PlasmaBlade™ X PS210-030S-LIGHT (UDI 00763000279189)Recalls & reported problems
Medtronic Navigation, Inc. · PLASMABLADEX PS210-030S-LIGHT 3.0SL PINK · Class 2
All models of PlasmaBlade X (7)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00763000279189), received between Nov 25, 2020 and May 25, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Failure to Shut Off” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Shut Off4
- Peeled/Delaminated2
- Energy Output Problem1
- No Device Output1
- Noise, Audible1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 25, 2026Failure to Shut Off
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA A MANUFACTURER REPRESENTATIVE REGARDING A HANDPIECE. IT WAS REPORTED THAT ONCE THE COAG BUTTON OF THE HANDPIECE WAS DEPRESSED, THE ENERGY CONTINUED TO RUN AS THE BUTTON WAS STUCK IN THE ON POSITION. A PATIENT WAS INVOLVED BUT THERE WAS NO IMPACT ON THE PATIENT OUTCOME. THERE WAS NO DELAY TO THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25268234 ↗MalfunctionFeb 7, 2024Failure to Shut Off; Failure to Shut Off
INFORMATION WAS RECEIVED FROM AN MANUFACTURER REPRESENTATIVE (REP) REGARDING A GENERATOR. IT WAS REPORTED THAT THE ENERGY WAS NOT SHUTTING OFF AFTER THE BUTTON WAS PRESSED. THE COAG BUTTON WAS STICKING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18660056 ↗MalfunctionOct 25, 2022No Device Output; Peeled/Delaminated; No Device Output; Peeled/Delaminated
INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING A HANDPIECE. IT WAS REPORTED THAT THE COATING ON THE END OF THE HANDPIECE PEELED OFF. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15667431 ↗MalfunctionFeb 21, 2022Peeled/Delaminated; Peeled/Delaminated
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A HANDPIECE. IT WAS REPORTED THAT TIP OF THE HANDPIECE WAS PEELING OFF ABRUPTLY DURING A CASE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13568319 ↗MalfunctionJan 12, 2021Energy Output Problem; Failure to Shut Off; Energy Output Problem; Failure to Shut Off
INFORMATION WAS RECEIVED FROM A MANUFACTURE REPRESENTATIVE (REP) REGARDING A GENERATOR AND HANDPIECES. IT WAS REPORTED THAT TA HANDPIECE MALFUNCTIONED DURING A CATH LAB ON THE DAY OF THE CALL. ADDITIONAL INFORMATION WAS RECEIVED FROM THE REP AND IT WAS FURTHER STATED THAT THE HAND PIECE CONTINUED TO PRODUCE POWER EVEN AFTER SURGEON STOPPED PRESSING BUTTON. A NEW HAND PIECE WAS OPENED AND THE ISSUE RESOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11156189 ↗MalfunctionNov 25, 2020Failure to Shut Off; Noise, Audible; Failure to Shut Off; Noise, Audible
INFORMATION WAS RECEIVED FROM A MANUFACTURING REPRESENTATIVE (REP) REGARDING A GENERATOR. IT WAS REPORTED THAT THE SITE HAD A COUPLE HANDPIECES THAT HAD AN ISSUE WHERE THE TONE AND BLADE CONTINUED AFTER THE SURGEON HAD STOPPED PRESSING THE BUTTON. IT WAS CLARIFIED THAT THE HANDPIECES WERE LIGHTED AND CONTINUED TO PRODUCE POWER AND NOISE AFTER THE SURGEON HAD REMOVED THEIR FINGER FROM THE BUTTON. THERE GENERATOR WAS STILL IN USE AND THERE WAS NO IMPACT TO PATIENT OUTCOME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10906972 ↗
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page