Skip to content

Dexcom G7 GSS (160) STP-AT-012Recalls & reported problems

Dexcom, Inc. · G7 US Retail Sensor/Transmitter 1-pack · Class 2

All models of Dexcom G7 (32) · Dexcom G7 GSS (160) (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Dexcom G7 sensor is a continuous glucose monitor that sends readings wirelessly to a phone. Of its 317 reports, all are malfunctions, mostly describing wireless communication problems, and some describe readings that differ from blood glucose meter results. No deaths and no injuries are reported. No recalls name this device. Reports fell to 0 in the last 12 months from 230 in the previous 12.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
317
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 317 adverse event reports that list this exact device (DI 00386270003614), received between Feb 2, 2024 and Jun 24, 2025 — classified as 317 malfunctions.

The most frequently reported specific device problem is “Wireless Communication Problem” (281 reports).

4 other recalls cover devices cleared under the same FDA submission (K213919); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 12 reportsAug 2024: 8 reportsSep 2024: 15 reportsOct 2024: 49 reportsNov 2024: 88 reportsDec 2024: 51 reportsJan 2025: 49 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 39 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202688
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Wireless Communication Problem281
  • Incorrect, Inadequate or Imprecise Result or Readings16
  • Detachment of Device or Device Component13
  • Appropriate Device Problem Term/Code Not Available1
  • Break1
  • No Device Output1
  • Premature End-of-Life Indicator1
  • Unintended Ejection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions303
  • Hyperglycemia12
  • Hemorrhage/Blood Loss/Bleeding1
  • Hypoglycemia1
  • Nausea1

Recent adverse event reports

The 25 most recent of 317 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 24, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. NO DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND THE PROBABLE CAUSE COULD NOT BE DETERMINED. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22324818 ↗
  • MalfunctionJun 11, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION DID NOT FIND BLOOD GLUCOSE CALIBRATION VALUES TO COMPARE TO CGM VALUES DURING THE REPORTED SENSOR SESSION OR THE CALIBRATIONS DURING THE REPORTED SENSOR SESSION WERE IN ACCURATE RANGE PER DEVICE SPECIFICATIONS. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22184986 ↗
  • MalfunctionMay 29, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION DID NOT FIND BLOOD GLUCOSE CALIBRATION VALUES TO COMPARE TO CGM VALUES DURING THE REPORTED SENSOR SESSION OR THE CALIBRATIONS DURING THE REPORTED SENSOR SESSION WERE IN ACCURATE RANGE PER DEVICE SPECIFICATIONS. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hypoglycemia

    MAUDE report 22102713 ↗
  • MalfunctionMay 29, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. NO DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND THE PROBABLE CAUSE COULD NOT BE DETERMINED. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22102709 ↗
  • MalfunctionMay 28, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE UNABLE/UNWILLING TO ANSWER ON NI. DATA WAS EVALUATED AND THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION DID NOT FIND BLOOD GLUCOSE CALIBRATION VALUES TO COMPARE TO CGM VALUES DURING THE REPORTED SENSOR SESSION OR THE CALIBRATIONS DURING THE REPORTED SENSOR SESSION WERE IN ACCURATE RANGE PER DEVICE SPECIFICATIONS. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22097399 ↗
  • MalfunctionMay 27, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE UPPER BUTTOCKS. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE C ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION FOUND A DIFFERENCE IN VALUES THAT FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22093112 ↗
  • MalfunctionMay 25, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. NO DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND THE PROBABLE CAUSE COULD NOT BE DETERMINED. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22082690 ↗
  • MalfunctionMay 20, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. NO DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND THE PROBABLE CAUSE COULD NOT BE DETERMINED. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22057859 ↗
  • MalfunctionMay 16, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO ARM ON (B)(6) 2025. DATA WAS EVALUATED AND THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION DID NOT FIND BLOOD GLUCOSE CALIBRATION VALUES TO COMPARE TO CGM VALUES DURING THE REPORTED SENSOR SESSION OR THE CALIBRATIONS DURING THE REPORTED SENSOR SESSION WERE IN ACCURATE RANGE PER DEVICE SPECIFICATIONS. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22040588 ↗
  • MalfunctionMay 16, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A NO READINGS/ SENSOR ERROR/ BRIEF SENSOR ISSUE OCCURRED. PERFORMANCE DATA WAS REVIEWED. NO READINGS/ SENSOR ERROR/ BRIEF SENSOR ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, SIGNAL LOSS OVER AN HOUR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE SIGNAL LOSS WAS DETERMINED TO BE THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22036912 ↗
  • MalfunctionMay 15, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22031160 ↗
  • MalfunctionMay 15, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. NORDIC BLUETOOTH PAIRING TEST WAS PERFORMED AND PASSED. PERFORMANCE DATA WAS NOT PROVIDED FOR EVALUATION. THE ALLEGATION COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED AS THE ALLEGATION WAS NOT FOUND. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22027852 ↗
  • MalfunctionMay 15, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION FOUND A DIFFERENCE IN VALUES THAT FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22024964 ↗
  • MalfunctionMay 15, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22024751 ↗
  • MalfunctionMay 15, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22024713 ↗
  • MalfunctionMay 13, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ABDOMEN, WHICH IS OFF-LABEL USAGE OF THE DEVICE. DATA WAS EVALUATED AND THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION DID NOT FIND BLOOD GLUCOSE CALIBRATION VALUES TO COMPARE TO CGM VALUES DURING THE REPORTED SENSOR SESSION OR THE CALIBRATIONS DURING THE REPORTED SENSOR SESSION WERE IN ACCURATE RANGE PER DEVICE SPECIFICATIONS. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22015762 ↗
  • MalfunctionMay 13, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. PRODUCT HAS BEEN RECEIVED BUT IS PENDING EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED UPON COMPLETION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22015415 ↗
  • MalfunctionMay 13, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22015023 ↗
  • MalfunctionMay 13, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT A BROKEN NEEDLE OR CANNULA OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22014227 ↗
  • MalfunctionMay 13, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22014218 ↗
  • MalfunctionMay 13, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED ON (B)(6) 2025. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22013906 ↗
  • MalfunctionMay 13, 2025Premature End-of-Life Indicator; Premature End-of-Life Indicator

    IT WAS REPORTED THAT EARLY SENSOR EXPIRATION OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22013663 ↗
  • MalfunctionMay 13, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. DATA WAS EVALUATED AND THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION DID NOT FIND BLOOD GLUCOSE CALIBRATION VALUES TO COMPARE TO CGM VALUES DURING THE REPORTED SENSOR SESSION OR THE CALIBRATIONS DURING THE REPORTED SENSOR SESSION WERE IN ACCURATE RANGE PER DEVICE SPECIFICATIONS. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia; Nausea

    MAUDE report 22012207 ↗
  • MalfunctionMay 13, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. IT WAS INDICATED THAT THE PATIENT RUBBED/BUMPED/LAID ON THE TRANSMITTER, WHICH IS MISUSE OF THE DEVICE. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION FOUND A DIFFERENCE IN VALUES THAT FALLS WITHIN THE C ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22012189 ↗
  • MalfunctionMay 13, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2025. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION FOUND A DIFFERENCE IN VALUES THAT FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 22011915 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K213919. One submission often covers many models, so check the product and lot details in each record.

  • Z-2446-2025 · Jul 24, 2025 · Open, Classified
    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog Number: SW12299 Software Version: version 2.8.0 and earlier Product
  • Z-2447-2025 · Jul 24, 2025 · Open, Classified
    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: SW12300 Software Version: version 2.8.0 and earlier Product Des
  • Z-2448-2025 · Jul 24, 2025 · Open, Classified
    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: versions 2.8.0 and earlier Produc
  • Z-1944-2025 · May 12, 2025 · Open, Classified
    Dexcom G7 Glucose Receiver, SKUs (part number): STK-AT-011(MT26403-0), STK-AT-012 (MT26403-0), STK-AT-013 (MT26403-0), STK-GT-001 (MT26403-1), STK-GT-008 (MT26403-5), STK-GT-013

Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices

All Integrated Continuous Glucose Monitoring System, Factory Calibrated devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page