report of death or permanent impairment associated with this event.

    Patient problems: Hyperglycemia

    MAUDE report 50952233 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 50 mg/dL compared to readings of 200 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 9 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949358 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 55 mg/dL compared to readings of 161 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 8 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949354 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 350 mg/dL compared to readings of 130 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 7 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949352 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 47 mg/dL compared to readings of 277 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the D Zone, showing the difference in values to be clinically significant. The length of sensor wear was unknown. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949347 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 54.00 mg/dL compared to readings of 155.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947949 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947906 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947904 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 56.00 mg/dL compared to readings of 159.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 7 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947898 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 39 mg/dL compared to readings of 140 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947404 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 300.00 mg/dL compared to readings of 134.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 11 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947242 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946869 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 400.00 mg/dL compared to readings of 120.00 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 8 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946811 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 2.70 mmol/L compared to readings of 12.00 mmol/L respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 9 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946751 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 170 mg/dL, 268 mg/dL compared to readings of 71 mg/dL, 106 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 7 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946701 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 351 mg/dL compared to readings of 170 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into C Zone, showing difference in values to be clinically significant. The length of sensor wear was 11 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946207 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 111 mg/dL compared to readings of 251 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into C Zone, showing difference in values to be clinically significant. The length of sensor wear was unknown. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946074 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 300.00 mg/dL compared to readings of 140.00 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 3 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946020 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 40 mg/dL compared to readings of 163 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into C Zone, showing difference in values to be clinically significant. The length of sensor wear was 8 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946014 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 64.00 mg/dL compared to readings of 176.00 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was unknown. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946000 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 67.00 mg/dL compared to readings of 199.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945925 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 400.00 mg/dL compared to readings of 127.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 4 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945771 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 83 mg/dL, 90 mg/dL, 60 mg/dL and 80 mg/dL compared to readings of 135 mg/dL, 130 mg/dL, 189 mg/dL and 107 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into C Zone, showing difference in values to be clinically significant. The length of sensor wear was unknown. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945767 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 53 mg/dL compared to readings of 249 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 6 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945703 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 53 mg/dL compared to readings of 316 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into D Zone, showing difference in values to be clinically significant. The length of sensor wear was 2 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945702 ↗

Other Integrated Continuous Glucose Monitoring system, factory calibrated devices