DeRoyal 88-000722Recalls & reported problems
Deroyal Industries, Inc. · ExtendEVAC Electrosurgical Pencil · Class 2
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00749756157919), received between Aug 14, 2019 and Aug 9, 2024 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Break” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break11
- Material Fragmentation11
- Electrical /Electronic Property Problem3
- Continuous Firing2
- Activation Problem1
- Activation, Positioning or Separation Problem1
- Device Remains Activated1
- Failure to Shut Off1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Burn(s)2
- Injury2
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 9, 2024Break; Material Fragmentation; Break; Material Fragmentation
A USER FACILITY FACILITY REPORTED, "CAUTERY TIP CLIP IS BREAKING OFF INSIDE THE PENCIL. TIP IS NO LONGER SECURE AND IS EITHER SUCKED UP INSIDE THE PENCIL OR FALLS OUT INTO THE PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19954154 ↗MalfunctionAug 9, 2024Break; Material Fragmentation; Break; Material Fragmentation
A USER FACILITY FACILITY REPORTED, "CAUTERY TIP CLIP IS BREAKING OFF INSIDE THE PENCIL. TIP IS NO LONGER SECURE AND IS EITHER SUCKED UP INSIDE THE PENCIL OR FALLS OUT INTO THE PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19953355 ↗MalfunctionAug 9, 2024Break; Material Fragmentation; Break; Material Fragmentation
A USER FACILITY REPORTED, "CAUTERY TIP CLIP IS BREAKING OFF INSIDE THE PENCIL. TIP IS NO LONGER SECURE AND IS EITHER SUCKED UP INSIDE THE PENCIL OR FALLS OUT INTO THE PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19953314 ↗MalfunctionAug 9, 2024Break; Material Fragmentation; Break; Material Fragmentation
A USER FACILITY REPORTED, "CAUTERY TIP CLIP IS BREAKING OFF INSIDE THE PENCIL. TIP IS NO LONGER SECURE AND IS EITHER SUCKED UP INSIDE THE PENCIL OR FALLS OUT INTO THE PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19953306 ↗MalfunctionAug 9, 2024Break; Material Fragmentation; Break; Material Fragmentation
A USER FACILITY REPORTED, "CAUTERY TIP CLIP IS BREAKING OFF INSIDE THE PENCIL. TIP IS NO LONGER SECURE AND IS EITHER SUCKED UP INSIDE THE PENCIL OR FALLS OUT INTO THE PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19953166 ↗MalfunctionAug 9, 2024Break; Material Fragmentation; Break; Material Fragmentation
A USER FACILITY REPORTED, "CAUTERY TIP CLIP IS BREAKING OFF INSIDE THE PENCIL. TIP IS NO LONGER SECURE AND IS EITHER SUCKED UP INSIDE THE PENCIL OR FALLS OUT INTO THE PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19953151 ↗MalfunctionAug 9, 2024Break; Material Fragmentation; Break; Material Fragmentation
A USER FACILITY FACILITY REPORTED, "CAUTERY TIP CLIP IS BREAKING OFF INSIDE THE PENCIL. TIP IS NO LONGER SECURE AND IS EITHER SUCKED UP INSIDE THE PENCIL OR FALLS OUT INTO THE PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19953138 ↗MalfunctionAug 9, 2024Break; Material Fragmentation; Break; Material Fragmentation
A DISTRIBUTING FACILITY REPORTED, "INSIDE PORTION OF THE PENCILS HAS PLACSTICK PIEC THAT RECEIVES ELECTROSURGICAL TIP AND THAT PLASTIC PIECE IS BREAKING.".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19953137 ↗MalfunctionAug 9, 2024Break; Material Fragmentation; Break; Material Fragmentation
A USER FACILITY FACILITY REPORTED, "CAUTERY TIP CLIP IS BREAKING OFF INSIDE THE PENCIL. TIP IS NO LONGER SECURE AND IS EITHER SUCKED UP INSIDE THE PENCIL OR FALLS OUT INTO THE PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19952997 ↗MalfunctionAug 9, 2024Break; Material Fragmentation; Break; Material Fragmentation
A USER FACILITY FACILITY REPORTED, "CAUTERY TIP CLIP IS BREAKING OFF INSIDE THE PENCIL. TIP IS NO LONGER SECURE AND IS EITHER SUCKED UP INSIDE THE PENCIL OR FALLS OUT INTO THE PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19952974 ↗MalfunctionAug 9, 2024Break; Material Fragmentation; Break; Material Fragmentation
A USER FACILITY REPORTED, "CAUTERY TIP CLIP IS BREAKING OFF INSIDE THE PENCIL. TIP IS NO LONGER SECURE AND IS EITHER SUCKED UP INSIDE THE PENCIL OR FALLS OUT INTO THE PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19952962 ↗MalfunctionSep 13, 2019Self-Activation or Keying; Activation, Positioning or Separation Problem; Activation Problem; Self-Activation or Keying; Activation, Positioning or Separation Problem; Activation Problem
THE ELECTROSURGICAL PENCIL WAS ACTIVATING WHILE NOT IN USE AND CAUSED A BURN APPROXIMATELY 2 CM TO THE PATIENT'S LEFT ANTERIOR THIGH.
Patient problems: Burn(s); Injury
MAUDE report 9047481 ↗MalfunctionSep 13, 2019Continuous Firing; Electrical /Electronic Property Problem; Continuous Firing; Electrical /Electronic Property Problem
THE CAUTERY PENCIL REMAINED ACTIVE WITHOUT FINGER CONTROL ACTIVATING THE PENCIL. IT WAS SET ON THE PATIENT AND CAUSED A BURN TO THE PATIENTS BREAST. THE CAUTERY PENCIL DEFECT WAS REPLICATED AFTER THE EVENT IN THE OR BY THE SURGEON.
Patient problems: Burn(s); Injury
MAUDE report 9047097 ↗MalfunctionSep 13, 2019Continuous Firing; Electrical /Electronic Property Problem; Continuous Firing; Electrical /Electronic Property Problem
THE COAGULATION BUTTON ON THE CAUTERY PENCIL REMAINED ACTIVATED EVEN THOUGH THE SURGEON WAS NOT PRESSING THE BUTTON.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9046794 ↗MalfunctionAug 14, 2019Electrical /Electronic Property Problem; Device Remains Activated; Failure to Shut Off; Electrical /Electronic Property Problem; Device Remains Activated; Failure to Shut Off
THE EXTENDEVAC PENCIL DID NOT SHUT OFF WHEN THE GENERATOR WAS TURNED OFF. THE PRODUCT WAS IN USE DURING A PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8894928 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.