PhotonBlade with Adaptive Smoke Evacuation PB2SERecalls & reported problems
Stryker Corporation · The PhotonBlade with Adaptive Smoke Evacuation is a monopolar RF device coupled with illumination that is indicated for cutting and coagulation of tissue during general surgical procedures and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Stryker PhotonBlade is a hand-held electrosurgical pencil used to cut and seal tissue during surgery, with a light and built-in smoke evacuation tubing. The 32 reports mostly describe the blade tip cracking or its coating flaking off during procedures, and the plastic clip that holds the smoke evacuation tubing breaking loose. No deaths or injuries are reported, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 32
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 00816728021261), received between Nov 8, 2021 and Sep 3, 2025 — classified as 32 malfunctions.
The most frequently reported specific device problem is “Fracture” (29 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture29
- Flaked4
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions32
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 3, 2025Fracture; Fracture
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22951465 ↗MalfunctionSep 11, 2024Flaked; Fracture; Flaked; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE, THAT THE PHOTON BLADE¿S INSULATION DELAMINATED OFF ELECTRODE TIP AND PLASTIC ON SMOKE EVACUATION LINE BROKE OFF. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20194387 ↗MalfunctionSep 9, 2024Flaked; Flaked
IT WAS REPORTED THAT PROTECTIVE COATING ON THE TIP CAME OFF DURING A SURGICAL PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20170414 ↗MalfunctionJul 10, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE THE PLASTIC CLIP HOLDING THE SMOKE EVACUATION TUBING TO THE PHOTONBLADE BROKE OFF. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19716174 ↗MalfunctionJul 3, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE, THE CLIP THAT HOLDS THE SMOKE EVACUATION TUBING TO THE PENCIL BROKE OFF. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19666832 ↗MalfunctionJul 3, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE, THE CLIP THAT HOLDS THE SMOKE EVACUATION TUBING TO THE PENCIL BROKE OFF. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19666826 ↗MalfunctionMay 23, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE THE CLIP THAT HOLDS THE SUCTION TUBE TO THE PHOTONBLADE BROKE OFF. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19382751 ↗MalfunctionMay 10, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A BREAST SURGERY, A CLIP HOLDING THE SMOKE EVACUATION TUBING TO THE PHOTONBLADE BROKE OFF. THE DOCTOR IS UNSURE IF THE CLIP FELL INTO THE PATIENT OR NOT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO MEDICAL INTERVENTION WAS REPORTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 19287844 ↗MalfunctionMay 3, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE THE CLIP ATTACHING THE SUCTION TUBING TO THE PHOTONBLADE HAND PIECE BROKE OFF AND FELL INTO THE PATIENT'S WOUND. THE SURGEON HAD TO REMOVE THE PLASTIC PIECE FROM THE SURGICAL SITE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19238740 ↗MalfunctionApr 3, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE, THE CLIP HOLDING THE SMOKE EVACUATION TUBING TO THE PHOTON BLADE BROKE OFF. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED. THIS IS EVENT #1 OF 2.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19038017 ↗MalfunctionApr 3, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE, THE CLIP HOLDING THE SMOKE EVACUATION TUBING TO THE PHOTON BLADE BROKE OFF. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED. THIS IS EVENT #2 OF 2.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19038010 ↗MalfunctionDec 18, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, TWO CLIPS THAT ATTACHES THE SMOKE EVACUATION TUBING TO THE PHOTON BLADE BROKE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18352600 ↗MalfunctionDec 4, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE CLAMP ON THE PHOTON BLADE THAT ATTACHES THE SMOKE TUBING BROKE, THE BROKEN FRAGMENTS WERE RETRIEVED FROM THE SURGICAL SITE RESULTING IN A FIVE TO TEN MINUTES DELAY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY; NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18255434 ↗MalfunctionNov 14, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, PART OF A PHOTON BLADE WITH ADAPTIVE SMOKE EVACUATION BROKE OFF. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18132401 ↗MalfunctionSep 22, 2023Fracture; Fracture
IT WAS REPORTED THAT THE CLAMP THAT ATTACHES THE SMOKE TUBING TO THE PHOTON BLADE MELTED AND FRACTURED IN 3 DIFFERENT PROCEDURES. IT WAS FURTHER REPORTED THAT FOR ALL 3 PROCEDURES, THERE WAS A TOTAL DELAY OF 2 HOURS TO RETRIEVE THE FRAGMENTS AND ENSURE NO FRAGMENTS REMAINED IN THE SURGICAL SITE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED. THIS MDR REPORT IS FOR THE SECOND EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17802204 ↗MalfunctionSep 22, 2023Fracture; Fracture
IT WAS REPORTED THAT THE CLAMP THAT ATTACHES THE SMOKE TUBING TO THE PHOTON BLADE MELTED AND FRACTURED IN 3 DIFFERENT PROCEDURES. IT WAS FURTHER REPORTED THAT FOR ALL 3 PROCEDURES, THERE WAS A TOTAL DELAY OF 2 HOURS TO RETRIEVE THE FRAGMENTS AND ENSURE NO FRAGMENTS REMAINED IN THE SURGICAL SITE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED. THIS MDR REPORT IS FOR THE THIRD EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17801873 ↗MalfunctionSep 22, 2023Fracture; Fracture
IT WAS REPORTED THAT THE CLAMP THAT ATTACHES THE SMOKE TUBING TO THE PHOTON BLADE MELTED AND FRACTURED IN 3 DIFFERENT PROCEDURES. IT WAS FURTHER REPORTED THAT FOR ALL 3 PROCEDURES, THERE WAS A TOTAL DELAY OF 2 HOURS TO RETRIEVE THE FRAGMENTS AND ENSURE NO FRAGMENTS REMAINED IN THE SURGICAL SITE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED. THIS MDR REPORT IS FOR THE FIRST EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17800321 ↗MalfunctionJul 31, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE PLASTIC CLIP ON THE PB2SE BLADE BROKE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17423921 ↗MalfunctionMar 27, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE C-CLAMP ON THE PB2SE BLADE BROKE, THE FRAGMENT WAS RECOVERED FROM THE SURGICAL SITE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16625152 ↗MalfunctionMar 27, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE C-CLAMP ON THE PB2SE BLADE BROKE, THE FRAGMENT WAS RECOVERED FROM THE SURGICAL SITE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16624967 ↗MalfunctionMar 27, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE C-CLAMP ON THE PB2SE BLADE BROKE, THE FRAGMENT WAS RECOVERED FROM THE SURGICAL SITE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16624965 ↗MalfunctionMar 27, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE C-CLAMP ON THE PB2SE BLADE BROKE, THE FRAGMENT WAS RECOVERED FROM THE SURGICAL SITE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16624963 ↗MalfunctionMar 13, 2023Break; Fracture; Break; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE PLASTIC CLIP BROKE AND FELL INTO THE PATIENT, THE BROKEN FRAGMENT WAS RETRIEVED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16528493 ↗MalfunctionNov 28, 2022Flaked; Flaked
IT WAS REPORTED THAT DURING A MASTECTOMY, THE COATING ON THE PHOTON BLADE FLAKED OFF. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15874599 ↗MalfunctionApr 26, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE CLIP FELL OFF THE PHOTON BLADE. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14210575 ↗
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page