Coblation AC2340-01Recalls & reported problems

Smith & Nephew, Inc. · SIDEWINDER BLADE 55 DEGREE · Class 2

All models of Coblation (62)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00817470000696), received between Feb 28, 2018 and May 5, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Arcing1
  • Arcing of Electrodes1
  • Device Alarm System1
  • Material Deformation1
  • Material Too Rigid or Stiff1
  • Noise, Audible1
  • Protective Measures Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 5, 2026Break

    IT WAS REPORTED THAT DURING A PROCEDURE, THE TIP OF THE SIDEWINDER BLADE ICW BROKE AFTER OPERATING THE DEVICE A FEW TIMES. THE BROKEN TIP DID NOT FELL INSIDE PATIENT. IT IS UNKNOWN HOW THE PROCEDURE WAS PERFORMED OR IF THERE WAS ANY DELAY. NO PATIENT INJURY OR FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25086254 ↗
  • MalfunctionJan 16, 2024Material Too Rigid or Stiff; Arcing of Electrodes; Material Deformation; Material Too Rigid or Stiff; Arcing of Electrodes; Material Deformation

    IT WAS REPORTED THAT, DURING A MENISCAL REPAIR, THE SIDEWINDER BLADE COULD NOT BEND. THERE WAS A BACK-UP DEVICE AVAILABLE AND A NON-SIGNIFICANT DELAY WAS REPORTED. NO PATIENT INJURIES OR OTHER COMPLICATIONS WERE REPORTED. PRELIMINARY RESULTS OF INVESTIGATION FOUND THAT THE BLACK SHRINK WRAP IS DEFORMED AND TORN AT THE DISTAL END AND SHOWS BARE METAL UNDERNEATH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18519120 ↗
  • MalfunctionJul 21, 2021Arcing; Protective Measures Problem; Arcing; Protective Measures Problem

    IT WAS REPORTED THAT THE QUANTUM2 DISPLAYED A HARDWIRE FAILURE ERROR SOON AFTER USING THE WAND IN AN ARTHROSCOPY. AFTER THE PROBLEM, A FUNCTIONALITY CHECK WAS DONE AND OCCURRED THAT A FLEXIBLE PART OF THE WAND OCCURRED AN ELECTRIC SPARK. THIS FAILURE MODE WAS CONFIRMED DURING FUNCTIONALLY CHECK FOR THE DEVICE IN THE AIR, NOT IN THE SALINE SOLUTION. NO PATIENT INJURY, DELAY OR OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12207521 ↗
  • MalfunctionNov 23, 2018Break

    IT WAS REPORTED THAT THE TIP DID NOT WORK. THERE WAS NOT DELAY AND NO PATIENT INJURIES. NO BACK-UP DEVICE AVAILABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8099991 ↗
  • MalfunctionFeb 28, 2018Device Alarm System; Break; Noise, Audible

    IT WAS REPORTED THAT DURING USE, A BREAKAGE SOUND WAS HEARD AND THE CONTROLLER ALARMED UNTIL THE WAND WAS DISCONNECTED. UPON INSPECTION OF THE WAND, IT WAS FOUND THAT THE SHEATH (INSULATION) WAS DAMAGED. ANOTHER WAND WAS OPENED TO COMPLETE THE CASE. NO PATIENT/USER INJURY OR OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7304372 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →