Dexcom, Inc. · G7 US Retail Sensor/Transmitter 1-pack · Class 2

All models of Dexcom G7 (32) · Dexcom G7 GSS (155) (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Dexcom G7 glucose sensor is a continuous glucose monitor. Its 406 adverse event reports often describe wireless communication problems, and some describe readings that differ from blood glucose meter results or sensors coming loose. The reports include no deaths and no injuries. No recalls name this device. Reports fell from 182 in the earlier year to 1 in the latest year.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
406
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 406 adverse event reports that list this exact device (DI 00386270003423), received between May 9, 2024 and Nov 13, 2025 — classified as 406 malfunctions.

The most frequently reported specific device problem is “Wireless Communication Problem” (319 reports).

4 other recalls cover devices cleared under the same FDA submission (K213919); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 3 reportsJun 2024: 1 reportsJul 2024: 24 reportsAug 2024: 48 reportsSep 2024: 81 reportsOct 2024: 66 reportsNov 2024: 36 reportsDec 2024: 23 reportsJan 2025: 29 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 91 reportsJun 2025: 3 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202691
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Wireless Communication Problem319
  • Detachment of Device or Device Component36
  • Incorrect, Inadequate or Imprecise Result or Readings26
  • Overheating of Device8
  • Appropriate Device Problem Term/Code Not Available7
  • Break6
  • No Audible Alarm2
  • Failure to Sense1

Patient problems

  • No Clinical Signs, Symptoms or Conditions383
  • Hyperglycemia18
  • Hypoglycemia4
  • Electric Shock1
  • Nausea1

Recent adverse event reports

The 25 most recent of 406 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 13, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23546911 ↗
  • MalfunctionJun 26, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT A BROKEN OR DETACHED NEEDLE OR CANNULA OCCURRED. THE SENSOR WAS INSERTED INTO THE ABDOMEN ON (B)(6) 2025. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22340598 ↗
  • MalfunctionJun 13, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. DATA WAS EVALUATED AND THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION DID NOT FIND BLOOD GLUCOSE CALIBRATION VALUES TO COMPARE TO CGM VALUES DURING THE REPORTED SENSOR SESSION OR THE CALIBRATIONS DURING THE REPORTED SENSOR SESSION WERE IN ACCURATE RANGE PER DEVICE SPECIFICATIONS. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hypoglycemia

    MAUDE report 22203084 ↗
  • MalfunctionJun 6, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. NO DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. THE REPORTED GLUCOSE VALUES FALL WITHIN THE E ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22160150 ↗
  • MalfunctionMay 27, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. NO DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND THE PROBABLE CAUSE COULD NOT BE DETERMINED. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22090032 ↗
  • MalfunctionMay 21, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. NO DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND THE PROBABLE CAUSE COULD NOT BE DETERMINED. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22061070 ↗
  • MalfunctionMay 21, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. DATA WAS EVALUATED AND THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION DID NOT FIND BLOOD GLUCOSE CALIBRATION VALUES TO COMPARE TO CGM VALUES DURING THE REPORTED SENSOR SESSION OR THE CALIBRATIONS DURING THE REPORTED SENSOR SESSION WERE IN ACCURATE RANGE PER DEVICE SPECIFICATIONS. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22060765 ↗
  • MalfunctionMay 20, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. NO DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND THE PROBABLE CAUSE COULD NOT BE DETERMINED. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22057975 ↗
  • MalfunctionMay 20, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM, ON (B)(6) 2025. DATA WAS EVALUATED AND THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION DID NOT FIND BLOOD GLUCOSE CALIBRATION VALUES TO COMPARE TO CGM VALUES DURING THE REPORTED SENSOR SESSION OR THE CALIBRATIONS DURING THE REPORTED SENSOR SESSION WERE IN ACCURATE RANGE PER DEVICE SPECIFICATIONS. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22057357 ↗
  • MalfunctionMay 20, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. DATA WAS EVALUATED AND THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION DID NOT FIND BLOOD GLUCOSE CALIBRATION VALUES TO COMPARE TO CGM VALUES DURING THE REPORTED SENSOR SESSION OR THE CALIBRATIONS DURING THE REPORTED SENSOR SESSION WERE IN ACCURATE RANGE PER DEVICE SPECIFICATIONS. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22057237 ↗
  • MalfunctionMay 20, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ABDOMEN ON (B)(6) 2025. NO DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND THE PROBABLE CAUSE COULD NOT BE DETERMINED. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hypoglycemia

    MAUDE report 22053902 ↗
  • MalfunctionMay 19, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT NO READINGS/ SENSOR ERROR/ BRIEF SENSOR ISSUE OCCURRED. PERFORMANCE DATA WAS REVIEWED. NO READINGS/ SENSOR ERROR/ BRIEF SENSOR ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, SIGNAL LOSS OVER ONE HOUR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE SIGNAL LOSS COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22044114 ↗
  • MalfunctionMay 16, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. DATA WAS EVALUATED AND THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION DID NOT FIND BLOOD GLUCOSE CALIBRATION VALUES TO COMPARE TO CGM VALUES DURING THE REPORTED SENSOR SESSION OR THE CALIBRATIONS DURING THE REPORTED SENSOR SESSION WERE IN ACCURATE RANGE PER DEVICE SPECIFICATIONS. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hypoglycemia

    MAUDE report 22039572 ↗
  • MalfunctionMay 16, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22038721 ↗
  • MalfunctionMay 16, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT NO READINGS/ SENSOR ERROR/ BRIEF SENSOR ISSUE OCCURRED. PERFORMANCE DATA WAS REVIEWED. NO READINGS/ SENSOR ERROR/ BRIEF SENSOR ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, SIGNAL LOSS OVER ONE HOUR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE SIGNAL LOSS COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22038079 ↗
  • MalfunctionMay 16, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT NO READINGS/ SENSOR ERROR/ BRIEF SENSOR ISSUE OCCURRED. PERFORMANCE DATA WAS REVIEWED. NO READINGS/ SENSOR ERROR/ BRIEF SENSOR ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, SIGNAL LOSS OVER ONE HOUR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE SIGNAL LOSS COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22038073 ↗
  • MalfunctionMay 16, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. NO DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND THE PROBABLE CAUSE COULD NOT BE DETERMINED. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22036723 ↗
  • MalfunctionMay 15, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22032709 ↗
  • MalfunctionMay 15, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO COMPLETE A DATA TRANSFER. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22032029 ↗
  • MalfunctionMay 15, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22031915 ↗
  • MalfunctionMay 15, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22031907 ↗
  • MalfunctionMay 15, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22030654 ↗
  • MalfunctionMay 15, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22028462 ↗
  • MalfunctionMay 15, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. NO DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND THE PROBABLE CAUSE COULD NOT BE DETERMINED. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hypoglycemia

    MAUDE report 22023806 ↗
  • MalfunctionMay 14, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22021407 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K213919. One submission often covers many models, so check the product and lot details in each record.

  • Z-2448-2025 · Jul 24, 2025 · Open, Classified
    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: versions 2.8.0 and earlier Produc
  • Z-2447-2025 · Jul 24, 2025 · Open, Classified
    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: SW12300 Software Version: version 2.8.0 and earlier Product Des
  • Z-2446-2025 · Jul 24, 2025 · Open, Classified
    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog Number: SW12299 Software Version: version 2.8.0 and earlier Product
  • Z-1944-2025 · May 12, 2025 · Open, Classified
    Dexcom G7 Glucose Receiver, SKUs (part number): STK-AT-011(MT26403-0), STK-AT-012 (MT26403-0), STK-AT-013 (MT26403-0), STK-GT-001 (MT26403-1), STK-GT-008 (MT26403-5), STK-GT-013

Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices

All Integrated Continuous Glucose Monitoring System, Factory Calibrated devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page