ReNew 3152Recalls & reported problems

Microline Surgical, Inc. · Mini Endocut Scissor Tip, Disposable · Class 2

All models of ReNew (45)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00811099010012), received between Jan 7, 2020 and Jul 16, 2025 — classified as 1 injury, 10 malfunctions.

The most frequently reported specific device problem is “Fracture” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture4
  • Arcing1
  • Break1
  • Defective Component1
  • Detachment of Device or Device Component1
  • Device Handling Problem1
  • Electrical /Electronic Property Problem1
  • Loss of or Failure to Bond1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Bowel Burn1
  • No Information1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 16, 2025Sparking; Sparking

    MICROLINE WAS MADE AWARE OF WHAT APPEARED TO BE AN ELECTRICAL SAFETY FAILURE EVENT OF A MICROLINE LAPARSCOPIC HANDPIECE AND A MICROLINE SCISSOR TIP DURING USE OF MONOPOLAR CAUTERIZATION DURING A PROCEDURE. THE PROCEDURE REPORTED WAS A HYSTERECTOMY AND BILATERAL SALPINGECTOMY. PATIENT WAS NOT HARMED, AND THE STAFF HAD NOTICED WHAT APPEARED TO BE MELTING AND CHARGING OF THE DISTAL SHAFT AND SCISSOR TIP OF THE MICROLINE PRODUCT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22520807 ↗
  • MalfunctionDec 26, 2024Fracture; Fracture

    THE HEAT SHRINK FELL INTO THE ABDOMINAL CAVITY AND WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21018736 ↗
  • InjuryNov 8, 2024Electrical /Electronic Property Problem; Arcing; Electrical /Electronic Property Problem; Arcing

    DURING THE PROCEDURE, THE ASSISTING SURGEON DISCOVERED A MONOPOLAR BURN TO THE BOWEL OF THE PATIENT. THE SURGEONS THEN BACKTRACKED THROUGH THE PROCEDURE TO CHECK FOR OTHER NON-NOTICED BURNS TO THE PATIENT. SURGEONS REPORTED FINDING 3 ADDITIONAL MONO POLAR BURNS TO THE PATIENT BOWEL, FOR A TOTAL OF 4 BURNS. SURGEONS DID NOT NOTICE THE BURNS UPON IMMEDIATE USE OF MONOPOLAR ELECTRO SURGICAL CAUTERY. NO VISUALIZATION OF ELECTRICAL ARCING WAS REPORTED AT THE TIME OF INCIDENT BY THE SURGEONS OR STAFF, HOWEVER THE DISTAL PORTION OF THE RENEW XR HANDPIECE DEMONSTRATES PHYSICAL DAMAGE. SOURCE OF MONOPOLAR ENERGY WAS RECORDED TO BE A VALLEY LAB FT10 SERIAL NUMBER (B)(6). SETTINGS USED WERE CUT 35 COAG BLEND, 35 COAG. THIS REPORT IS FOR THE TIP THAT WAS ATTACHED TO THE HANDPIECE AT THE TIME OF THE INCIDENT.

    Patient problems: Bowel Burn

    MAUDE report 20644137 ↗
  • MalfunctionOct 23, 2024Fracture; Fracture

    THE HEAT SHRINK FELL INTO THE ABDOMINAL CAVITY DURING AN OPERATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20514188 ↗
  • MalfunctionMar 7, 2024Fracture; Fracture

    DURING THE SURGERY, THE HEAT SHRINK OF THE TIP CAME OFF. WE CONFIRMED THE HEAT SHRINK WAS COMPLETELY OFF AND THE METAL PARTS WERE PARTIALLY EXPOSED. WE ASSUME THE HEAT SHRINK CAME OFF DUE TO THE CONTACT OF THE COMBINATION INSTRUMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18857214 ↗
  • MalfunctionFeb 23, 2023Fracture; Fracture

    A CAPA (170) WAS OPENED TO IDENTIFY THE ROOT CAUSE AND PROVIDE A CORRECTIVE ACTION. THE PRELIMINARY RESULT OF THIS INVESTIGATION INDICATE THAT THERE WAS AN ISSUE RELATED TO TEMPERATURE AND THE DWELL TIME IN THE OVEN

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16430054 ↗
  • MalfunctionJan 10, 2023Loss of or Failure to Bond; Break; Loss of or Failure to Bond; Break

    DURING THE SURGERY, HEAT SHRINK CAME OFF AND FELL INTO THE ABDOMINAL CAVITY. IT WAS REMOVED FROM THE ABDOMINAL CAVITY AND THERE WAS NO HEALTH PROBLEM FOR THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16133936 ↗
  • MalfunctionDec 15, 2022Material Fragmentation; Material Fragmentation

    DURING THE SURGERY, HEAT SHRINK CAME OFF AND FELL INTO THE ABDOMINAL CAVITY. IT WAS REMOVED FROM THE ABDOMINAL CAVITY AND THERE WAS NO HEALTH PROBLEM FOR THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15991792 ↗
  • MalfunctionAug 24, 2022Defective Component; Defective Component

    TUBING CASING BROKE OFF IN PATIENT COMPLAINT: SCISSOR TIP WAS SCREWED ONTO HANDLE 0 ISSUES NOTED. SURGEON USED SCISSORS INSIDE PT. - PULLED OUT SCISSORS AND SAID "OH ITS GRABBING WHEN I PULLED IT OUT" I INSPECTED THE TIP - NOTED 'SOMETHING WAS DIFFERENT' BUT DIDN'T REALIZE THE TUBING CASING WAS OFF UNTIL SURGEON LOOKED IN ABDOMEN & FOUND THE PIECE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15283243 ↗
  • MalfunctionSep 28, 2021Device Handling Problem; Device Handling Problem

    DURING A PROCEDURE, OUR CUSTOMER FOUND THAT THE PART WHICH SHOULD BE ADHERED CAME OFF. THE SURGEON IMMEDIATELY PICKED IT UP THUS THERE WAS NEITHER HEALTH DAMAGE NOR EXTENSION OF THE PROCEDURE. A SPARE INSTRUMENT WAS USED TO CONTINUE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12542613 ↗
  • MalfunctionJan 7, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    A THIRD OF THE PROTECTIVE SHEATH CAME OFF THE END OF THE SCISSOR TIP.

    Patient problems: No Information

    MAUDE report 9560100 ↗

Other Electrosurgical, Cutting & Coagulation & Accessories devices

All Electrosurgical, Cutting & Coagulation & Accessories devices →