Skip to content

Coblation EICA4835-01Recalls & reported problems

Smith & Nephew, Inc. · REFLEX ULTRA PTR WITH INTEGRATED CABLE · Class 2

All models of Coblation (62)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00817470003871), received between Mar 10, 2019 and Feb 13, 2025 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Output Problem2
  • Failure to Conduct1
  • Premature End-of-Life Indicator1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2
  • Foreign Body In Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 13, 2025Break; Break

    IT WAS REPORTED THAT DURING AN ENT PROCEDURE, THE REFLEX ULTRA PTR COBLATOR II WAND BLADE BROKE INSIDE THE PATIENT¿S NOSE-TURBINATE. THE SURGEON STATED THAT ATTEMPTING TO REMOVE THE BROKEN PIECE COULD CAUSE MORE HARM, SO THE PIECE WAS LEFT INSIDE THE PATIENT¿S NOSE. IT IS UNKNOWN WHETHER A BACK-UP DEVICE WAS AVAILABLE OR IF THERE WAS A SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 21375932 ↗
  • MalfunctionDec 6, 2023Break; Break

    IT WAS REPORTED THAT DURING AN ENT, A REFLEX ULTRA WAND TIP BENT DOWN UPON WITHDRAWAL FROM THE PATIENT´S TURBINATE AND AFTERWARDS THE TIP BROKE OFF OUTSIDE OF THE SURGICAL FIELD. SURGERY CONTINUED WITHOUT ANY DELAY; HOWEVER IT IS UNKNOWN HOW WAS COMPLETED. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18279595 ↗
  • MalfunctionNov 30, 2023Failure to Conduct; Premature End-of-Life Indicator; Failure to Conduct; Premature End-of-Life Indicator

    IT WAS REPORTED THAT DURING AN ENT PROCEDURE, WHEN TWO (2) REFLEX ULTRA WERE PLUGGED IN, THEY DISPLAYED "EXPIRED HANDPIECE". THE PROCEDURE WAS COMPLETED WITH A COMPETITOR DEVICE AND WITH A SURGICAL DELAY GREATER THAN 30 MINUTES. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18236644 ↗
  • MalfunctionJul 17, 2020Output Problem; Output Problem

    IT WAS REPORTED THAT DURING A TURBINATE REDUCTION THE REFLEX ULTRA COBLATOR DEFAULT WAS 0 AND 0 INSTEAD 4 AND 2. THE WAND WOULD NOT ABLATE OR CREATE PLASMA. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10292670 ↗
  • MalfunctionMar 10, 2019Output Problem

    IT WAS REPORTED THAT DURING TURBINATE REDUCTION SURGERY, WHEN THE ABLATE FOOT PENAL WAS INITIALLY DEPRESSED, THE ABLATE SETTING WENT TO ¿0¿ AND NOTHING OCCURRED. A NEW WAND WAS OPENED BUT THE SAME ISSUE OCCURRED. PROCEDURE WAS CANCELLED. NO PATIENT INJURIES WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8406503 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page