Stryker Corporation · The PhotonBlade is an RF device coupled with illumination that is indicated for cutting and coagulation of soft tissue during surgical procedures. It is intended to be used with a variety of standard commercially available electrosurgical units. · Class 2

All models of PhotonBlade (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00816728020158), received between Jun 6, 2017 and Oct 18, 2024 — classified as 2 injuries, 16 malfunctions.

The most frequently reported specific device problem is “Flaked” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Flaked5
  • Adverse Event Without Identified Device or Use Problem4
  • Energy Output To Patient Tissue Incorrect3
  • Fracture2
  • Break1
  • Device Remains Activated1
  • Out-Of-Box Failure1
  • Sparking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient5
  • Superficial (First Degree) Burn5
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 18, 2024Flaked; Flaked

    IT WAS REPORTED THAT DURING A PROCEDURE, THE COATING ON THE TIP OF THE PHOTONBLADE FELL INSIDE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20480498 ↗
  • MalfunctionMay 1, 2024Flaked; Flaked

    IT WAS REPORTED THAT DURING THE PROCEDURE THE COATING ON THE ELECTRODE TIP WAS FLAKING OFF. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19220899 ↗
  • MalfunctionDec 6, 2022Flaked; Flaked

    IT WAS REPORTED THAT DURING A MASTECTOMY, THE COATING ON THE PHOTON BLADE TIP FLAKED OFF WITHIN TEN MINUTES OF USE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15926954 ↗
  • MalfunctionOct 19, 2022Flaked; Flaked

    IT WAS REPORTED THAT DURING A PROCEDURE, THE PROTECTIVE COATING ON THE PHOTON BLADE STARTED CHIPPING OFF. IT WAS ALSO REPORTED THAT THERE WAS DELAY OF A FEW MINUTES TO REMOVE THE CHIPS. IT WAS FURTHER REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15631383 ↗
  • MalfunctionSep 6, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE TIP OF THE PHOTON BLADE BROKE. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15362326 ↗
  • MalfunctionMar 28, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A TOTAL HIP ARTHROPLASTY BLUE RING AROUND THE LIGHT FELL OFF INTO THE PATIENT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13929376 ↗
  • MalfunctionOct 12, 2021Flaked; Flaked

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE WHILE THE BREAST WAS OPEN, THE COATING MATERIAL FROM THE PHOTON BLADE FLAKED OFF. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12614995 ↗
  • MalfunctionJul 4, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    APPROXIMATELY (B)(6) 2018, DURING A PACEMAKER CHANGE OUT THE INSULATION OF A LEAD WAS ACCIDENTALLY DAMAGED. A REPAIR KIT WAS USED TO FIX THE LEAD INTRAOPERATIVELY. THE CASE WAS COMPLETED WITH NO FURTHER INCIDENT AND NO IMPACT TO THE PATIENT. THIS INCIDENT WAS REPORTED TO INVUITY ON (B)(4) 2018.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7661760 ↗
  • MalfunctionMay 16, 2018Device Remains Activated; Device Remains Activated

    DURING A CASE THE PHOTONBLADE REMAINED ACTIVE AFTER THE BUTTON WAS RELEASED. THERE WAS NO PATIENT INJURY. THE DEVICE WAS UNPLUGGED AND THE CASE WAS COMPLETED USING A DIFFERENT DEVICE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7520882 ↗
  • MalfunctionFeb 8, 2018Break; Break

    TOWARD THE END OF A BI-LATERAL NIPPLE SPARING MASTECTOMY, THE COLLAR THAT LOCKS/UNLOCKS THE TELESCOPING SHAFT BROKE. TWO MAIN HALVES WERE RETRIEVED IMMEDIATELY AND A THIRD SMALLER PIECE COULD NOT BE ACCOUNTED FOR. THE SURGEON DID AN EXAMINATION OF THE SURGICAL CAVITY AND DID NOT SEE A THIRD PIECE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7252965 ↗
  • MalfunctionFeb 6, 2018Out-Of-Box Failure; Out-Of-Box Failure

    WHEN THE PHOTON BLADE WAS INITIALLY CONNECTED TO THE GENERATOR, A SOLID AUDIBLE TONE WAS EMITTED BY THE ESU. IT WAS OBSERVED THAT THE PATIENT RECEIVED TWO VERY MINOR BURNS. THE CASE WAS COMPLETED USING A SECOND DEVICE.

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 7245971 ↗
  • MalfunctionJan 5, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO AN INVUITY REPRESENTATIVE ON (B)(4) 2017, THAT THE INSULATION OF A LEAD WAS ACCIDENTALLY DAMAGED DURING A BATTERY REPLACEMENT OF AN ICD. A SUTURE SLEEVE WAS APPLIED OVER THE DAMAGED AREA OF THE INSULATION. THE CASE WAS COMPLETED WITH NO FURTHER INCIDENT AND NO IMPACT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7170453 ↗
  • MalfunctionJan 5, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO AN INVUITY REPRESENTATIVE ON (B)(6) 2017 THAT DURING A BATTERY REPLACEMENT OF AN ICD THE INSULATION OF A LEAD WAS ACCIDENTALLY MELTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7170446 ↗
  • InjuryJun 29, 2017Energy Output To Patient Tissue Incorrect; Energy Output To Patient Tissue Incorrect

    DURING A BREAST BIOPSY PROCEDURE, THE PHYSICIAN NOTED A SMALL AMOUNT OF ENERGY DISCHARGE COMING FROM THE SHAFT OF THE DEVICE. THIS RESULTED IN A SMALL BURN (<1 MM) AT THE OPENING OF THE INCISION. THE DOCTOR REMOVED THE AFFECTED TISSUE AND CLOSED THE INCISION.

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 6679528 ↗
  • MalfunctionJun 29, 2017Energy Output To Patient Tissue Incorrect; Energy Output To Patient Tissue Incorrect

    DURING A LUMPECTOMY PROCEDURE, THE PHYSICIAN NOTED A BURN NEAR THE PERI-AREIOLAR INCISION. THERE WAS NO MEDICAL OR SURGICAL INTERVENTION REQUIRED.

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 6679521 ↗
  • MalfunctionJun 29, 2017Energy Output To Patient Tissue Incorrect; Energy Output To Patient Tissue Incorrect

    DURING A LUMPECTOMY PROCEDURE, THE PHYSICIAN NOTED A SUPERFICIAL BURN ON THE SKIN. THERE WAS NO FOLLOW-UP TREATMENT REPORTED.

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 6679511 ↗
  • MalfunctionJun 29, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING AN ICD GEN CHANGE PROCEDURE, THE INSULATION OF A LEAD WAS ACCIDENTALLY CUT. THE LEAD WAS REPAIRED AND ITS FUNCTION WAS TESTED; THE CASE WAS COMPLETED WITHOUT INCIDENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6679508 ↗
  • InjuryJun 6, 2017Therapy Delivered to Incorrect Body Area; Sparking; Therapy Delivered to Incorrect Body Area; Sparking

    DURING A LUMPECTOMY PROCEDURE, THE PHYSICIAN NOTICED THAT THE DISTAL TIP OF THE TELESCOPING WAVEGUIDE FELT HOT. THE PHYSICIAN EXAMINED THE INCISION AREA AND NOTICED A BURN NEAR THE PERI-AREOLA REGION THAT REQUIRED A SURGICAL INTERVENTION.

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 6614502 ↗

Other Electrosurgical, Cutting & Coagulation & Accessories devices

All Electrosurgical, Cutting & Coagulation & Accessories devices →