Hydro-Surg 5600240Recalls & reported problems
Davol Inc. · Davol Advantage Reusable Quick-Disconnect J-Hook Electrosurgical Attachment, 5mm x 33cm · Class 2
All models of Hydro-Surg (8)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741077234), received between Nov 12, 2021 and Apr 13, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Degraded” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Degraded1
- Material Puncture/Hole1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 13, 2022Material Puncture/Hole; Material Puncture/Hole
AS REPORTED, THE BARD/DAVOL J-HOOK ELECTROSURGICAL REUSABLE TIP (WHICH REQUIRES REPROCESSING CLEANING/STERILIZATION BETWEEN CASES) WAS USED AND STERILIZED 3 TIMES PRIOR TO NOTING THE RUBBER SHEATH AROUND THE TIP WAS SEPARATED. AS REPORTED, THERE WAS NO ISSUE WITH THE DEVICE DURING USAGE AND NO REPORTED PATIENT INJURY. AS REPORTED, THE USER FACILITY HAS NOT MADE CHANGES TO THE STERILIZATION PROCESS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14106357 ↗MalfunctionNov 12, 2021Degraded; Degraded
AS REPORTED, THE BARD/DAVOL J-HOOK ELECTROSURGICAL TIP (RE-USABLE) FAILED AFTER 4 USES. AS REPORTED, THE METAL TIP IS USUALLY 2MM AWAY AND THERE WAS DAMAGE/TEARS/WEAR TO THE INSULATION COATING. IT WAS REPORTED THAT THE DEVICE FAILED THE INSULATION TEST WHEN TESTED WITH AN INSULATION TESTER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12806400 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page