CapSure®SP Novus 5592-53 (UDI 00681490135733)Recalls & reported problems
Medtronic, Inc. · LEAD 5592-53 STEROID BIP SILICONE ATRL · Class 3 · Implant
All models of CapSureSP Novus (14)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic CapSure SP Novus heart lead has 138 adverse event reports, including 3 deaths and 53 injuries. Many describe the lead failing to detect heart signals or picking up false ones, along with electrical noise. No recalls name this device. Reports fell to 7 over the last year from 21 in the year before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 140
- Deaths reported
- 3
- Injuries reported
- 54
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 140 adverse event reports that list this exact device (DI 00681490135733), received between Jun 8, 2017 and Sep 10, 2026 — classified as 3 deaths, 54 injuries, 83 malfunctions.
The most frequently reported specific device problem is “Under-Sensing” (61 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Under-Sensing61
- Over-Sensing30
- Adverse Event Without Identified Device or Use Problem22
- High Capture Threshold11
- Signal Artifact/Noise11
- Fracture10
- High impedance10
- Low impedance9
Patient problems
- No Clinical Signs, Symptoms or Conditions66
- No Known Impact Or Consequence To Patient35
- Unspecified Infection17
- No Consequences Or Impact To Patient7
- Dyspnea3
- Occlusion2
- Pocket Erosion2
- Undesired Nerve Stimulation2
Recent adverse event reports
The 25 most recent of 140 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 10, 2026Over-Sensing
It was reported that the Right Atrial (RA) lead exhibited Far Field R-Wave (FFRW) oversensing. The RA lead remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50709137 ↗InjuryJul 29, 2026Failure to Capture; Under-Sensing; Decreased Sensitivity
It was reported that the Right Atrial (RA) lead exhibited under sensing during Atrial Tachycardia (AT)/Atrial Fibrillation (AF) stored episodes resulting in termination of AT/AF detected event(s) in progress. A low P wave amplitude was noted. The RA also exhibited no capture. It was reported that the Right Ventricular (RV) lead exhibited under sensing. A low R wave amplitude was noted. The RA lead was programmed off. The RV lead remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50074897 ↗MalfunctionMay 1, 2026Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD APPEARED TO BE UNDER SENSING DURING ATRIAL TACHYCARDIA (AT)/ATRIAL FIBRILLATION (AF) STORED EPISODES, RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25054985 ↗InjuryApr 24, 2026Break; Appropriate Device Problem Term/Code Not Available
IT WAS FURTHER REPORTED THAT THE RA LEAD WAS CAPPED AND REPLACED DUE TO A VISIBLE INSULATION BREACH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24986187 ↗MalfunctionApr 17, 2026Signal Artifact/Noise; Over-Sensing; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED ELECTROMAGNETIC INTERFERENCE (EMI)/NOISE, OVERSENSING AND HIGH THRESHOLDS. THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED EMI/NOISE ALSO. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24918378 ↗MalfunctionMar 25, 2026Electrical /Electronic Property Problem; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ ATRIAL FIBRILLATION (AT/AF) STORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS. POLARITY SWITCH WAS ALSO OBSERVED. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24685742 ↗MalfunctionMar 4, 2026Pacing Problem; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENTS RESULTING IN INAPPROPRIATE PACING AND RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24510865 ↗MalfunctionJan 8, 2026Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS. IT WAS FURTHER REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED A POLARITY SWITCH. BOTH LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24001820 ↗MalfunctionDec 11, 2025Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED T-WAVE OVERSENSING (TWOS) AND THE RIGHT ATRIAL (RA) LEAD EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING. THE LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH THRESHOLDS. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23779319 ↗InjuryOct 21, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE PATIENT EXPERIENCED SYNCOPE. IT WAS ALSO REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING ON STORED ELECTROGRAMS (EGM). IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD TRIGGERED AN ALERT FOR RISING IMPEDANCE AND EXHIBITED OVERSENSING ON STORED EGMS RESULTING IN INAPPROPRIATE PACING. THE RA LEAD REMAINS IN USE. THE RV LEAD WAS CAPPED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Syncope/Fainting
MAUDE report 23347981 ↗InjuryOct 15, 2025High Capture Threshold; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS. THE LEADS WERE CAPPED AND REPLACEMENT WITH A LEADLESS PACEMAKER. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23278446 ↗InjuryAug 12, 2025Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE PATIENT EXPERIENCED ELEVATED HEART RATES. IT WAS FURTHER REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING. THE LEAD WAS REPROGRAMMED AND REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Arrhythmia
MAUDE report 22782374 ↗InjuryAug 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO INFECTION, AND FOR THE DEVICE ERODING THROUGH THE PATIENT'S SKIN. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 22753143 ↗MalfunctionJul 31, 2025Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED OVERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) WHICH COULD BE MISCLASSIFYING AT/AF EVENTS. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22673401 ↗InjuryJul 14, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INFECTION. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 22502602 ↗MalfunctionJul 1, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED POTENTIAL UNDERSENSING ON ELECTROGRAMS (EGM). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22369686 ↗InjuryJun 10, 2025Fracture; Adverse Event Without Identified Device or Use Problem; Fracture; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH UNDEFINED IMPEDANCE AND OVERSENSING OF ATRIAL PACING. IT WAS ALSO NOTED THE RV LEAD WAS OVERSENSING AND EXHIBITED RISING THRESHOLDS. IT WAS NOTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED AN ATRIAL LEAD POSITION CHECK FAILURE. THE RA AND RV LEAD REMAINS IN USE. IT WAS FURTHER REPORTED THAT THE RV LEAD EXHIBITED RISING IMPEDANCE. THE RA AND RV LEADS EXHIBITED HIGH THRESHOLDS. THE RV LEAD WAS NOTED TO BE FRACTURED. THE LEADS WERE EXPLANTED AND REPLACED. IT WAS FURTHER REPORTED THAT THE RA LEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22175953 ↗MalfunctionMay 30, 2025Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED OVERSENSING ON STORED ELECTROGRAMS (EGM). THE OVERSENSING WAS SUSPECTED TO BE ELECTROMAGNETIC INTERFERENCE (EMI) DETECTED BY THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) ON THE ATRIAL CHANNEL, HOWEVER AN EMI SOURCE HAD NOT YET BEEN CONFIRMED. THE ICD AND RA LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22114792 ↗MalfunctionMay 28, 2025Electrical /Electronic Property Problem; Over-Sensing; Electrical /Electronic Property Problem; Over-Sensing
IT WAS REPORTED THAT THE PATIENT EXPERIENCED SHORTNESS OF BREATH. UPON DEVICE ASSESSMENT IT WAS NOTED THAT THE RIGHT VENTRICULAR (RV) AND RIGHT ATRIAL (RA) LEADS BOTH EXHIBITED OVER SENSING. THE RV LEAD EXHIBITED OVER SENSING OF SHORT V-V INTERVALS. AN ATRIAL LEAD POSITION CHECK FAILURE WAS ALSO REPORTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Dyspnea
MAUDE report 22093976 ↗InjuryMay 2, 2025Low impedance; Low impedance
IT WAS REPORTED THAT THE PACING LEADS EXHIBITED LOW IMPEDANCES AND THE "PATIENT HAD TOLD THEM THAT ONE OF THE LEADS WAS NOT FUNCTIONING PROPERLY AND WAS NOT CONNECTED". WITH THE EXCEPTION OF THE LOW IMPEDANCES THE LEAD TESTING WAS NORMAL. THE LEADS WERE REPROGRAMMED AND REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21948933 ↗MalfunctionMar 25, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT DURING USE OF THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D), IT WAS NOTED THAT THE DEVICE IS NOT CLASSIFYING EPISODE AS VENTRICULAR TACHYCARDIA (VT), IT IS SIMPLY DETECTED IN THE VT MONITOR ZONE WITHOUT DISCRIMINATORS APPLIED. THE CRT-D REMAINS IN USE. THE RIGHT ATRIAL (RA) LEAD EXHIBITED SIGNIFICANT ATRIAL UNDERSENSING NOTED DURING EPISODE. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21693872 ↗MalfunctionMar 13, 2025Pacing Problem; Under-Sensing; Pacing Problem; Under-Sensing
IT WAS FURTHER REPORTED THAT RA LEAD UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENTS WAS RESULTING IN TERM INATION OF AT/AF DETECTED EVENT(S) IN PROGRESS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21592900 ↗MalfunctionMar 3, 2025Pacing Problem; Under-Sensing; Pacing Problem; Under-Sensing
IT WAS REPORTED THAT THE PATIENT PRESENTED WITH SHORTNESS OF BREATH. THE RIGHT ATRIAL (RA) LEAD APPEARED TO BE UNDERSENSING RESULTING IN INAPPROPRIATE PACING. THE LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Dyspnea
MAUDE report 21505063 ↗InjuryMar 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INFECTION. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 21502184 ↗MalfunctionFeb 25, 2025Over-Sensing; Under-Sensing; Over-Sensing; Under-Sensing
IT WAS REPORTED THAT THERE WAS RIGHT ATRIAL (RA) LEAD OVERSENSING AND UNDERSENSING ON STORED FAST VENTRICULAR TACHYCARDIA (FVT). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21459526 ↗
Other Permanent defibrillator electrodes devices
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes333 deaths30,581 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M62Medtronic, Inc. · Permanent defibrillator electrodes221 deaths14,405 reports
- Capsurefix® Novus 5076-52Medtronic, Inc. · Permanent defibrillator electrodes111 deaths9,860 reports
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes80 deaths8,280 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M55Medtronic, Inc. · Permanent defibrillator electrodes116 deaths7,750 reports
- Durata™ 7122Q-58St. Jude Medical, Inc. · Permanent defibrillator electrodes151 deaths6,202 reports
- Durata™ 7122Q-65St. Jude Medical, Inc. · Permanent defibrillator electrodes86 deaths4,773 reports
- Capsurefix® Novus 5076-58Medtronic, Inc. · Permanent defibrillator electrodes47 deaths4,587 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page