ReNew 3222Recalls & reported problems

Microline Surgical, Inc. · Fenestrated Grasper Tip, Disposable · Class 2

Quick summary

This Microline ReNew 3222 is a disposable fenestrated grasper tip used in surgery. The 24 reports mostly describe the tip breaking, snapping or coming apart during procedures, sometimes leaving small pieces inside the patient that were later retrieved. There are 24 reports mentioning this identifier, with no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 00811099010067), received between Jul 2, 2018 and Oct 2, 2025 — classified as 24 malfunctions.

The most frequently reported specific device problem is “Fracture” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture10
  • Detachment of Device or Device Component4
  • Insufficient Device Problem Information4
  • Break3
  • Material Fragmentation2
  • Material Integrity Problem1
  • Material Separation1

Patient problems

  • No Known Impact Or Consequence To Patient14
  • No Clinical Signs, Symptoms or Conditions9
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 2, 2025Break; Break

    DURING THE CASE, ONE SIDE OF THE GRASPER TIP SNAPPED OFF INSIDE THE PATIENT. THE SURGEON WAS ABLE TO SAFELY RETRIEVE THE BROKEN PIECE, AND THE PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23198471 ↗
  • MalfunctionSep 24, 2025Fracture; Fracture

    DAMAGE OCCURRED DURING PROCEDURE WHEN GRASPING A ROBOTIC SPONGE (CIGAR) APPROX.4MM IN THICKNESS. UNABLE TO DETERMINE IF A MICRO PIECE OF METAL IS LOST IN PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23137245 ↗
  • MalfunctionMar 24, 2025Fracture; Fracture

    DISPOSABLE TIP WAS NOTICED LOOSE IN THE ABDOMEN. TIP WAS REMOVED BY THE SURGEON ((B)(6)) AND DAMGE TO THE TIP LINER WAS NOTICED, AND REPORTED TO THE LOCAL REP. BATCH IS GOING TO BE REMOVED AND REPLACED AS REPORTED FAULT HAS BEEN SUGGESTED WITH ONE OF THE LOT TIPS - WHICH WAS NOTICEABLY LOOSE UPON INSPECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21679335 ↗
  • MalfunctionAug 12, 2024Fracture; Fracture

    (B)(6) MEDICAL CENTER HAD A 3222, FENESTRATED GRASPER BREAK WHEN IN USE DURING A PROCEDURE. PART OF THE GRASPER WINGS BROKE APART AND FELL INTO THE PATIENTS ABDOMEN. THE BROKEN PORTION HAD TO BE SEARCHED FOR AND RETRIEVED. ALL BROKEN PORTIONS WERE REPORTED AS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19964940 ↗
  • MalfunctionApr 11, 2024Fracture; Fracture

    HEAT SHRINK FELL OFF DURING PROCEDURE INTO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19088096 ↗
  • MalfunctionJan 10, 2023Break; Break

    GRASPER BROKE IN HALF IN PATIENT, X-RAY WAS DONE AND ALL PIECES WERE REMOVED FROM PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16131399 ↗
  • MalfunctionSep 9, 2022Break; Break

    HEAT SHRINK FELL OFF DURING PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15393574 ↗
  • MalfunctionMar 1, 2021Fracture; Fracture

    ONE SIDE OF THE GRASPER BROKE OFF IN PATIENT DURING LAPAROSCOPIC CASE. IT WAS A CLEAN BREAK AND ONE PIECE WAS RETRIEVED FROM THE PATIENT. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11396096 ↗
  • MalfunctionSep 11, 2020Fracture; Fracture

    WHEN GRASPING A VERY THICK-WALLED GALLBLADDER, A ""BRANCH"" OF THE GRASPING FORCEPS BROKE OFF, LANDED BEHIND THE LIVER, WAS LOCALIZED BY X-RAY, AND WAS EASILY REMOVED LAPAROSCOPICALLY. THE REST OF THE OPERATION WAS UNPROBLEMATIC.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10520242 ↗
  • MalfunctionJan 7, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING A PROCEDURE WITH A MICROLINE 3222 FENESTRATED GRASPER, DR. (B)(6) NOTICED THE HEAT SHRINK OF THE DISPOSABLE GRASPER TIP HAD COME OFF OF THE TIP AND WAS FREELY FLOATING AROUND THE ABDOMEN OF THE PATIENT. THE HEAT SHRINK WAS RETRIEVED, AND ALL POSSIBLE BOXES THAT THE TIP COULD OF CAME FROM REMOVED FROM CIRCULATION. LOT NUMBERS IMPACTED: 00147564, 00148767, 00148821.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9560133 ↗
  • MalfunctionJan 7, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING A PROCEDURE WITH A MICROLINE 3222 FENESTRATED GRASPER, DR. (B)(6) NOTICED THE HEAT SHRINK OF THE DISPOSABLE GRASPER TIP HAD COME OFF OF THE TIP AND WAS FREELY FLOATING AROUND THE ABDOMEN OF THE PATIENT. THE HEAT SHRINK WAS RETRIEVED, AND ALL POSSIBLE BOXES THAT THE TIP COULD OF CAME FROM REMOVED FROM CIRCULATION. LOT NUMBERS IMPACTED: 00147564, 00148767, 00148821.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9560067 ↗
  • MalfunctionJan 7, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING A PROCEDURE WITH A MICROLINE 3222 FENESTRATED GRASPER, DR. (B)(6) NOTICED THE HEAT SHRINK OF THE DISPOSABLE GRASPER TIP HAD COME OFF OF THE TIP AND WAS FREELY FLOATING AROUND THE ABDOMEN OF THE PATIENT. THE HEAT SHRINK WAS RETRIEVED, AND ALL POSSIBLE BOXES THAT THE TIP COULD OF CAME FROM REMOVED FROM CIRCULATION. LOT NUMBERS IMPACTED: 00147564, 00148767, 00148821.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9560048 ↗
  • MalfunctionAug 26, 2019Material Fragmentation; Material Fragmentation

    DURING A PROCEDURE IT WAS DISCOVERED THAT THE HEAT SHRINK BROKE AND FELL INTO THE PATIENT'S ABDOMEN. THE PIECES WERE RETRIEVED. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8931830 ↗
  • MalfunctionAug 8, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    THE JAW BROKE ON THE FENESTRATED GRASPER DURING A PROCEDURE AND FELL INTO THE PATIENT. THE PIECE WAS RETRIEVED. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8875205 ↗
  • MalfunctionApr 16, 2019Fracture; Fracture

    JAW BROKE. THERE WAS NO HARM TO THE PATIENT AND THE ANESTHESIA TIME WAS NOT EXTENDED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8520157 ↗
  • MalfunctionApr 16, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    THE HEAT SHRINK CAME OFF THE TIP AND FELL INTO THE PATIENT'S ABDOMEN DURING A PROCEDURE. THE PIECE WAS ABLE TO BE RETRIEVED. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8520154 ↗
  • MalfunctionMar 29, 2019Fracture; Fracture

    SURGEON WAS COMPLETING A LIVER RESECTION WHEN ONE SIDE OF THE GRASPER BROKE OFF AND FELL INTO THE ABDOMEN. THE PIECE WAS ABLE TO BE RETRIEVED. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8463686 ↗
  • MalfunctionMar 29, 2019Fracture; Fracture

    THE GRASPER TIP WAS CLOSED ON TISSUE WHEN THE SURGEON FELT A POP AND THE JAW BROKE JUST ABOVE THE YOLK. THE PIECE WAS ABLE TO BE RETRIEVED. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8463685 ↗
  • MalfunctionMar 29, 2019Fracture; Fracture

    GRASPER TIP BROKE OFF INSIDE PATIENT. ENTIRE JAW OF ONE SIDE BROKE OFF. PROCEDURE WAS EXTENDED BY 5 MINUTES. THE PIECE WAS ABLE TO BE RETRIEVED FROM THE PATIENT. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8463681 ↗
  • MalfunctionFeb 22, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    THE HEAT SHRINK PORTION OF THE DISPOSABLE TIP CAME OFF IN THE ABDOMINAL CAVITY. THE PIECE WAS ABLE TO BE RETRIEVED. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8363443 ↗
  • MalfunctionFeb 22, 2019Material Integrity Problem; Material Integrity Problem

    ONE OF THE JAWS ON THE GRASPERS BROKE OFF DURING A PROCEDURE AND NEEDED TO BE RETRIEVED FROM INSIDE OF THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8363442 ↗
  • MalfunctionFeb 8, 2019Material Separation; Insufficient Device Problem Information; Material Separation; Insufficient Device Problem Information

    BLACK COLLAR COMPONENT OF THE DISPOSABLE TIP HAS DETACHED FROM THE TIP AND FALLEN INSIDE THE PATIENT. THE COMPONENT WAS FOUND RECOVERED AND COLLECTED

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8322178 ↗
  • MalfunctionAug 22, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    DURING A SURGICAL PROCEDURE, THE HEAT SHRINK CAME OFF FROM THE RENEW FENESTRATED GRASPER TIP, BUT DID NOT FALL INTO THE PATIENT. THE MALFUNCTIONED TIP WAS NOT USED IN THE PATIENT AND THE SURGEON USED A NEW TIP. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7806356 ↗
  • MalfunctionJul 2, 2018Material Fragmentation; Material Fragmentation

    AT THE END OF A HYSTERECTOMY PROCEDURE, THE SURGEON NOTICED THAT A FRAGMENT OF THE HEAT SHRINK INSULATION TUBE FELL INTO THE PATIENT'S PERITONEAL CAVE. THE FRAGMENT WAS ABLE TO BE RETRIEVED. THE PROCEDURE AND ANESTHESIA TIME WERE EXTENDED FOR AN UNKNOWN AMOUNT OF TIME. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7654785 ↗

Other Electrosurgical, Cutting & Coagulation & Accessories devices

All Electrosurgical, Cutting & Coagulation & Accessories devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.