The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00817371020014), received between Nov 17, 2023 and Nov 17, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Human-Device Interface Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Human-Device Interface Problem1

Patient problems

  • Burn(s)1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 17, 2023Human-Device Interface Problem; Human-Device Interface Problem

    PHYSICIAN'S OFFICE ACCT# (B)(4) REPORTED A BURN ON (B)(6) 2021 TO SALES REPRESENTATIVE. THERMITIGHT PROCEDURE WAS PERFORMED ON (B)(6) 2023 ON THE PATIENT'S NECK AND SUBMENTAL AREA AFTER SUBMENTAL LIPSUCTION. THE TREATMENT SET TEMPERATURE WAS 65C, AND DURING TREATMENT THE READINGS WERE BETWEEN 65-67C. THE SURGICAL TECHNICIAN MONITORING THE "EXTERNAL TEMP" READ A TEMPERATURE RANGE OF 35-40C. DURING TREATMENT THERE WAS NO INDICATION OF BURNS, SKIN WAS WARM AND PERFUSING APPROPRIATELY. TREATMENT TIME OF THE PROCEDURE WAS 20 MINUTES. AFTER THE TREATMENT WHILE CLEANING, IT WAS NOTICED THAT THE PATIENT'S SKIN WAS PINK AND PEELING ON THE LEFT SIDE OF THE PATIENT'S NECK BUT DID NOT EXHIBIT OPEN AREAS INDICATIVE OF COOL COMPRESS AND VASELINE. THE PATIENT WAS "WRAPPED" PER USUAL POST PROCEDURE. THE DAY FOLLOWING THE PROCEDURE, THE PATIENT RETURN WITH SIGNIFICANT BLISTERING ON RIGHT SIDE OF NECK AND OPEN AREAS ON LEFT SIDE OF THE NECK AND BURNS IN THE SUBMENTAL AREA. IMAGES OF THE PATIENT WAS PROVIDED, OFFICE PERSONNEL NOTED THAT THE PROTOCOL HAS BEEN FOLLOWED, AND THAT THERE WERE TWO PEOPLE MONITORING TEMPERATURE WITH NO INDICATION THAT THE SKIN WAS "TOO HOT" CITING CONCERNS ABOUT THE INFRARE

    Patient problems: Burn(s)