Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System STP-FT-011Recalls & reported problems

Dexcom, Inc. · G7 15 Day US DME 1 Pack · Class 2

All models of Dexcom G7 (32) · Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Dexcom G7 glucose monitoring system is a sensor worn on the body. Its 477 adverse event reports mostly describe wireless connection problems and glucose readings that come out too low or too high compared with a blood glucose meter. There are no deaths and 10 injuries. Reports rose from none in the prior 12 months to 477 in the last 12. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
627
Deaths reported
0
Injuries reported
14
Recalls naming it
0

Summary

The FDA's MAUDE database contains 627 adverse event reports that list this exact device (DI 00386270004758), received between Jan 5, 2026 and Sep 30, 2026 — classified as 14 injuries, 613 malfunctions.

The most frequently reported specific device problem is “Wireless Communication Problem” (427 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 6 reportsMar 2026: 35 reportsApr 2026: 60 reportsMay 2026: 77 reportsJun 2026: 99 reportsJul 2026: 91 reportsAug 2026: 103 reportsSep 2026: 154 reportsOct 2026: 0 reports202420252026154
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Wireless Communication Problem427
  • Low Readings117
  • Detachment of Device or Device Component28
  • High Readings26
  • Break5
  • Inappropriate or Unexpected Reset5
  • Premature End-of-Life Indicator5
  • Appropriate Device Problem Term/Code Not Available4

Patient problems

  • No Clinical Signs, Symptoms or Conditions481
  • Hyperglycemia116
  • Hypoglycemia22
  • Nausea7
  • Polydipsia6
  • Shaking/Tremors6
  • Fatigue5
  • Vomiting5

Recent adverse event reports

The 25 most recent of 627 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026High Readings

    Patient problems: Hypoglycemia

    MAUDE report 50954772 ↗
  • MalfunctionSep 30, 2026High Readings

    Patient problems: Hypoglycemia

    MAUDE report 50954318 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949144 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that signal loss occurred. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948806 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that signal loss occurred. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948784 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that signal loss occurred. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947680 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that signal loss occurred. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946413 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that signal loss occurred. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943800 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that signal loss occurred. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942387 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that a signal loss occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942327 ↗
  • MalfunctionSep 30, 2026Low Readings

    It was reported that an inaccuracy between the Continuous Glucose Monitor (CGM) and the Blood Glucose (BG) meter occurred FREQUENTLY between 9/10/2026 and 9/11/2026. The sensor was inserted into the arm on (b)(6) 2026. It was indicated that the patient entered BG values outside the 40¿400 mg/dL (2.2¿22.2 mmol/L) range, which is misuse of the device. Data was provided for investigation. The allegation was confirmed due to the finding of inaccuracy between the Continuous Glucose Monitor (CGM) and the Blood Glucose (BG) meter occurred FREQUENTLY within the investigation window. The probable cause of the inaccuracy between the Continuous Glucose Monitor (CGM) and the Blood Glucose (BG) meter could not be determined. The reported glucose values fall within the C zone of the Parkes Error Grid. The data investigation found a difference in values that fall within the D zone of the Parkes Error Grid. No injury or medical intervention was reported.

    Patient problems: Hyperglycemia

    MAUDE report 50941913 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Erythema; Nodule; Pain; Skin Infection; Skin Inflammation/ Irritation; Swelling/ Edema

    MAUDE report 50941001 ↗
  • MalfunctionSep 29, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939010 ↗
  • MalfunctionSep 29, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938946 ↗
  • MalfunctionSep 29, 2026Low Readings

    It was reported that an inaccuracy between the Continuous Glucose Monitor (CGM) and the Blood Glucose (BG) meter occurred. The sensor was inserted into the arm on (b)(6) 2026. It was indicated that the patient entered BG values outside the 40¿400 mg/dL (2.2¿22.2 mmol/L) range, which is misuse of the device. No data was provided for evaluation. Confirmation of the allegation and a probable cause could not be determined. The reported glucose values fall within the D zone of the Parkes Error Grid. No injury or medical intervention was reported.

    Patient problems: Hyperglycemia

    MAUDE report 50934350 ↗
  • MalfunctionSep 29, 2026Wireless Communication Problem

    It was reported that signal loss occurred. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause was the transmitter and app were unable to complete a data transfer. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50931898 ↗
  • MalfunctionSep 28, 2026Low Readings

    It was reported that an inaccuracy between the Continuous Glucose Monitor (CGM) and the Blood Glucose (BG) meter occurred. The sensor was inserted into the arm on (b)(6) 2026. Data was evaluated and the allegation was undetermined. The probable cause could not be determined via data. The reported glucose values fall within the D zone of the Parkes Error Grid. The data investigation did not find blood glucose calibration values to compare to CGM values during the reported sensor session or the calibrations during the reported sensor session were in accurate range per device specifications. No injury or medical intervention was reported.

    Patient problems: Hyperglycemia

    MAUDE report 50923154 ↗
  • MalfunctionSep 28, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923123 ↗
  • MalfunctionSep 28, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920967 ↗
  • MalfunctionSep 28, 2026Wireless Communication Problem

    It was reported that"Signal Loss" occurred. No product or data was provided for investigation. The allegation and the probable cause cannot be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917102 ↗
  • MalfunctionSep 28, 2026Wireless Communication Problem

    It was reported that a signal loss occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916507 ↗
  • MalfunctionSep 28, 2026Low Readings

    It was reported that an inaccuracy between the Continuous Glucose Monitor (CGM) and the Blood Glucose (BG) meter occurred. The sensor was inserted into the arm on (b)(6)-2026. Data was evaluated and the allegation was confirmed. The probable cause could not be determined via data. The reported glucose values fall within the D zone of the Parkes Error Grid. The data investigation found a difference in values that fall within the D of the Parkes Error Grid. No injury or medical intervention was reported.

    Patient problems: Hyperglycemia

    MAUDE report 50914768 ↗
  • MalfunctionSep 25, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50903980 ↗
  • MalfunctionSep 25, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50903550 ↗
  • MalfunctionSep 25, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions