ReNew 3941Recalls & reported problems
Microline Surgical, Inc. · ReNew XR Handpiece, 34cm · Class 2
All models of ReNew (45)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00811099013761), received between May 20, 2024 and Feb 19, 2025 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Arcing” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Arcing2
- Electrical /Electronic Property Problem2
- Insufficient Device Problem Information1
- Sparking1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Bowel Burn1
- Burn(s)1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 19, 2025Sparking; Sparking
DURING A THL PROCEDURE DR. SANTAMARIA FLORES AND OR STAFF NOTICED THAT THE MICROLINE HANDPIECE AND SCISSOR TIP BEGAN SPARKING AT THE END OF THE CASE DURING MONOPOLAR CAUTERIZATION. SETTINGS WERE REPORTED AS "35 CUT BLEND, COAG 35." THE INSTRUMENT WAS REMOVED FROM THE PATIENT, AND THE DISTAL END APPEARED MELTED AND CHARRED. THE INSTRUMENT WAS PUT ASIDE, AND THE MANUFACTURER WAS CONTACTED. NO BURNS OR INJURY TO PATIENT WERE REPORTED AT TIME OF FILING THIS AIR REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21413264 ↗InjuryJan 17, 2025Electrical /Electronic Property Problem; Arcing; Electrical /Electronic Property Problem; Arcing
WHILE OPERATING IN THE ABDOMINAL CAVITY A BURN ON OVARIAN TISSUE WAS NOTICED BY SURGEONS. THEY STATED THEY HAD THE DISP. SCISSOR TIP ON THE MICROLINE PISTOL HANDLE IN VIEW AT ALL TIMES AND DIDN'T SEE IT TOUCH THE TISSUE THAT APPEAR TO HAVE THE SCORCH MARK ON IT. INSTRUMENT WAS MARKED AND LABELED AS BROKEN; PLACED BACK IN THE TRAY FOR PROCESSING. PROCESSING CENTER INDICATES THAT THIS WAS PROCESSED PER PROCEDURE AND PASSED ALL QUALITY TEST OF LAPAROSCOPIC EQUIPMENT.
Patient problems: Burn(s)
MAUDE report 21179271 ↗InjuryNov 8, 2024Electrical /Electronic Property Problem; Arcing; Electrical /Electronic Property Problem; Arcing
DURING THE PROCEDURE, THE ASSISTING SURGEON DISCOVERED A MONOPOLAR BURN TO THE BOWEL OF THE PATIENT. THE SURGEONS THEN BACKTRACKED THROUGH THE PROCEDURE TO CHECK FOR OTHER NON-NOTICED BURNS TO THE PATIENT. SURGEONS REPORTED FINDING 3 ADDITIONAL MONO POLAR BURNS TO THE PATIENT BOWEL, FOR A TOAL OF 4 BURNS. SURGEONS DID NOT NOTICE THE BURNS UPON IMMEDIATE USE OF MONOPOLAR ELECTRO SURGICAL CAUTERY. NO VISUALIZATION OF ELECTRICAL ARCING WAS REPORTED AT THE TIME OF INCIDENT BY THE SURGEONS OR STAFF, HOWEVER THE DISTAL PORTION OF THE RENEW XR HANDPIECE DEMONSTRATES PHYSICAL DAMAGE. SOURCE OF MONOPOLAR ENERGY WAS RECORDED TO BE A VALLEY LAB FT10 SERIAL NUMBER (B)(6). SETTINGS USED WERE CUT 35 COAG BLEND, 35 COAG.
Patient problems: Bowel Burn
MAUDE report 20643496 ↗MalfunctionMay 20, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
(B)(6) HOSPITAL IN (B)(6) , RI EXPERIENCED AN ELETRCIAL ARCING INCIDENT WHILE USING THE MONOPOLAR FUNCTION OF A MICROLINE XR HANDPIECE, AND A MICROLINE SCISSOR TIP. THE HANDPIECE, AND SCISSOR TIP HAVE BEEN RETRIEVED AND SENT BACK TO MICROLINE SURGICAL HEADQUARTERS FOR FURTHER INVESTIGATION. MONOPOLAR SETTINGS USED INSIDE OF THE PATIENT WERE 60/40, AND THEN 35/45. STAFF ATTHE FACILITY STATED THAT THE PATIENT WASN'T HARMED, HOWEVER THE INSTRUMENT DID ARC WHILE INSIDE THE PATIENT CAVITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19353023 ↗
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page