Dexcom, Inc. · G7 GSS Patch Septum · Class 2

All models of Dexcom G7 (32) · Dexcom G7 GSS (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Dexcom G7 glucose sensor is part of a continuous glucose monitoring system. Reports mostly describe signal loss, readings that did not match blood glucose meter checks, and parts detaching. There are 7,118 reports, with no deaths and 45 injuries. No recalls name this device. Reports fell to 1 in the last 12 months from 175 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
7,118
Deaths reported
0
Injuries reported
45
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7,118 adverse event reports that list this exact device (DI 00386270003386), received between Jun 29, 2023 and Jan 3, 2026 — classified as 45 injuries, 7,073 malfunctions.

The most frequently reported specific device problem is “Wireless Communication Problem” (5,265 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 754 reportsDec 2023: 868 reportsJan 2024: 1,487 reportsFeb 2024: 1,069 reportsMar 2024: 645 reportsApr 2024: 373 reportsMay 2024: 217 reportsJun 2024: 193 reportsJul 2024: 117 reportsAug 2024: 126 reportsSep 2024: 137 reportsOct 2024: 86 reportsNov 2024: 63 reportsDec 2024: 17 reportsJan 2025: 33 reportsFeb 2025: 12 reportsMar 2025: 8 reportsApr 2025: 8 reportsMay 2025: 7 reportsJun 2025: 4 reportsJul 2025: 11 reportsAug 2025: 8 reportsSep 2025: 4 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261,487
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Wireless Communication Problem5,265
  • Incorrect, Inadequate or Imprecise Result or Readings605
  • Detachment of Device or Device Component570
  • Premature End-of-Life Indicator276
  • Appropriate Device Problem Term/Code Not Available192
  • Break91
  • Connection Problem79
  • Unintended Application Program Shut Down16

Patient problems

  • No Clinical Signs, Symptoms or Conditions6,574
  • Hyperglycemia397
  • Hypoglycemia91
  • Foreign Body In Patient38
  • Loss of consciousness12
  • Shaking/Tremors10
  • Diaphoresis9
  • Dizziness9

Recent adverse event reports

The 25 most recent of 7,118 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 3, 2026Protective Measures Problem

    DEXCOM BECAME AWARE THAT THE USER EXPERIENCED A FAILURE BUT DID NOT RECEIVE THE EXPECTED SENSOR FAILURE ALERT ON THEIR G7 CGM APPLICATION. DATA WAS PROVIDED AND AN INVESTIGATION WAS PERFORMED. THE ROOT CAUSE WAS DETERMINED TO BE A CODING ISSUE. THE G7 CGM APPLICATION ONLY DETECTS THAT THE SESSION HAS ENDED, THUS, IT DOES NOT ALERT THE USER THAT THE SENSOR HAS FAILED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23952876 ↗
  • MalfunctionSep 11, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23030877 ↗
  • MalfunctionSep 9, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PRODUCT HAS BEEN RECEIVED BUT IS PENDING EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED UPON COMPLETION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23012325 ↗
  • MalfunctionSep 5, 2025Premature End-of-Life Indicator; Premature End-of-Life Indicator

    IT WAS REPORTED THAT EARLY SENSOR EXPIRATION OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22978618 ↗
  • MalfunctionSep 3, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22953415 ↗
  • MalfunctionAug 29, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A PAIRING ISSUE OCCURRED. PERFORMANCE DATA WAS REVIEWED. A PAIRING ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, SIGNAL LOSS OVER AN HOUR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE SIGNAL LOSS WAS DETERMINED TO BE THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22928435 ↗
  • InjuryAug 27, 2025Low Readings; Low Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE WEARABLE WAS INSERTED INTO THE ARM ON (B)(6) 2025. ON (B)(6) 2025, THE PATIENT WAS SHOWING SYMPTOMS OF DIABETIC KETOACIDOSIS (DKA). SHE FELT LIGHTHEADED, "SMELLING OF KETONES", NAUSEA AND VOMITING. THE CGM WAS DISPLAYING 315 MG/DL WHILE THE BG WAS OVER 600 MG/DL. THE PATIENT WAS TAKEN TO THE EMERGENCY DEPARTMENT AND TREATED WITH IV FLUIDS. LABORATORY TESTS WERE ALSO DONE. RESULTS SHOWED THAT THE PATIENT WAS NOT IN DKA. THE PATIENT WAS DISCHARGED THE SAME EVENING WHEN THEY STABILIZED. NO DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND THE PROBABLE CAUSE COULD NOT BE DETERMINED. THE REPORTED GLUCOSE VALUES FALL WITHIN THE B ZONE OF THE PARKES ERROR GRID. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.

    Patient problems: Dizziness; Hyperglycemia; Nausea; Vomiting

    MAUDE report 22909680 ↗
  • MalfunctionAug 27, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A PAIRING ISSUE OCCURRED. PERFORMANCE DATA WAS REVIEWED. A PAIRING ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, SIGNAL LOSS OVER ONE HOUR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE SIGNAL LOSS COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22906156 ↗
  • MalfunctionAug 27, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A PAIRING ISSUE OCCURRED. PERFORMANCE DATA WAS REVIEWED. A PAIRING ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, SIGNAL LOSS OVER AN HOUR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE SIGNAL LOSS WAS DETERMINED TO BE THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22905651 ↗
  • MalfunctionAug 27, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A PAIRING ISSUE OCCURRED. PERFORMANCE DATA WAS REVIEWED. A PAIRING ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, SIGNAL LOSS OVER ONE HOUR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE SIGNAL LOSS COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22903023 ↗
  • MalfunctionAug 20, 2025Break; Break

    IT WAS REPORTED THAT A BROKEN SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO SERIOUS OR PROLONGED PATIENT HARM OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Foreign Body In Patient; Pain

    MAUDE report 22857151 ↗
  • MalfunctionAug 18, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22835499 ↗
  • MalfunctionAug 6, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22730215 ↗
  • MalfunctionJul 30, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22666504 ↗
  • MalfunctionJul 29, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A NO READINGS/ SENSOR ERROR/ BRIEF SENSOR ISSUE OCCURRED. PERFORMANCE DATA WAS REVIEWED. NO READINGS/ SENSOR ERROR/ BRIEF SENSOR ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, SIGNAL LOSS OVER AN HOUR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE SIGNAL LOSS WAS DETERMINED TO BE THE TRANSMITTER AND APP WERE UNABLE TO COMPLETE A DATA TRANSFER. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22646382 ↗
  • MalfunctionJul 28, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT NO READINGS/ SENSOR ERROR/ BRIEF SENSOR ISSUE OCCURRED. PERFORMANCE DATA WAS REVIEWED. NO READINGS/ SENSOR ERROR/ BRIEF SENSOR ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, SIGNAL LOSS OVER ONE HOUR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE SIGNAL LOSS COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22625535 ↗
  • MalfunctionJul 25, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. PRODUCT HAS BEEN RECEIVED BUT IS PENDING EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED UPON COMPLETION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22612484 ↗
  • MalfunctionJul 22, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT UNSPECIFIED ERROR MESSAGE OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED AS SIGNAL LOSS GREATER THAN AN HOUR WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE SIGNAL LOSS. THE PROBABLE CAUSE OF THE SIGNAL LOSS COULD NOT BE DETERMINED. THE REPORTED EVENT OF A UNSPECIFIED ERROR MESSAGE IS REPORTABLE BASED ON THE FINDING OF SIGNAL LOSS GREATER THAN AN HOUR. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22570024 ↗
  • MalfunctionJul 22, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22566893 ↗
  • MalfunctionJul 17, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22536146 ↗
  • MalfunctionJul 17, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT NO READINGS/ SENSOR ERROR/ BRIEF SENSOR ISSUE OCCURRED. PERFORMANCE DATA WAS REVIEWED. NO READINGS/ SENSOR ERROR/ BRIEF SENSOR ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, SIGNAL LOSS OVER ONE HOUR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE SIGNAL LOSS COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22535570 ↗
  • MalfunctionJul 14, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22502618 ↗
  • MalfunctionJul 14, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A PAIRING ISSUE OCCURRED. PERFORMANCE DATA WAS REVIEWED. A PAIRING ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, SIGNAL LOSS OVER AN HOUR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE SIGNAL LOSS WAS DETERMINED TO BE THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22501176 ↗
  • MalfunctionJul 10, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22472545 ↗
  • MalfunctionJun 11, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. PERFORMANCE DATA WAS NOT PROVIDED FOR EVALUATION. THE ALLEGATION COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED AS THE ALLEGATION WAS NOT FOUND. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22185726 ↗

Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices