Aquamantys™ 23-301-1Recalls & reported problems

Medtronic Navigation, Inc. · BIPOLAR SEALER 23-301-1 AQM MBS · Class 2

All models of Aquamantys (12)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000280918), received between Apr 4, 2024 and Aug 26, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Failure to Prime” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Prime1
  • Obstruction of Flow1
  • Use of Device Problem1

Patient problems

  • Burn(s)1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2025Failure to Prime; Obstruction of Flow; Failure to Prime; Obstruction of Flow

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER REGARDING A GENERATOR. IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE HANDPIECE FAILED TO PRIME AND THERE WAS NO SALINE FLOW IN TUBING. INITIALLY TUBING WAS PLACED INTO THE PUMP SEGMENT AND THE PRIME BUTTON WAS PRESSED. THE PUMP SEGMENT ROLLERS WERE AUDIBLY TURNING INSIDE BUT THERE WAS NO FLUID MOVEMENT. THE PRIME FEATURE WAS STOPPED AND THE TUBING WAS REMOVED FROM THE PUMP. TROUBLESHOOTING METHODS WERE PERFORMED WITH TUBING REMOVED AND PRIME BUTTON PRESSED BRIEFLY WITH THE PUMP SEGMENT OPEN TO VISUALLY INSPECT MOVEMENT OF THE ROLLERS (NO ISSUE NOTED), THE IV BAG WAS SQUEEZED BUT NO FLUID WAS PASSING THROUGH THE TUBING, A NEW 1000ML BAG OF SALINE WAS SPIKED AND THE DRIP CHAMBER WAS SQUEEZED BUT DID NOT MOVE FLUID DOWN THE TUBING, A FINAL ATTEMPT WAS MADE BY PLACING THE TUBING BACK INTO THE PUMP SEGMENT BUT THE TUBING WOULD NOT PRIME OR DELIVER FLUID TO THE TIP OF THE DEVICE. FAILED TO RESOLE THE ISSUE AFTER MULTIPLE ATTEMPTS. A NEW HANDPIECE WAS DELIVERED TO THE STERILE FIELD AND THEN CONNECTED TO THE AEX GENERATOR, IT WORKED WITHOUT ISSUE. NO REPORTED IMPACT ON PATIENT. ADDITIONAL INFORMATION WAS RECEIVED. IT WAS REPORTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22892147 ↗
  • InjuryApr 4, 2024Use of Device Problem; Use of Device Problem

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A HANDPIECE. IT WAS REPORTED THAT DURING A PROCEDURE THE PATIENT HAD SUPERFICIAL BURNS FROM HOT SALINE. THE SURGEON NOTED THAT THE BURN ON THE LEFT IMF FROM THE HANDPIECE HAD HEALED IN 2/3 BUT NOT IN 1/3 AND WOULD REQUIRE A SURGICAL SCAR REVISION AT A LATER DATE. IT WAS ALSO NOTED THAT THE ISSUE WAS NOT WITH THE PRODUCT BUT WAS USER ERROR.

    Patient problems: Burn(s)

    MAUDE report 19044765 ↗

Other Electrosurgical, cutting & coagulation & accessories devices

All Electrosurgical, cutting & coagulation & accessories devices →