Skip to content

DeRoyal 88-000002Recalls & reported problems

Deroyal Industries, Inc. · Electrosurgical Pencil · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00749756053280), received between Aug 9, 2024 and Aug 9, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Manufacturing, Packaging or Shipping Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Manufacturing, Packaging or Shipping Problem1
  • Self-Activation or Keying1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 9, 2024Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    A USER FACILITY REPORTED, "COAGULATOR BUTTON AND CUT BUTTON WORKED INCORRECTLY, THE BUTTONS DID THE OPPOSITE OF WHAT WAS LABELED."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19954378 ↗
  • MalfunctionAug 9, 2024Self-Activation or Keying; Self-Activation or Keying

    A USER FACILITY REPORTED, "OUR FACILITY USES YOUR BOVIE PENCILS, REF# 88-000002, AND WE HAD AN ISSUE DURING A TOTAL KNEE REPLACEMENT THIS MORNING. THE BOVIE'S "CUT" ACTIVATED ON ITS OWN AND STAYED ACTIVATED. THIS HAS HAPPENED IN THE PAST BUT THE PENCIL AND PAD WERE NOT SAVED FROM THE PREVIOUS INCIDENT. OUR BOVIE MACHINE WAS INSPECTED AND CLEARED FROM THE PREVIOUS INCIDENT. AFTER TODAY'S INCIDENT, THE BOVIE WAS SWITCHED WITH A DIFFERENT BRAND AND THE ISSUE WAS NOT DUPLICATED. THE DEFECTIVE BOVIE WAS RE-INSERTED AND IT ACTIVATED WITHOUT THE PENCIL BEING PUSHED. THE LOT NUMBER OF THIS BOVIE IS 58744504. WE HAVE PRESERVED THE BOVIE PENCIL AND CAUTERY TIP FOR FURTHER INSPECTION."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19951318 ↗