VOYANT® 5mm Fusion Device EB211Recalls & reported problems
Applied Medical Resources Corporation · Sealer & Divider Handpiece · Class 2
All models of VOYANT 5mm Fusion Device (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00607915138431), received between Jul 18, 2023 and Aug 8, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Self-Activation or Keying” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Self-Activation or Keying2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 8, 2025Self-Activation or Keying; Self-Activation or Keying
PROCEDURE PERFORMED: BARIATRIC SURGERY. EVENT DESCRIPTION: AFTER THE LAST FUSION THE ACTIVATION BUTTON CONTINUED TO LAUNCH FUSION CYCLE AFTER USING THE DEVICE SEEING SEVERAL TIMES THE FUSION ACTIVATION BUTTON REMAINED BLOCKED. PATIENT STATUS: NO PROBLEM. INTERVENTION: DEVICE REPLACEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22746415 ↗MalfunctionJul 18, 2023Self-Activation or Keying; Self-Activation or Keying
PROCEDURE PERFORMED: BYPASS EVENT DESCRIPTION: TRANSLATION: ONCE CONNECTED, MESSAGES APPEARED REPEATEDLY. IMPOSSIBLE TO USE THE DM. AFTER PLUGGING IN THE DEVICE, MESSAGES APPEARED FROM THE FIRST USE (CLEAN DEVICE). IMPOSSIBILITY OF USE ...REPEATED MESSAGES. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL 01DEC22: TRANSLATION: THERE WERE NO CONSEQUENCES, INJURY OR OTHERWISE, FOR THE PATIENT BUT A BAD COAGULATION THAT GENERATED SEVERAL CATCHES AND INCREASED THE STRESS OF THE SURGERY. NO CONSEQUENCES TO REPORT FOR THE PATIENT. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL 6DEC23: THIS STATEMENT MEANS NO CONSEQUENCE FOR THE PATIENT AND THE SURGERY TOOK LONGER THAN EXPECTED DUE TO THIS PROBLEM WITH THE VOYANT DEVICE. THE DEVICE ACTIVATED BUT THE COAGULATION WAS BAD AND SEVERAL MESSAGES APPEARED REPEATEDLY ON THE GENERATOR SUCH AS ¿CLEAN THE DEVICE. ¿ TYPE OF INTERVENTION: DEVICE REPLACEMENT. PATIENT STATUS: THERE WERE NO CONSEQUENCES, INJURY OR OTHERWISE. THE SURGERY TOOK LONGER THAN EXPECTED DUE TO THIS PROBLEM WITH THE VOYANT DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17344422 ↗
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page