Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System STP-FT-013Recalls & reported problems

Dexcom, Inc. · G7 15 Day US Medicare Sensor 1pk · Class 2

All models of Dexcom G7 (32) · Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System (6)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Dexcom G7 sensor is a continuous glucose monitor worn on the body to track blood sugar. The 1,507 reports most often describe wireless communication problems, low readings, the sensor detaching, and high readings. Some reports mention a foreign body in the patient. The reports list no deaths and 30 injuries. Reports in the last 12 months number 1,507, compared with 0 in the 12 months before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,723
Deaths reported
0
Injuries reported
36
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,723 adverse event reports that list this exact device (DI 00386270004741), received between Dec 29, 2025 and Sep 30, 2026 — classified as 36 injuries, 1,686 malfunctions.

The most frequently reported specific device problem is “Wireless Communication Problem” (1,210 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 7 reportsFeb 2026: 33 reportsMar 2026: 74 reportsApr 2026: 142 reportsMay 2026: 196 reportsJun 2026: 313 reportsJul 2026: 294 reportsAug 2026: 160 reportsSep 2026: 503 reportsOct 2026: 0 reportsNov 2026: 0 reports202420252026503
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Wireless Communication Problem1,210
  • Low Readings223
  • Detachment of Device or Device Component123
  • High Readings60
  • Break30
  • Premature End-of-Life Indicator21
  • Incorrect, Inadequate or Imprecise Result or Readings19
  • Inappropriate or Unexpected Reset8

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,398
  • Hyperglycemia223
  • Hypoglycemia61
  • Foreign Body In Patient27
  • Foreign Body Reaction10
  • Nausea10
  • Loss of consciousness9
  • Malaise9

Recent adverse event reports

The 25 most recent of 1,723 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954737 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954618 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954290 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952928 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952891 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952280 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that signal loss occurred. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951959 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that signal loss occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951431 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that a signal loss occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951104 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that signal loss occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950718 ↗
  • MalfunctionSep 30, 2026Low Readings

    It was reported that an inaccuracy between the Continuous Glucose Monitor (CGM) and the Blood Glucose (BG) meter occurred. The sensor was inserted into the arm on (b)(6)2026. Data was evaluated and the allegation was confirmed. The probable cause could not be determined via data. The reported glucose values fall within the D zone of the Parkes Error Grid. The data investigation found a difference in values that falls within the D zone of the Parkes Error Grid. No injury or medical intervention was reported.

    Patient problems: Hyperglycemia

    MAUDE report 50950366 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that a signal loss occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949453 ↗
  • MalfunctionSep 30, 2026Low Readings

    It was reported that an inaccuracy between the Continuous Glucose Monitor (CGM) and the Blood Glucose (BG) meter occurred. The sensor was inserted into the Arm, on, on (b)(6) 2026. No data was provided for evaluation. The allegation and a probable cause could not be determined. The reported glucose values fall within the D zone of the Parkes Error Grid. No injury or medical intervention was reported.

    Patient problems: Hyperglycemia

    MAUDE report 50949181 ↗
  • MalfunctionSep 30, 2026Low Readings

    It was reported that an inaccuracy between the Continuous Glucose Monitor (CGM) and the Blood Glucose (BG) meter occurred. The sensor was inserted into the arm on (b)(6)2026. Data was evaluated and the allegation was undetermined. The probable cause could not be determined via data. The reported glucose values fall within the D zone of the Parkes Error Grid. The data investigation did not find blood glucose calibration values to compare to CGM values during the reported sensor session or the calibrations during the reported sensor session were in accurate range per device specifications. No injury or medical intervention was reported.

    Patient problems: Hyperglycemia

    MAUDE report 50948830 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that signal loss occurred OCCASIONALLY between 9/6/2026 and 9/14/2026. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour was found RARELY within the investigation window. The probable cause was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948809 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that signal loss occurred. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948794 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that signal loss occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948788 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that a No Readings/ Sensor Error/ Brief Sensor Issue occurred. Performance data was reviewed. No Readings/ Sensor Error/ Brief Sensor Issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over an hour was found in connection to the reported allegation. The probable cause of the signal loss was determined to be the transmitter and app were unable to complete a data transfer. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948414 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    Product has been received but is pending evaluation. A follow up report will be submitted upon completion. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945802 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that warm-up takes longer than expected occurred. A review of the share logs was performed and the allegation was confirmed. The probable cause was determined to be signal loss. In addition, the probable cause of the signal loss could not be determined. The reported event of warm-up continued longer than intended duration is reportable based on the finding of signal loss greater than one hour. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944482 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that a No Readings/ Sensor Error/ Brief Sensor Issue occurred. Performance data was reviewed. No Readings/ Sensor Error/ Brief Sensor issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over an hour was found in connection to the reported allegation. The probable cause of the signal loss was determined to be the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943819 ↗
  • MalfunctionSep 30, 2026High Readings

    It was reported that an inaccuracy between the Continuous Glucose Monitor (CGM) and the Blood Glucose (BG) meter occurred. The sensor was inserted into the arm on (b)(6)-2026. Data was evaluated and the allegation was confirmed. The probable cause could not be determined via data. The reported glucose values fall within the D zone of the Parkes Error Grid. However, the data investigation found a difference in values that falls within the C zone of the Parkes Error Grid. No injury or medical intervention was reported.

    Patient problems: Hypoglycemia

    MAUDE report 50942335 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that signal loss occurred. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942331 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941419 ↗
  • MalfunctionSep 29, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941380 ↗

FDA clearance and classification

The FDA regulates product code QBJ (integrated continuous glucose monitoring system, factory calibrated) under 21 CFR 862.1355 as a Class II device. 21 CFR 862.1355 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.

Common questions about the Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System STP-FT-013

Has the Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System STP-FT-013 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00386270004741), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System STP-FT-013?

The FDA's MAUDE database lists 1,723 adverse event reports for this device, received between Dec 29, 2025 and Sep 30, 2026. In those reports, the most common specific device problems are “Wireless Communication Problem” (1,210), “Low Readings” (223) and “Detachment of Device or Device Component” (123); the most common patient problems are “Hyperglycemia” (223), “Hypoglycemia” (61) and “Foreign Body In Patient” (27). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System STP-FT-013?

None of the reports are classified as deaths, 36 as injuries and 1,686 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System STP-FT-013 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System STP-FT-013, and is it still sold?

The FDA's device database lists Dexcom, Inc. as the company responsible for it (brand Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System, model STP-FT-013). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices

All Integrated Continuous Glucose Monitoring System, Factory Calibrated devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page