Aquamantys™ 23-112-1 (UDI 00763000738501)Recalls & reported problems

Medtronic Navigation, Inc. · AQM 6.0 BIPOLAR SEALER UNIT · Class 2

All models of Aquamantys (12)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000738501), received between Mar 17, 2026 and Aug 25, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Noise, Audible1
  • Smoking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 25, 2026Smoking; Noise, Audible

    Information was received from healthcare provider (HCP) via a manufacturer representative regarding a handpiece. It was reported that during an above-the-knee amputation procedure, a generator produced a loud noise during activation (other case). The generator was switched out intraoperatively, and a similar loud noise was observed with the replacement generator during activation while using the same device. A loud pump noise was also reported to be always present when the handpiece was activated. The generator was then reported to have started ¿smoking¿ while the button was pressed, and use of the device was discontinued intraoperatively after the smoking was observed. The machines were changed during the case and the device was re-primed, and the procedure was completed without further use of the device. Patient was present and unknown patient impact. Additional Information was received. It was reported that the generator would not pump fluid. The cause of the noise was the bad tubing of the handpiece. The handpiece was not stopped immediately after the noise was heard; use continued until it started smoking. The noise continued during use and activation, not just during initial

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50433970 ↗
  • MalfunctionMar 17, 2026Break

    ADDITIONAL INFORMATION WAS RECEIVED. IT WAS REPORTED THAT THERE WAS A SURGICAL OF 3-4 MINUTES TO GET REPLACEMENT AND NO REPORTED IMPACT ON PATIENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24615165 ↗

Other Electrosurgical, cutting & coagulation & accessories devices

All Electrosurgical, cutting & coagulation & accessories devices →