ReNew 3925Recalls & reported problems
Microline Surgical, Inc. · Rigid Shaft Handpiece, 42cm, Reusable · Class 2
All models of ReNew (45)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00811099010616), received between Dec 28, 2018 and Oct 5, 2021 — classified as 2 injuries.
The most frequently reported specific device problem is “Arcing” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Arcing1
- Smoking1
Patient problems
- Bowel Burn1
- Burning Sensation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryOct 5, 2021Smoking; Smoking
REUSABLE LAPAROSCOPIC INSTRUMENTS AND DISPOSABLE TIP WERE INSPECTED PRIOR TO PUTTING TOGETHER FOR ANY VISUAL DEFECTS. NONE NOTED. WHILE SURGEON WAS USING A EXTRALONG MICROLINE LAPAROSCOPIC HANDLED INSTRUMENT WITH A DISPOSABLE SCISSOR TIP, SURGICAL FIRST ASSISTANT NOTICE A PLUME COMING FROM THE THREADED PORTION, THE DISTAL END OF INSTRUMENT. SURGEON WAS ASKED TO STOP AND IT WAS OBSERVED THAT THERE WAS 1/3 OF CM THERMAL DAMAGE. GENERAL SURGEON WAS CONSULTED AND OVERSEWED THE INJURY. THE INJURY DID NOT ENTER THE SEROSAL LAYER AND THE GENERAL SURGEON DID STATE THAT IT COULD HAVE BEEN LEFT ALONE BUT SINCE HE WAS THERE IT WAS OVERSEWN. THE AREA OF INJURY WAS THE DESCENDING COLON.
Patient problems: Bowel Burn
MAUDE report 12578150 ↗InjuryDec 28, 2018Arcing; Arcing
THERE WAS A BURNING LESION TO THE STOMACH AS A RESULT OF A DEFECT IN THE INSULATION FROM THE SHAFT OF THE LAPAROSCOPIC INSTRUMENT. ANESTHESIA TIME WAS EXTENDED FOR 15 MINUTES. NO FOLLOW-UP MEDICAL TREATMENT WAS REQUIRED.
Patient problems: Burning Sensation
MAUDE report 8202329 ↗
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.