ReNew 3152Recalls & reported problems
Microline Surgical, Inc. · Mini Endocut Scissor Tip, Disposable · Class 2
All models of ReNew (45)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00811099010012), received between Jan 7, 2020 and Jul 16, 2025 — classified as 1 injury, 10 malfunctions.
The most frequently reported specific device problem is “Fracture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture4
- Arcing1
- Break1
- Defective Component1
- Detachment of Device or Device Component1
- Device Handling Problem1
- Electrical /Electronic Property Problem1
- Loss of or Failure to Bond1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Bowel Burn1
- No Information1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 16, 2025Sparking; Sparking
MICROLINE WAS MADE AWARE OF WHAT APPEARED TO BE AN ELECTRICAL SAFETY FAILURE EVENT OF A MICROLINE LAPARSCOPIC HANDPIECE AND A MICROLINE SCISSOR TIP DURING USE OF MONOPOLAR CAUTERIZATION DURING A PROCEDURE. THE PROCEDURE REPORTED WAS A HYSTERECTOMY AND BILATERAL SALPINGECTOMY. PATIENT WAS NOT HARMED, AND THE STAFF HAD NOTICED WHAT APPEARED TO BE MELTING AND CHARGING OF THE DISTAL SHAFT AND SCISSOR TIP OF THE MICROLINE PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22520807 ↗MalfunctionDec 26, 2024Fracture; Fracture
THE HEAT SHRINK FELL INTO THE ABDOMINAL CAVITY AND WAS RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21018736 ↗InjuryNov 8, 2024Electrical /Electronic Property Problem; Arcing; Electrical /Electronic Property Problem; Arcing
DURING THE PROCEDURE, THE ASSISTING SURGEON DISCOVERED A MONOPOLAR BURN TO THE BOWEL OF THE PATIENT. THE SURGEONS THEN BACKTRACKED THROUGH THE PROCEDURE TO CHECK FOR OTHER NON-NOTICED BURNS TO THE PATIENT. SURGEONS REPORTED FINDING 3 ADDITIONAL MONO POLAR BURNS TO THE PATIENT BOWEL, FOR A TOTAL OF 4 BURNS. SURGEONS DID NOT NOTICE THE BURNS UPON IMMEDIATE USE OF MONOPOLAR ELECTRO SURGICAL CAUTERY. NO VISUALIZATION OF ELECTRICAL ARCING WAS REPORTED AT THE TIME OF INCIDENT BY THE SURGEONS OR STAFF, HOWEVER THE DISTAL PORTION OF THE RENEW XR HANDPIECE DEMONSTRATES PHYSICAL DAMAGE. SOURCE OF MONOPOLAR ENERGY WAS RECORDED TO BE A VALLEY LAB FT10 SERIAL NUMBER (B)(6). SETTINGS USED WERE CUT 35 COAG BLEND, 35 COAG. THIS REPORT IS FOR THE TIP THAT WAS ATTACHED TO THE HANDPIECE AT THE TIME OF THE INCIDENT.
Patient problems: Bowel Burn
MAUDE report 20644137 ↗MalfunctionOct 23, 2024Fracture; Fracture
THE HEAT SHRINK FELL INTO THE ABDOMINAL CAVITY DURING AN OPERATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20514188 ↗MalfunctionMar 7, 2024Fracture; Fracture
DURING THE SURGERY, THE HEAT SHRINK OF THE TIP CAME OFF. WE CONFIRMED THE HEAT SHRINK WAS COMPLETELY OFF AND THE METAL PARTS WERE PARTIALLY EXPOSED. WE ASSUME THE HEAT SHRINK CAME OFF DUE TO THE CONTACT OF THE COMBINATION INSTRUMENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18857214 ↗MalfunctionFeb 23, 2023Fracture; Fracture
A CAPA (170) WAS OPENED TO IDENTIFY THE ROOT CAUSE AND PROVIDE A CORRECTIVE ACTION. THE PRELIMINARY RESULT OF THIS INVESTIGATION INDICATE THAT THERE WAS AN ISSUE RELATED TO TEMPERATURE AND THE DWELL TIME IN THE OVEN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16430054 ↗MalfunctionJan 10, 2023Loss of or Failure to Bond; Break; Loss of or Failure to Bond; Break
DURING THE SURGERY, HEAT SHRINK CAME OFF AND FELL INTO THE ABDOMINAL CAVITY. IT WAS REMOVED FROM THE ABDOMINAL CAVITY AND THERE WAS NO HEALTH PROBLEM FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16133936 ↗MalfunctionDec 15, 2022Material Fragmentation; Material Fragmentation
DURING THE SURGERY, HEAT SHRINK CAME OFF AND FELL INTO THE ABDOMINAL CAVITY. IT WAS REMOVED FROM THE ABDOMINAL CAVITY AND THERE WAS NO HEALTH PROBLEM FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15991792 ↗MalfunctionAug 24, 2022Defective Component; Defective Component
TUBING CASING BROKE OFF IN PATIENT COMPLAINT: SCISSOR TIP WAS SCREWED ONTO HANDLE 0 ISSUES NOTED. SURGEON USED SCISSORS INSIDE PT. - PULLED OUT SCISSORS AND SAID "OH ITS GRABBING WHEN I PULLED IT OUT" I INSPECTED THE TIP - NOTED 'SOMETHING WAS DIFFERENT' BUT DIDN'T REALIZE THE TUBING CASING WAS OFF UNTIL SURGEON LOOKED IN ABDOMEN & FOUND THE PIECE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15283243 ↗MalfunctionSep 28, 2021Device Handling Problem; Device Handling Problem
DURING A PROCEDURE, OUR CUSTOMER FOUND THAT THE PART WHICH SHOULD BE ADHERED CAME OFF. THE SURGEON IMMEDIATELY PICKED IT UP THUS THERE WAS NEITHER HEALTH DAMAGE NOR EXTENSION OF THE PROCEDURE. A SPARE INSTRUMENT WAS USED TO CONTINUE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12542613 ↗MalfunctionJan 7, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
A THIRD OF THE PROTECTIVE SHEATH CAME OFF THE END OF THE SCISSOR TIP.
Patient problems: No Information
MAUDE report 9560100 ↗
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page