Skip to content

Dexcom G7 GSS (1) STP-AT-018Recalls & reported problems

Dexcom, Inc. · G7 GSS US (1pk) -161 (Government Payor) · Class 2

Quick summary

This Dexcom G7 sensor is a continuous glucose monitor. The 937 reports mostly describe signal loss between the sensor and phone app, sensors detaching, and low or inaccurate readings. Some mention high or low blood sugar, and 23 mention a foreign object in the patient. The reports include no deaths and 28 injuries, and no recalls name this device. Reports fell to 248 over the last year from 689 in the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
939
Deaths reported
0
Injuries reported
28
Recalls naming it
0

Summary

The FDA's MAUDE database contains 939 adverse event reports that list this exact device (DI 00386270004550), received between Feb 14, 2025 and Sep 23, 2026 — classified as 28 injuries, 911 malfunctions.

The most frequently reported specific device problem is “Wireless Communication Problem” (352 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 155 reportsJun 2025: 148 reportsJul 2025: 148 reportsAug 2025: 99 reportsSep 2025: 63 reportsOct 2025: 72 reportsNov 2025: 49 reportsDec 2025: 53 reportsJan 2026: 30 reportsFeb 2026: 28 reportsMar 2026: 34 reportsApr 2026: 48 reportsMay 2026: 2 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 3 reportsOct 2026: 0 reports202420252026155
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Wireless Communication Problem352
  • Detachment of Device or Device Component343
  • Low Readings78
  • Incorrect, Inadequate or Imprecise Result or Readings49
  • Break43
  • Premature End-of-Life Indicator33
  • High Readings27
  • Adverse Event Without Identified Device or Use Problem4

Patient problems

  • No Clinical Signs, Symptoms or Conditions769
  • Hyperglycemia102
  • Hypoglycemia33
  • Foreign Body In Patient23
  • Loss of consciousness5
  • Elevated Ketones/ Diabetic Ketoacidosis3
  • Erythema3
  • Skin Infection3

Recent adverse event reports

The 25 most recent of 939 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50873491 ↗
  • MalfunctionSep 8, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50677658 ↗
  • MalfunctionSep 2, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50570702 ↗
  • MalfunctionAug 28, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50507396 ↗
  • MalfunctionJun 18, 2026Wireless Communication Problem

    IT WAS REPORTED THAT UNSPECIFIED ISSUE OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6)2026. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED AS SIGNAL LOSS GREATER THAN AN HOUR WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE SIGNAL LOSS DUE TO THE TRANSMITTER AND APP WERE NOT BEING ABLE TO ESTABLISH A CONNECTION. THE REPORTED EVENT OF A UNSPECIFIED ISSUE IS REPORTABLE BASED ON THE FINDING OF SIGNAL LOSS GREATER THAN AN HOUR. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25555825 ↗
  • MalfunctionJun 16, 2026Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25520726 ↗
  • MalfunctionMay 26, 2026Wireless Communication Problem

    IT WAS REPORTED THAT NO READINGS OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2026. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED AS SIGNAL LOSS GREATER THAN AN HOUR WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE SIGNAL LOSS DUE TO THE TRANSMITTER AND APP WERE NOT BEING ABLE TO ESTABLISH A CONNECTION. THE REPORTED EVENT OF A NO READINGS IS REPORTABLE BASED ON THE FINDING OF SIGNAL LOSS GREATER THAN AN HOUR. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25278648 ↗
  • MalfunctionMay 5, 2026Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25089184 ↗
  • MalfunctionApr 30, 2026Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO COMPLETE A DATA TRANSFER. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25052407 ↗
  • MalfunctionApr 30, 2026Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25038904 ↗
  • MalfunctionApr 28, 2026Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25021085 ↗
  • MalfunctionApr 28, 2026Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25017981 ↗
  • MalfunctionApr 27, 2026High Readings

    SUBSEQUENT TO THE INITIAL MDR, A CORRECTION IS REQUIRED. IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM, ON (B)(6) 2026. IT WAS INDICATED THAT THE PATIENT RUBBED/BUMPED/LAID ON THE TRANSMITTER, WHICH IS MISUSE OF THE DEVICE. DATA WAS EVALUATED AND THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION DID NOT FIND BLOOD GLUCOSE CALIBRATION VALUES TO COMPARE TO CGM VALUES DURING THE REPORTED SENSOR SESSION OR THE CALIBRATIONS DURING THE REPORTED SENSOR SESSION WERE IN ACCURATE RANGE PER DEVICE SPECIFICATIONS. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hypoglycemia

    MAUDE report 25010624 ↗
  • MalfunctionApr 27, 2026Low Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2026. DATA WAS EVALUATED AND THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION DID NOT FIND BLOOD GLUCOSE CALIBRATION VALUES TO COMPARE TO CGM VALUES DURING THE REPORTED SENSOR SESSION OR THE CALIBRATIONS DURING THE REPORTED SENSOR SESSION WERE IN ACCURATE RANGE PER DEVICE SPECIFICATIONS. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 25007856 ↗
  • MalfunctionApr 27, 2026Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. IT WAS INDICATED THAT THE PATIENT WAS USING AN UNSUPPORTED OPERATING SYSTEM, WHICH IS MISUSE OF THE DEVICE. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25000579 ↗
  • MalfunctionApr 23, 2026Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24975049 ↗
  • MalfunctionApr 22, 2026Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED FREQUENTLY BETWEEN (B)(6) 2026 AND (B)(6) 2026. DATA WAS PROVIDED FOR INVESTIGATION. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER ONE HOUR RARELY WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24949456 ↗
  • MalfunctionApr 21, 2026Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PRODUCT WAS PROVIDED FOR INVESTIGATION. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. NORDIC BLUETOOTH PAIRING TEST WAS PERFORMED AND FAILED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24942214 ↗
  • MalfunctionApr 21, 2026Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24940322 ↗
  • MalfunctionApr 21, 2026Low Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2026. DATA WAS EVALUATED AND THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION DID NOT FIND BLOOD GLUCOSE CALIBRATION VALUES TO COMPARE TO CGM VALUES DURING THE REPORTED SENSOR SESSION OR THE CALIBRATIONS DURING THE REPORTED SENSOR SESSION WERE IN ACCURATE RANGE PER DEVICE SPECIFICATIONS. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 24938803 ↗
  • MalfunctionApr 21, 2026Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24937215 ↗
  • MalfunctionApr 20, 2026High Readings

    IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED FREQUENTLY BETWEEN (B)(6) 2026. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2026. DATA WAS PROVIDED FOR INVESTIGATION. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED OCCASIONALLY WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE OF THE INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER COULD NOT BE DETERMINED. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION FOUND A DIFFERENCE IN VALUES THAT FALLS WITHIN THE B ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Hypoglycemia

    MAUDE report 24925569 ↗
  • MalfunctionApr 17, 2026Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24918505 ↗
  • MalfunctionApr 17, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT A DETACHED OR MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2026. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 24917789 ↗
  • MalfunctionApr 17, 2026Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24916058 ↗

Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices

All Integrated Continuous Glucose Monitoring System, Factory Calibrated devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.