Hydro-Surg 5600200Recalls & reported problems
Davol Inc. · Davol Advantage Reusable Quick-Disconnect Needle Tip Electrosurgical Attachment, 5mmx33cm · Class 2
All models of Hydro-Surg (8)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741077227), received between May 22, 2019 and May 22, 2019 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component2
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 22, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED PER THE USER FACILITY THAT THE BLACK COATING ON THE REUSABLE NEEDLE POINT CAUTERY TIPS IS FALLING OFF AFTER JUST ONE STERILIZATION. CONTACT CONFIRMED THAT A PIECE DETACHED WHEN CLEANING THE TIP, AND THERE IS CONCERN THIS COULD RESULT IN THE PIECE COMING OFF IN USE. THERE WAS NO REPORTED PATIENT INJURY/COMPLICATION AS A RESULT OF THE PROBLEM EXPERIENCED. ADDENDUM: IN FOLLOW UP WITH THE USER FACILITY CONTACTS THEY REPORTED THE DEVICE WAS USED IN LAP TOTAL HYSTERECTOMY PROCEDURE. THE PROBLEM PRESENTED DURING FIRST USE OF THE DEVICE, AND FOLLOWING THE CLEANING STERILIZATION PROTOCOL OF THE NON-STERILE REUSABLE DEVICE. THE ELECTRO CAUTERY TIP WAS BEING CLEARED OF DEBRIS BY THE OR TECH USING A WET SPONGE. THE INSULATION MATERIAL AFFIXED TO THE DISTAL END OF THE DEVICE APPEARED TO HAVE PHYSICALLY CRACKED AND SPLIT AND THE MATERIAL CAME OFF IN THE SPONGE IN TWO SMALL PIECES. THE PIECES WERE DISCARDED. STAFF CONFORMED THAT THERE WAS NOTHING THAT FELL INTO THE BODY, AND NO PATIENT IMPACT AS A RESULT OF THE PROBLEM THE WAS EXPERIENCED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8633473 ↗MalfunctionMay 22, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED PER THE USER FACILITY THAT THE BLACK COATING ON THE REUSABLE NEEDLE POINT CAUTERY TIPS IS FALLING OFF AFTER JUST ONE STERILIZATION. CONTACT CONFIRMED THAT A PIECE DETACHED WHEN CLEANING THE TIP, AND THERE IS CONCERN THIS COULD RESULT IN THE PIECE COMING OFF DURING USE. THERE WAS NO REPORTED PATIENT INJURY/COMPLICATION AS A RESULT OF THE PROBLEM EXPERIENCED. ADDENDUM: IN FOLLOW UP WITH THE USER FACILITY CONTACTS THEY REPORTED THE DEVICE WAS USED IN LAP TOTAL HYSTERECTOMY PROCEDURE. THE PROBLEM PRESENTED DURING FIRST USE OF THE DEVICE, AND FOLLOWING THE CLEANING STERILIZATION PROTOCOL OF THE NON-STERILE REUSABLE DEVICE. THE ELECTRO CAUTERY TIP WAS BEING CLEARED OF DEBRIS BY THE OR TECH USING A WET SPONGE. THE INSULATION MATERIAL AFFIXED TO THE DISTAL END OF THE DEVICE APPEARED TO HAVE PHYSICALLY CRACKED AND SPLIT AND THE MATERIAL CAME OFF IN THE SPONGE IN TWO SMALL PIECES. THE PIECES WERE DISCARDED. STAFF CONFORMED THAT THERE WAS NOTHING THAT FELL INTO THE BODY, AND NO PATIENT IMPACT AS A RESULT OF THE PROBLEM THE WAS EXPERIENCED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8633409 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page