ClosureFast™ CF7-3-60 (UDI 00763000702298)Recalls & reported problems

Medtronic, Inc. · CATHETER CF7-3-60 CLOSURE FAST V05 · Class 2

All models of ClosureFast (8)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000702298), received between Jun 19, 2025 and Oct 22, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Overheating of Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device2
  • Deformation Due to Compressive Stress1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 22, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING TREATMENT OF VENOUS INSUFFICIENCY IN THE ANTERIOR LATERAL ACCESSORY VEIN USING THE CLOSUREFAST CATHETER, THE VEIN DID NOT CLOSE AFTER THE INITIAL TREATMENT. THE CATHETER LUMEN WAS FLUSHED PRIOR TO USE. THE CATHETER OVERHEATED TO 150 DEGREES CELSIUS, AND THE GENERATOR DISPLAYED AN ERROR MESSAGE INDICATING THAT THE TREATMENT WAS STOPPED DUE TO EXCESSIVELY HIGH TEMPERATURE. THE ERROR OCCURRED WITHIN THE FIRST CYCLE OF TREATMENT. COMPRESSION WAS UTILIZED. THE PHYSICIAN STARTED A CYCLE OUTSIDE THE BODY AGAIN JUST TO CHECK IF THE ERROR OCCURS AGAIN AND IT HAPPENED. TUMESCENT INFILTRATION AND LOCAL ANESTHESIA WERE UTILIZED, AND HAND COMPRESSION WAS APPLIED. THERE WAS NO PARAMETERS CHANGED ON THE GENERATOR. ANOTHER CATHETER WAS USED WITHIN THE SAME PROCEDURE TO COMPLETE THE TREATMENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23355275 ↗
  • InjuryJun 19, 2025Overheating of Device; Deformation Due to Compressive Stress; Overheating of Device; Deformation Due to Compressive Stress

    PHYSICIAN WAS ATTEMPTING TO USE A CLOSUREFAST CF7-3-60 TO TREAT A PATIENT. THE PREPARATION OF THE CATHETER, USE OF THE DEVICE, AND POST-USAGE PROCEDURES WAS PERFORMED IN ACCORDANCE WITH THE PACKAGE INSERT AND THE USER MANUAL. THE CATHETER LUMEN WAS FLUSHED PRIOR TO USE. THE GUIDEWIRE WAS USED DURING CATHETER INSERTION AND PLACEMENT. THE TEMPERATURE AT THE TIME OF HEAT GENERATION INCREASED TO THE SET TEMPERATURE. IT WAS REPORTED, FOR ONE LESION, TWO PUNCTURES WERE MADE TO TREAT IT, AND AFTER CAUTERIZING THE AREA NEAR THE GROIN, THE CATHETER WAS REMOVED FROM THE BODY, WHEREUPON IT WAS DISCOVERED THAT THE METAL COIL HAD BECOME KINKED. AS THE KINKED PART HAD ALSO BECOME SCORCHED, IT WAS DECIDED TO REFRAIN FROM FURTHER USE AND TO REPLACE THE CATHETER. IT IS UNCLEAR WHETHER THE CATHETER WAS ALREADY KINKED AT THE TIME IT WAS REMOVED. ANOTHER CATHETER WAS OPENED AND USED TO COMPLETE THE PROCEDURE. NO ADDITIONAL TREATMENT WAS REQUIRED. NO INJURY TO THE PATIENT AS A RESULT OF THE ISSUE/SCORCHED CATHETER. COMPRESSION WAS BEING APPLIED WHEN THE ISSUE OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22289224 ↗

Other Electrosurgical, cutting & coagulation & accessories devices

All Electrosurgical, cutting & coagulation & accessories devices →