ReNew 3122Recalls & reported problems

Microline Surgical, Inc. · Microscissor Tip, Disposable · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00811099010036), received between Feb 20, 2020 and Dec 26, 2024 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Fracture” (6 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 4 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 5, 2024Open, Classified

    A lot of ReNew Microscissor Tips exhibited a 2% rate of failure in terms of the heat shrink breaking off.

    Z-0663-2025 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Fracture6
  • Loss of or Failure to Bond1
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 26, 2024Fracture; Fracture

    IN ONE CASE, THE HEAT SHRINK FELL INTO THE ABDOMINAL CAVITY AND WAS RETRIEVED AND DISPOSED AT THE HOSPITAL. IN THE OTHER TWO CASES, THE HEAT SHRINK FELL OUTSIDE THE SURGICAL FIELD SO NO HARM TO THE PATIENT. THESE TWO PIECES WERE RETRIEVED AND AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21018735 ↗
  • MalfunctionNov 27, 2024Fracture; Fracture

    THE HEAT SHRINK GOT CAUGHT ON THE TROCAR AND CAME OFF AND RETRIEVED. IT DID NOT FALL INTO THE ABDOMINAL CAVITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20798320 ↗
  • MalfunctionOct 24, 2024Fracture; Fracture

    THE HEAT SHRINK CAME OFF DURING THE SURGERY. THE HEAT SHRINK WAS REMOVED FROM THE ABDOMINAL CAVITY, SO THERE WAS NO HEALTH PROBLEM. AS THE EXACT SAME PHENOMENON TOOK PLACE WITH THE SAME LOT, OUR CUSTOMER STOPPED USING THE LOT AND RETURNED TO US.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20522575 ↗
  • MalfunctionOct 24, 2024Fracture; Fracture

    THE HEAT SHRINK CAME OFF DURING THE SURGERY. THE HEAT SHRINK WAS REMOVED FROM THE ABDOMINAL CAVITY, SO THERE WAS NO HEALTH PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20522510 ↗
  • MalfunctionOct 15, 2024Fracture; Fracture

    THE HEATSHRINK CAME OFF DURING THE SURGERY. THE HEAT SHRINK WAS REMOVED FROM THE ABDOMINAL CAVITY, SO THERE WAS NO HEALTH PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20456762 ↗
  • MalfunctionOct 15, 2024Fracture; Fracture

    THE HEATSHRINK CAME OFF DURING THE SURGERY. THE HEAT SHRINK WAS REMOVED FROM THE ABDOMINAL CAVITY, SO THERE WAS NOT HEALTH PROBLEM.THE TIP WAS DISCARDED SINCE IT WAS USED FOR THE INFECTIOUS DISEASE PATIENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20455207 ↗
  • MalfunctionDec 15, 2022Material Fragmentation; Material Fragmentation

    DURING THE SURGERY, HEAT SHRINK CAME OFF AND FELL INTO THE ABDOMINAL CAVITY. IT WAS REMOVED FROM THE ABDOMINAL CAVITY AND THERE WAS NO HEALTH PROBLEM FOR THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15991858 ↗
  • MalfunctionFeb 20, 2020Loss of or Failure to Bond; Loss of or Failure to Bond

    DURING SURGERY, THE SURGEON FOUND THAT THE HEAT SHRINK HAD COME OFF IN THE PATIENT'S ABDOMINAL CAVITY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9732214 ↗

Other Electrosurgical, Cutting & Coagulation & Accessories devices

All Electrosurgical, Cutting & Coagulation & Accessories devices →