PlasmaBlade™ X PS210-030S (UDI 00643169936683)Recalls & reported problems
Medtronic Navigation, Inc. · PLASMABLADE X PS210-030S PINK · Class 2
All models of PlasmaBlade X (7)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00643169936683), received between Nov 20, 2020 and Aug 25, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Energy Output Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Energy Output Problem2
- Melted2
- Peeled/Delaminated2
- Arcing of Electrodes1
- Fire1
- Smoking1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 25, 2026Smoking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50433037 ↗MalfunctionFeb 7, 2022Energy Output Problem; Peeled/Delaminated; Energy Output Problem; Peeled/Delaminated
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A GENERATOR AND TWO HANDPIECES. IT WAS REPORTED THAT THERE WAS AN ISSUES WITH TWO HANDPIECES THAT WERE NOT WORKING PROPERLY. THE FIRST HANDPIECE WAS NOT CUTTING THE TISSUE, BUT COAG WAS WORKING. IT LOOKED LIKE COATING WAS PEELING OFF. THE SECOND HANDPIECE CUT DIED WITHIN 10 MINUTES OF USE. SURGEON STATED TISSUE WAS FIBROUS. THE CASE WAS COMPLETED WITHOUT INJURY NO IMPACT TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13472414 ↗MalfunctionFeb 7, 2022Energy Output Problem; Peeled/Delaminated; Energy Output Problem; Peeled/Delaminated
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A GENERATOR AND TWO HANDPIECES. IT WAS REPORTED THAT THERE WAS AN ISSUES WITH TWO HANDPIECES THAT WERE NOT WORKING PROPERLY. THE FIRST HANDPIECE WAS NOT CUTTING THE TISSUE, BUT COAG WAS WORKING. IT LOOKED LIKE COATING WAS PEELING OFF. THE SECOND HANDPIECE CUT DIED WITHIN 10 MINUTES OF USE. SURGEON STATED TISSUE WAS FIBROUS. THE CASE WAS COMPLETED WITHOUT INJURY NO IMPACT TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13472402 ↗MalfunctionJan 11, 2022Melted; Melted
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A HANDPIECE AND A GENERATOR. IT WAS REPORTED THAT DURING A CERVICAL SPINE PROCEDURE, THE HANDPIECE MELTED IN THE SURGEONS HAND DURING A PROCEDURE. ADDITIONAL INFORMATION ALSO STATED THAT THEY REPLACED THE HANDPIECE TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13226022 ↗MalfunctionNov 24, 2020Fire; Melted; Fire; Melted
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL (HCP) REGARDING A GENERATOR AND HANDPIECE. IT WAS REPORTED THAT THE HANDPIECE CAUGHT FIRE AND THE TIP MELTED WHILE THE HANDPIECE WAS BEING USED DURING A CASE. THERE WAS NO IMPACT ON PATIENT OUTCOME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10893112 ↗MalfunctionNov 20, 2020Arcing of Electrodes; Arcing of Electrodes
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
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10873258 ↗
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page