ReNew 3142Recalls & reported problems

Microline Surgical, Inc. · Endocut Scissor Tip, Disposable · Class 2

All models of ReNew (45)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00811099010005), received between Jul 17, 2018 and Nov 6, 2025 — classified as 2 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Arcing” (2 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 1, 2017Terminated

    Certain lots of disposable laparoscopic instrument tips may potentially have a compromised sterile barrier.

    Z-1741-2018 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Arcing2
  • Fracture2
  • Insufficient Device Problem Information2
  • Defective Component1
  • Device Fell1
  • Electrical /Electronic Property Problem1
  • Failure to Cut1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Burn(s)2
  • Insufficient Information1
  • Internal Organ Perforation1
  • No Known Impact Or Consequence To Patient1
  • Superficial (First Degree) Burn1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 6, 2025Fracture; Fracture

    COMPONENT BROKE DURING USE.

    Patient problems: Insufficient Information

    MAUDE report 23487803 ↗
  • MalfunctionAug 1, 2025Use of Device Problem; Device Fell; Use of Device Problem; Device Fell

    A WHITE SILICONE BAND WAS INADVERTENTLY LEFT ON THE DEVICE TIP AND SUBSEQUENTLY DISLODGED INTO THE PATIENT'S CHEST CAVITY DURING THE INITIAL PROCEDURE ON (B)(6) 2025. THE SURGEON CLOSED THE CHEST WITHOUT NOTICING THE BAND. DURING A SUBSEQUENT REOPERATION FOR AN UNRELATED REASON ON (B)(6) 2025, THE SILICONE BAND WAS DISCOVERED AND REMOVED. THE SILICON BAND WAS REMOVED SAFELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22690324 ↗
  • MalfunctionMar 20, 2025Failure to Cut; Failure to Cut

    DURING A LAPAROSCOPIC OVARIAN CYSTECTOMY, THE DEVICE FAILED TO OPEN AND CLOSE DURING A CRITICAL PART OF THE PROCEDURE. THE TIP WAS REMOVED AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21653178 ↗
  • MalfunctionDec 11, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    PATIENT WAS TAKEN BACK FOR A DIAGNOSTIC LAPAROSCOPIC PROCEDURE FOR ENDOMETRIOSIS LESIONS. TWO LESIONS WERE REMOVED WITH THE MONOPOLAR SCISSORS AND LAPAROSCOPIC GRASPER. THE LESION WAS REMOVED AND UPON MOVING FURTHER AWAY FROM THE SITE WITH THE CAMERA, IT WAS IDENTIFIED THAT THE MONOPOLAR SCISSOR HAD BURNED THE FUNDAL SEROSA OF THE UTERUS IN 2 LOCATIONS AS WELL AS THE LEFT FALLOPIAN TUBE SUPERFICIALLY. THIS WAS DONE AT THE CONNECTION OF THE MONOPOLAR SCISSORS TO THE LAPAROSCOPIC INSTRUMENT AND A FAULT IN THE INSULATION OF THIS CONNECTION. THE SCISSOR TIPS AND THE LAPAROSCOPIC INSTRUMENT WERE IMMEDIATELY SWITCHED OUT FOR A NEW SET AND WERE WORKING APPROPRIATELY.

    Patient problems: Burn(s)

    MAUDE report 20907063 ↗
  • MalfunctionJun 27, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    THE NURSE REALIZED THAT THE EDGE OF THE BACK HUB WAS DENTED. IT SEEMS THAT THIS PART WAS INITIALLY DENTED SINCE THE HANDPIECE WASINTACT AND THE EDGE OF THE DENT PART WAS SMOOTH. WE CANNOT DENY THE POSSIBILITY OF THE IMPACT ON INSULATION PROPERTIES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19624683 ↗
  • InjuryApr 11, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ACCORDING TO THE REPORTER, DURING A LAP REMOVAL OF ADNEXAL/OVARY, A NON-MEDTRONIC DEVICE WAS BEING USED WITH THE GENERATOR AND CORD. AT THE POINT WHERE THE MICROLINE DISPOSABLE ATTACHED TO THE INSTRUMENT MELTED WHICH CAUSED A BURN AND PERFORATION OF THE UTERUS. WHEN INSTRUMENT WAS REMOVED EVERYTHING APPEARED TO BE CONNECTED PROPERLY. THE PERFORATION WAS SUTURED AND ADDITIONAL HOSPITAL STAY WAS NOT REQUIRED. THIS REPORT REFLECTS INFORMATION RECEIVED BY FDA IN THE FORM OF A NOTIFICATION PER 803.22 (B)(2).

    Patient problems: Internal Organ Perforation; Superficial (First Degree) Burn

    MAUDE report 19087564 ↗
  • MalfunctionJul 8, 2022Defective Component; Defective Component

    NAME PLATE COLLAR SPLIT IN HALF INTRAOPERATIVELY UNABLE TO FIND ONE HALF OF THE PIECE. THE DISLODGED (PLASTIC PORTION) PORTION OF THE TIP WAS NEVER RECOVERED AFTER THE PROCEDURE. SALES REP INDICATES THAT THE TIP SPLIT LONGITUDINALLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14958471 ↗
  • MalfunctionSep 3, 2021Fracture; Fracture

    DURING A LAP CHOLE THE BLACK PLASTIC (HEAT SHRINK) ON A DISPOSABLE SCISSOR TIP CAME APART AND HAD TO BE RETRIEVED FROM THE PATIENT'S ABDOMEN. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12420575 ↗
  • MalfunctionJun 27, 2019Arcing; Arcing

    MALE PATIENT IN THE OR FOR A LAPAROSCOPIC RIGHT INGUINAL HERNIA REPAIR. DURING THE PROCEDURE, THE SURGEON FIRED THE HANDPIECE OF THE LAPAROSCOPIC SCISSORS PRIOR TO USING THEM ON THE PATIENT. WHEN HE DID THAT, THE HANDPIECE ARCED AND CAUSED A SPARK AND SMOKE TO COME FROM THE TIP OF THE HANDPIECE, WHICH BURNED THROUGH THE RUBBER INSULATION ON THE SIDE. THE INSTRUMENT WAS IMMEDIATELY REMOVED AND A NEW SCISSOR TIPS AND HANDLE WERE OBTAINED. THE PROCEDURE CONTINUED AS PLANNED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8741259 ↗
  • InjuryJul 17, 2018Arcing; Arcing

    DURING THE RELEASE OF THE GALLBLADDER WITH THE LAPAROSCOPIC SCISSORS, A SPARK APPEARED AT THE DISPOSABLE TIP, WHICH BURNED THE DUODENUM. THE PROCEDURE AND ANESTHESIA TIME WERE EXTENDED FOR AN UNKNOWN AMOUNT OF TIME.

    Patient problems: Burn(s)

    MAUDE report 7695753 ↗

Other Electrosurgical, Cutting & Coagulation & Accessories devices

All Electrosurgical, Cutting & Coagulation & Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page