Maquet Cardiovascular LLC · HemoPro 2 Extension Cable · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Maquet HemoPro 2 extension cable has 193 reports, 5 injuries and no deaths. Many reports describe the cable not delivering energy when activated, and others describe cables breaking, cracking or losing connection during use. Reports fell from 41 in the prior 12 months to 14 in the most recent 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
194
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 194 adverse event reports that list this exact device (DI 00607567700925), received between Jul 24, 2020 and Jun 25, 2026 — classified as 5 injuries, 189 malfunctions.

The most frequently reported specific device problem is “Failure to Deliver Energy” (74 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 8 reportsJan 2024: 4 reportsFeb 2024: 3 reportsMar 2024: 3 reportsApr 2024: 7 reportsMay 2024: 4 reportsJun 2024: 1 reportsJul 2024: 5 reportsAug 2024: 0 reportsSep 2024: 4 reportsOct 2024: 5 reportsNov 2024: 2 reportsDec 2024: 4 reportsJan 2025: 2 reportsFeb 2025: 1 reportsMar 2025: 14 reportsApr 2025: 4 reportsMay 2025: 3 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 5 reportsSep 2025: 1 reportsOct 2025: 2 reportsNov 2025: 6 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 2 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Deliver Energy74
  • Break55
  • Electrical /Electronic Property Problem21
  • Connection Problem20
  • Intermittent Continuity11
  • Insufficient Device Problem Information6
  • Failure to Disconnect5
  • Melted5

Patient problems

  • No Clinical Signs, Symptoms or Conditions152
  • No Consequences Or Impact To Patient28
  • Insufficient Information22
  • No Patient Involvement5

Recent adverse event reports

The 25 most recent of 194 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 25, 2026Failure to Deliver Energy; Insufficient Device Problem Information

    THE HOSPITAL REPORTED THAT THE VASOVIEW HEMOPRO 2 WAS DEFECTIVE AT THE TIME OF THE CASE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 25620162 ↗
  • MalfunctionMay 11, 2026Failure to Deliver Energy

    THE HOSPITAL REPORTED THAT THE HEMOPRO2 EXTENSION CABLE WAS NOT TRANSMITTING ENERGY. IT WAS REMOVED FROM THE CIRCUIT AND REPLACED WITH ANOTHER CABLE. NO PATIENT HARM OCCURRED, AS THE EVENT TOOK PLACE IN A CADAVER LABORATORY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25139955 ↗
  • MalfunctionMay 6, 2026Failure to Deliver Energy

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25096432 ↗
  • MalfunctionApr 3, 2026Break

    THE HOSPITAL REPORTED THAT AFTER A VEIN HARVEST, THE SURGEON TRIED TO REMOVE HEMOPRO2 EXTENSION CABLE FROM THE BUTT-END OF HEMOPRO 2 CUATERY DEVICE AND THE CORD BROKE. PER ATTACHED PHOTO THE CORD IS BROKEN. NO PATIENT HARM. NO DELAY IN CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24781338 ↗
  • MalfunctionMar 18, 2026Failure to Deliver Energy

    THE HOSPITAL REPORTED THAT THE HEMOPRO 2 EXTENSION CABLE IS NOT WORKING. THE POWER SUPPLY POWER SETTING IS UNKNOWN. NO POWER ENERGY TO DISPOSABLE. CHANGING OUT THE POWER SUPPLY WAS NOT MENTIONED/UNKNOWN. PATIENT IN OPERATING ROOM NO PATIENT EFFECTS OR DELAYS REPORTED. OPENED UP ANOTHER HP2 EXTENSION CABLE TO COMPLETE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24625476 ↗
  • MalfunctionMar 17, 2026Electrical /Electronic Property Problem

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24614592 ↗
  • MalfunctionFeb 4, 2026Failure to Deliver Energy

    THE HOSPITAL REPORTED THAT THE HEMOPRO2 EXTENSION CABLE WAS NOT WORKING. IT WAS REPLACED WITH ANOTHER DEVICE TO COMPLETE THE CASE. NO PATIENT IMPACT OR DELAYS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24257058 ↗
  • MalfunctionFeb 4, 2026Electrical /Electronic Property Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24256543 ↗
  • MalfunctionDec 16, 2025Crack; Crack

    THE HOSPITAL REPORTED THAT DURING THE SETUP OF THE CASE, THE NEW HEMOPRO2 EXTENSION CABLE WAS CRACKED AT THE CONNECTION. THIS WAS DISCOVERED WHEN THE TRAY WAS OPENED. A NEW CORD WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23815115 ↗
  • InjuryNov 27, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    RELATED TO TW (B)(4). THE HOSPITAL REPORTED THAT A NEW CORD WAS SHIPPED TO THEM LAST WEEK. THE HARVESTER NOTICED AND COMMENTED ON THE "LENGTH AND SKINNIER" THAN THE OTHER CORDS USED PRIOR. THE CONNECTION ENDS WERE THE SAME SO IT WAS ASSUMED THAT THESE WERE CORRECT AND USABLE. AFTER THE FIRST 2-3 BURNS (FAT TISSUE NOT BRANCHES) HE NOTICED THAT THE NON-WIRED JAW HAD BEEN "MELTED" AND HAD THE PIECE HAD BECOME DISLODGED INTO THE PATIENTS TUNNEL. HE WAS ABLE TO RETRIEVE THE PIECE USING THE C-RING AND FLUSHED THE TUNNEL WITH COPIUOS AMOUNT OF SALINE WITHOUT SEEING ANY OTHER PIECES IN THE TUNNEL SO HE DOES BELIEVE THE HE GOT THE PIECE(S) TOTALLY OUT OF THE LEG. HE THEN SWITCHED OUT THE VH-4030 AND THE VH-4000 AND TOOK THE VEIN OUT WITHOUT ANY ISSUES. THERE WAS A DELAY OF ABOUT 6-8 MINUTES.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23664678 ↗
  • MalfunctionNov 27, 2025Failure to Deliver Energy; Failure to Deliver Energy

    THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, HEMOPRO2 EXTENSION CABLE DID NOT WORK. NO POWER/ENERGY. UNKNOWN HOW OLD THE EXTENSION CABLE WAS OR HOW MANY STERILIZATIONS. NO ISSUES WITH POWER SUPPLY REPORTED. POWER SETTING UNKNOWN. AUDIBLE BEEP AND LIGHTS ON UNKNOWN. POWER SUPPLY WASN'T THE ISSUE REPORTED AND WAS WORKING. THE SURGEON OPENED UP A NEW VH-4030 TO COMPLETE THE CASE. PATIENT WAS IN THE ROOM. NO PATIENT EFFECTS OR DELAYS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23664030 ↗
  • MalfunctionNov 25, 2025Failure to Deliver Energy; Failure to Deliver Energy

    THE HOSPITAL REPORTED THAT DURING EVH PROCEDURE, NONE OF THE HEMOPRO2 EXTENSION CABLES AVAILABLE WERE WORKING WITH VH-4000 OR VH-3010 EXCEPT ONE. THE CUSTOMER STATED THAT ALL ¿OLD¿ CORDS NEEDED TO BE REPLACED WITH THE NEW CORDS THAT WERE COMPATIBLE WITH THE NEW VH-3010. A SITE VISIT WAS CONDUCTED DURING PROCEDURE. DURING INSPECTION, THE ACCT MGR OBSERVED THAT THE SINGLE CORD REPORTED AS FUNCTIONAL WAS NOT A GETINGE MANUFACTURED (NON-OEM) CORD AND HE HAD NEVER SEEN SUCH CORD. THIS WAS COMMUNICATED WITH BOTH CIRCULATING NURSE AND HARVESTER. FUNCTIONAL TESTING WAS PERFORMED ON THE NON-OEM CORD AND CONFIRMED IT OPERATES WITH THE VH-3010 AND THE HARVESTING TOOL OF THE VH-4000 BY BOTH VISUAL STEAMING OF THE HARVESTING TOOL JAW AND AUDIBLE BEEPING OF THE VH-3010. SUBSEQUENT TESTING OF ALL THE GETINGE OEM CORDS THAT WERE REPORTED AS NOT WORKING DEMONSTRATED NORMAL FUNCTION. ALL OEM CORDS WERE VERIFIED AS FULLY OPERATIONAL AND THAT WAS COMMUNICATED TO CUSTOMER. THE CUSTOMER IS REQUESTING REPLACEMENT OF ALL CORDS THAT THEY REPORTED AS NONFUNCTIONAL AS WELL AS ORDERING INFORMATION OF THE NON-OEM CORD. THE ACCT MGR TOLD THE CUSTOMER THAT HE HAD NO INFORMATION ON THE NON-OEM CORD. CASE WAS COMP

    Patient problems: Insufficient Information

    MAUDE report 23645772 ↗
  • MalfunctionNov 25, 2025Failure to Deliver Energy; Failure to Deliver Energy

    N/A.

    Patient problems: Insufficient Information

    MAUDE report 23645172 ↗
  • MalfunctionNov 25, 2025Failure to Deliver Energy; Failure to Deliver Energy

    N/A.

    Patient problems: Insufficient Information

    MAUDE report 23644954 ↗
  • MalfunctionNov 21, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE DEFECTIVE HEMOPRO2 EXTENSION CABLE WAS SWAPPED OUT FOR A NEW ONE AS THE CAUTERY FUNCTIONALITY WAS COMPROMISED, IT WAS RESTORED ONCE THE DEFECTIVE CABLE WAS SWAPPED. THERE WAS NO DELAY. THERE WAS NO PATIENT HARM.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23618616 ↗
  • MalfunctionOct 24, 2025Failure to Deliver Energy; Failure to Deliver Energy

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23383508 ↗
  • MalfunctionOct 16, 2025Failure to Deliver Energy; Insufficient Device Problem Information; Failure to Deliver Energy; Insufficient Device Problem Information

    THE HOSPITAL IS REQUESTING A REPLACEMENT HEMOPRO2 EXTENSION CABLE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23309410 ↗
  • MalfunctionSep 2, 2025Failure to Deliver Energy; Failure to Deliver Energy

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22939138 ↗
  • MalfunctionAug 22, 2025Failure to Deliver Energy; Failure to Deliver Energy

    THE HOSPITAL REPORTED THAT DURING THE PROCEDURE, THE HEMOPRO2 EXTENSION CABLE WOULD NOT WORK. NO POWER. OPENED UP ANOTHER VH-4030 TO COMPLETE CASE. NO PATIENT EFFECTS OR DELAYS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22876992 ↗
  • MalfunctionAug 13, 2025Failure to Deliver Energy; Failure to Deliver Energy

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22795547 ↗
  • MalfunctionAug 13, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, HEMOPRO2 EXTENSION CABLE ENERGY STOPPED BEING DELIVERED DURING THE LIGATION PHASE. POWER SOURCE SWAPPED WITHOUT RESOLUTION OF THE PROBLEM. VH-4030 WAS THEN SWAPPED WHILE USING THE ORIGINAL POWER SOURCE AND HARVESTER WAS ABLE TO COMPLETE THE PROCEDURE WITHOUT FURTHER INCIDENT. NO PATIENT INJURY REPORTED. MINIMAL DELAY DURING TROUBLESHOOT BUT NO MATERIAL DELAY TO PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22794347 ↗
  • MalfunctionAug 1, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, HEMOPRO2 EXTENSION CABLE STOPPED WORKING. CHANGED TO ANOTHER HP2 EXTENSION CABLE TO COMPLETE CASE. NO PATIENT EFFECTS OR DELAYS REPORTED. NO PROCEDURAL DELAYS REPORTED. TROUBLESHOOTING STEPS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22689279 ↗
  • MalfunctionAug 1, 2025Failure to Deliver Energy; Failure to Deliver Energy

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22689193 ↗
  • MalfunctionJul 31, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22675590 ↗
  • MalfunctionJun 24, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, HEMOPRO2 EXTENSION CABLE STOPPED WORKING. TROUBLESHOOTING WAS ATTEMPTED. NEW POWER SUPPLY WAS TRIED, THEN A NEW EXTENSION CABLE, WHICH WORKED. THERE WAS A PROCEDERAL DELAY DUE TO THE TROUBLE SHOOTING EFFORTS. THE CASE WAS COMPLETED USING A NEW DEVICE. THERE WAS NO PATIENT HARM OR EFFECTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22318279 ↗

Other Electrosurgical, Cutting & Coagulation & Accessories devices