ReNew 3941Recalls & reported problems

Microline Surgical, Inc. · ReNew XR Handpiece, 34cm · Class 2

All models of ReNew (45)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00811099013761), received between May 20, 2024 and Feb 19, 2025 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Arcing” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Arcing2
  • Electrical /Electronic Property Problem2
  • Insufficient Device Problem Information1
  • Sparking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Bowel Burn1
  • Burn(s)1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 19, 2025Sparking; Sparking

    DURING A THL PROCEDURE DR. SANTAMARIA FLORES AND OR STAFF NOTICED THAT THE MICROLINE HANDPIECE AND SCISSOR TIP BEGAN SPARKING AT THE END OF THE CASE DURING MONOPOLAR CAUTERIZATION. SETTINGS WERE REPORTED AS "35 CUT BLEND, COAG 35." THE INSTRUMENT WAS REMOVED FROM THE PATIENT, AND THE DISTAL END APPEARED MELTED AND CHARRED. THE INSTRUMENT WAS PUT ASIDE, AND THE MANUFACTURER WAS CONTACTED. NO BURNS OR INJURY TO PATIENT WERE REPORTED AT TIME OF FILING THIS AIR REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21413264 ↗
  • InjuryJan 17, 2025Electrical /Electronic Property Problem; Arcing; Electrical /Electronic Property Problem; Arcing

    WHILE OPERATING IN THE ABDOMINAL CAVITY A BURN ON OVARIAN TISSUE WAS NOTICED BY SURGEONS. THEY STATED THEY HAD THE DISP. SCISSOR TIP ON THE MICROLINE PISTOL HANDLE IN VIEW AT ALL TIMES AND DIDN'T SEE IT TOUCH THE TISSUE THAT APPEAR TO HAVE THE SCORCH MARK ON IT. INSTRUMENT WAS MARKED AND LABELED AS BROKEN; PLACED BACK IN THE TRAY FOR PROCESSING. PROCESSING CENTER INDICATES THAT THIS WAS PROCESSED PER PROCEDURE AND PASSED ALL QUALITY TEST OF LAPAROSCOPIC EQUIPMENT.

    Patient problems: Burn(s)

    MAUDE report 21179271 ↗
  • InjuryNov 8, 2024Electrical /Electronic Property Problem; Arcing; Electrical /Electronic Property Problem; Arcing

    DURING THE PROCEDURE, THE ASSISTING SURGEON DISCOVERED A MONOPOLAR BURN TO THE BOWEL OF THE PATIENT. THE SURGEONS THEN BACKTRACKED THROUGH THE PROCEDURE TO CHECK FOR OTHER NON-NOTICED BURNS TO THE PATIENT. SURGEONS REPORTED FINDING 3 ADDITIONAL MONO POLAR BURNS TO THE PATIENT BOWEL, FOR A TOAL OF 4 BURNS. SURGEONS DID NOT NOTICE THE BURNS UPON IMMEDIATE USE OF MONOPOLAR ELECTRO SURGICAL CAUTERY. NO VISUALIZATION OF ELECTRICAL ARCING WAS REPORTED AT THE TIME OF INCIDENT BY THE SURGEONS OR STAFF, HOWEVER THE DISTAL PORTION OF THE RENEW XR HANDPIECE DEMONSTRATES PHYSICAL DAMAGE. SOURCE OF MONOPOLAR ENERGY WAS RECORDED TO BE A VALLEY LAB FT10 SERIAL NUMBER (B)(6). SETTINGS USED WERE CUT 35 COAG BLEND, 35 COAG.

    Patient problems: Bowel Burn

    MAUDE report 20643496 ↗
  • MalfunctionMay 20, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    (B)(6) HOSPITAL IN (B)(6) , RI EXPERIENCED AN ELETRCIAL ARCING INCIDENT WHILE USING THE MONOPOLAR FUNCTION OF A MICROLINE XR HANDPIECE, AND A MICROLINE SCISSOR TIP. THE HANDPIECE, AND SCISSOR TIP HAVE BEEN RETRIEVED AND SENT BACK TO MICROLINE SURGICAL HEADQUARTERS FOR FURTHER INVESTIGATION. MONOPOLAR SETTINGS USED INSIDE OF THE PATIENT WERE 60/40, AND THEN 35/45. STAFF ATTHE FACILITY STATED THAT THE PATIENT WASN'T HARMED, HOWEVER THE INSTRUMENT DID ARC WHILE INSIDE THE PATIENT CAVITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19353023 ↗

Other Electrosurgical, Cutting & Coagulation & Accessories devices

All Electrosurgical, Cutting & Coagulation & Accessories devices →