COBLATOR II Controller 120V (SRVC controller-only) 13546-59Recalls & reported problems
Smith & Nephew, Inc. · COBLATOR II Controller Only 120V Refurb · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00817470001792), received between Jun 21, 2020 and Jul 13, 2020 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Defective Component” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K030108); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Defective Component2
- Connection Problem1
Patient problems
- No Information1
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 13, 2020Connection Problem; Connection Problem
IT WAS REPORTED THAT THE COBLATOR CONTROLLER WAS NOT WORKING. IT IS UNKNOWN WHEN THE ISSUE WAS DISCOVERED, IF A DELAY HAPPENED AND IF A BACKUP DEVICE WAS AVAILABLE. HOWEVER, NO PATIENT INJURIES WERE REPORTED. RESULTS OF INVESTIGATION HAVE CONCLUDED THAT THIS UNIT DOES NOT RECOGNIZE THE WAND WHICH MAKES IT A REPORTABLE EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Information
MAUDE report 10268534 ↗MalfunctionJul 7, 2020Defective Component; Defective Component
IT WAS REPORTED THAT THE FOOT PEDAL INSERTION WAS PUSHED ON, THE GENERATOR WAS INOPERABLE. IT IS UNKNOWN WHETHER THERE WAS A BACK UP AVAILABLE OR DELAY IN THE CASE AND IF THE EVENT HAPPENED DURING SURGERY. NO PATIENT INJURIES WERE REPORTED. RESULTS OF INVESTIGATION HAVE CONCLUDED THAT THIS UNIT HAD THE FOOT PEDAL PORT DAMAGED WHICH MAKES IT A REPORTABLE EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10239823 ↗MalfunctionJun 21, 2020Defective Component; Defective Component
IT WAS REPORTED THAT THE CONTROLLER CONNECTOR FOR THE FOOT SWITCH WAS DAMAGED. NO CASE OR PATIENT INVOLVED. WAS INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Patient Involvement
MAUDE report 10177246 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K030108. One submission often covers many models, so check the product and lot details in each record.
- Z-1323-2015 · Feb 20, 2015 · TerminatedCoblator II System, 120V. Product Usage: The Coblator II (Model RF8000E) is a bipolar, radiofrequency electrosurgical system designed for use in otorhinolaryngology (ENT) surger
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page