Renuvion®/J-Plasma® APYX-270BRecalls & reported problems
Apyx Medical Corporation · Renuvion®/J-Plasma Precise® Blade Tip, 5mm-Ø, 27cm-Length, Pistol Grip · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00607151050016), received between May 11, 2022 and May 11, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method1
- Patient Device Interaction Problem1
Patient problems
- Chronic Obstructive Pulmonary Disease (COPD)1
- Pneumothorax1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMay 11, 2022Improper or Incorrect Procedure or Method; Patient Device Interaction Problem; Improper or Incorrect Procedure or Method; Patient Device Interaction Problem
A (B)(6) FEMALE PATIENT WITH MEDICAL HISTORY OF CHRONIC PULMONARY OBSTRUCTIVE DISEASE (COPD) SECONDARY TO SMOKING, UNDERWENT A RENUVION PROCEDURE ON (B)(6) 2022, WHICH TOOK PLACE (B)(6). THE SURGERY WAS PERFORMED USING THE APYX-270B WITH GENERATOR SETTINGS AT 70% POWER AND 3 LITERS OF FLOW PER MINUTE UNDER GENERAL ANESTHESIA WITH ORAL TRACHEAL INTUBATION. ACCUMULATION OF HELIUM GAS DURING THE PROCEDURE WAS MASSAGED AND SUCTIONED AT THE END OF THE PROCEDURE WITH SOME IMPROVEMENT AND THE PATIENT WAS DISCHARGED HOME. SUBSEQUENTLY, THE PATIENT SUFFERED RESPIRATORY DISTRESS RESULTING IN EXTENSIVE SUBCUTANEOUS EMPHYSEMA (FACE, NECK, ARMS, BACK, ABDOMEN AND THIGHS), MARGINAL RIGHT PNEUMOTHORAX AND COMPLETE LEFT PNEUMOTHORAX DIAGNOSE BY A CT-SCAN. THE PATIENT WAS ADMITTED TO A HOSPITAL INTENSIVE CARE UNIT AND WAS TREATED WITH A CHEST TUBE TO THE LEFT LUNG. FURTHER REVIEW BY THE APYX MEDICAL ADVISORY BOARD DETERMINED THE FOLLOWING: IMPROPER TUNNELING WAS PERFORMED PRIOR TO USE OF THE RENUVION HANDPIECE. INSUFFICIENT NUMBER OF PORTALS AND THEIR INTERCONNECTION TO FACILITATE THE EVACUATION OF GASES DURING THE PROCEDURE. HIGH GAS FLOW CONFIGURATION (3 LITERS PER MINUTE IN ALL TREATED AREAS). F
Patient problems: Chronic Obstructive Pulmonary Disease (COPD); Pneumothorax
MAUDE report 14360199 ↗
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page