Renuvion APYX-RS3Recalls & reported problems
Apyx Medical Corporation · Renuvion® System Generator · Class 2
All models of Renuvion (4)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00607151050115), received between Jul 15, 2022 and Jun 7, 2025 — classified as 5 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information3
- Patient Device Interaction Problem1
- Self-Activation or Keying1
- Use of Device Problem1
Patient problems
- Full thickness (Third Degree) Burn3
- Burn(s)1
- Low Oxygen Saturation1
- Pain1
- Partial thickness (Second Degree) Burn1
- Pneumothorax1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJun 7, 2025Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information
THE PATIENT APPEARED TO EXPERIENCE SUBCUTANEOUS EMPHYSEMA WITH DROP IN PERIPHERAL OXYGEN SATURATION DURING A PROCEDURE IN WHICH RENUVION WAS USED AS PART OF THE OVERALL SURGICAL TREATMENT. THE PATIENT APPEARED TO EXPERIENCE SUBCUTANEOUS EMPHYSEMA WITH DROP IN PERIPHERAL OXYGEN SATURATION DURING A PROCEDURE IN WHICH RENUVION WAS USED AS PART OF THE OVERALL SURGICAL TREATMENT.
Patient problems: Low Oxygen Saturation; Pain; Pneumothorax
MAUDE report 22162119 ↗InjuryMar 3, 2025Self-Activation or Keying; Self-Activation or Keying
THE PATIENT WAS BURNED BY THE UNPROMPTED ACTIVATION OF AN UNIDENTIFIED MONOPOLAR HANDPIECE DURING A PROCEDURE IN WHICH AN APYX GENERATOR WAS USED.
Patient problems: Full thickness (Third Degree) Burn
MAUDE report 21510528 ↗InjuryOct 24, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
InjuryMar 23, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
AN APYX MEDICAL SALES REPRESENTATIVE REPORTED ON 13FEB2023 THAT THEY RECEIVED A REPORT FROM (B)(6) COSMETIC SURGERY INDICATING A PATIENT EXPERIENCED BURNS AFTER A SURGICAL PROCEDURE PERFORMED ON (B)(6) 2022 IN WHICH RENUVION WAS ALSO USED AS PART OF THE OVERALL SURGICAL PROCEDURE. APYX MEDICAL CONTACTED THE OFFICE ON 01MAR2023. THE SURGEON STATED THERE WAS NO DEVICE MALFUNCTION BUT THAT THEY BELIEVE RENUVION, USED IN CONJUNCTION WITH VASER LIPOSUCTION, MAY HAVE CAUSED OR CONTRIBUTED TO THE PATIENT'S BURNS.
Patient problems: Full thickness (Third Degree) Burn
MAUDE report 16601548 ↗InjuryJul 15, 2022Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT A 60 YEAR OLD VERY THIN FEMALE PATIENT WITH NO MEDICAL HISTORY AND NO PREVIOUS PROCEDURES TO THE TREATED AREA UNDERWENT MICRO-AIRE LIPOSUCTION FOLLOWED BY SUBDERMAL COAGULATION WITH RENUVION TO THE SUBMENTAL AREA. THE SURGEON REPORTED A NOTICEABLE BLISTER FORMATION IN THE TREATMENT AREA FOLLOWING 1 TREATMENT PASS OF THE RENUVION DEVICE WITH GENERATOR SETTINGS OF 20% POWER AND 4 LITER PER MINUTES (LPM) OF HELIUM FLOW. UPON DISCOVERY OF THE BLISTER, THE GENERATOR SETTINGS WERE CHANGED TO 60% POWER AND 1.5 LPM OF HELIUM FLOW, AND THE TREATMENT WAS CONTINUED TO BOTH THE RIGHT AND LEFT SIDE OF THE NECK. THE INITIAL GENERATOR SETTINGS OF 20% AND 4 LPM OF THE RENUVION DEVICE WOULD NOT HAVE CONTRIBUTED TO FORMATION OF A BLISTER AFTER ONE TREATMENT PASS AT APPROPRIATE DEVICE TREATMENT SPEEDS. CONTINUING TREATMENT FOLLOWING FORMATION OF A BLISTER IN THE TREATMENT AREA IS NOT RECOMMENDED. THE CAUSE OF THE EVENT IS UNKNOWN AT THIS TIME, BUT CONTINUING TREATMENT FOLLOWING FORMATION OF A BLISTER IN THE TREATMENT AREA IS NOT RECOMMENDED. THE GENERATOR WILL BE SENT BACK TO APYX MEDICAL CORPORATION FOR A FULL EVALUATION.
Patient problems: Full thickness (Third Degree) Burn; Partial thickness (Second Degree) Burn
MAUDE report 15028217 ↗
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page