Skip to content

ReNew 3292Recalls & reported problems

Microline Surgical, Inc. · Modified Raptor Grasper Tip, Disposable · Class 2

All models of ReNew (45)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00811099010104), received between Jun 21, 2023 and Mar 25, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 25, 2025Fracture; Fracture

    DURING PROCEDURE FOR PATIENT, WHILE THE RESIDENT WAS USING LAPAROSCOPIC INSTRUMENT WITH MICROLINE GRASPER TIP INSIDE THE PATIENT'S BODY, ONE OF THE PIECE OF GRASPER TIP CAME OFF. RESIDENT WAS ABLE TO RETRIEVE IT. PROVIDED SCRUB NURSE WITH ANOTHER MICROLINE GRASPER TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21690235 ↗
  • MalfunctionJun 3, 2024Fracture; Fracture

    DURING PROCEDURE FOR PATIENT, WHILE USING LAPAROSCOPIC INSTRUMENT WITH MICROLINE GRASPER TIP INSIDE THE PATIENT'S BODY, ONE OF THE PIECE OF GRASPER TIP CAME OFF. THE PIECE WAS REMOVED FROM THE PATIENT. THE DEVICE WILL NOT BE RETURNED TO MSI

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19449710 ↗
  • MalfunctionJun 3, 2024Fracture; Fracture

    DURING PROCEDURE FOR PATIENT, WHILE WAS USING LAPAROSCOPIC INSTRUMENT WITH MICROLINE GRASPER TIP INSIDE THE PATIENT'S BODY, ONE OF THE PIECE OF GRASPER TIP CAME OFF. THE PIECE WAS RETRIREVED. THE DEVICE WILL NOT BE SENT BACK TO MICROLINE SURGICAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19449210 ↗
  • MalfunctionJun 21, 2023Break; Break

    SHRINK WRAP FELL OFF OF GRASPER INTO PATIENT ABDOMEN. (RETRIEVED AND REMOVED)

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17173102 ↗

Other Electrosurgical, Cutting & Coagulation & Accessories devices

All Electrosurgical, Cutting & Coagulation & Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page