Skip to content

OsteoCool™ 2.0 RF Ablation System OC04-100Recalls & reported problems

Medtronic Sofamor Danek, Inc. · HUB OC04-100 OC 2.0 100W · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000764043), received between Jul 29, 2026 and Jul 29, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Display or Visual Feedback Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Display or Visual Feedback Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 29, 2026Display or Visual Feedback Problem

    Information was received from the Healthcare professional via a manufacturer representative regarding a device used for balloon kyph oplasty. It was reported that the when the OC probes were connected to the connector hub, the connector hub did not recognize the probes as connected. Never got an error message. Multiple probes were attempted; however, the generator screen continued to display "connect probes." As a result, the ablation could not be initiated. The physician elected to proceed with the BKP and was able to complete the case. No harm to the patient. The generator was unplugged and restarted twice. The probes were exchanged, and each of the four probe connections was tested and didn't worked, the issue was the connector hub and not the generator itself. There was no patient involvement reported. There were no further complications reported regarding the event. Additional information was received as event occurred during pre-op.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50064299 ↗

Other Electrosurgical, cutting & coagulation & accessories devices

All Electrosurgical, cutting & coagulation & accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page