Electrosurgical Tip Cleaner; PL 138029Recalls & reported problems

Xodus Medical, Inc. · Class 2

All models of Electrosurgical Tip Cleaner; PL (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00787551000535), received between May 24, 2021 and Aug 30, 2021 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Delivered as Unsterile Product” (2 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Sep 19, 2019Class IITerminated

    The sterile barrier of some devices may have been affected

    Z-0149-2020 · Root cause: Packaging · Recall details · FDA record ↗

Most reported problems

Device problems

  • Delivered as Unsterile Product2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 30, 2021Delivered as Unsterile Product; Delivered as Unsterile Product

    THIS IS A VOLUNTARY DISTRIBUTOR REPORT DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, CATALOG # 138029 LOT 19MAR14A, FOR A POSSIBLE BREACH OF STERILITY DUE TO AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH ANY PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION IN (B)(6). THE COMPLETION OF THE PRELIMINARY DEVICE EVALUATION CONFIRMED AN INSUFFICIENT HEATSEAL; THEREFORE, THIS COMPLAINT MEETS THE CRITERIA FOR A REPORTABLE EVENT. THIS REPORT IS BEING RAISED AS A VOLUNTARY DISTRIBUTOR REPORT ON THE BASIS OF DEVICE MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12387379 ↗
  • MalfunctionMay 24, 2021Delivered as Unsterile Product; Delivered as Unsterile Product

    THIS IS A VOLUNTARY DISTRIBUTOR REPORT. DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, ELECTROSURGICAL TIP CLEANER, CATALOG 138029, LOT 19FEB27 FOR A POSSIBLE INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION IN JAPAN. DUE TO THE POTENTIAL SEVERITY OF A BREACH IN STERILITY, THIS COMPLAINT MEETS THE CRITERIA FOR A REPORTABLE EVENT. THIS REPORT IS BEING RAISED AS A VOLUNTARY DISTRIBUTOR REPORT ON THE BASIS OF DEVICE MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11873050 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →