Skip to content

Coblation EICA5874-01Recalls & reported problems

Smith & Nephew, Inc. · EVAC 70 XTRA HP WAND · Class 2

All models of Coblation (62)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00817470004007), received between Jan 30, 2018 and Feb 2, 2026 — classified as 5 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Arcing” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Arcing2
  • Melted2
  • Sparking2
  • Decrease in Suction1
  • Defective Device1
  • Failure to Deliver Energy1
  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Burn(s)2
  • Burn, Thermal1
  • Hemorrhage/Blood Loss/Bleeding1
  • No Information1
  • No Known Impact Or Consequence To Patient1
  • Partial thickness (Second Degree) Burn1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 2, 2026Melted

    IT WAS REPORTED THAT, DURING A T&A PROCEDURE, THE TIP OF THE PRECISE COBLATOR MELTED. IT IS UNKNOWN IF THERE WAS A BACKUP DEVICE AVAILABLE, HOWEVER THERE WAS A DELAY GREATER THAN 30 MINUTES. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24240953 ↗
  • MalfunctionFeb 2, 2026Melted

    IT WAS REPORTED THAT, DURING A TONSILLECTOMY AND ADENOIDECTOMY, THE TIP OF AN EVAC 70 XTRA WAND MELTED. THE PROCEDURE WAS RESUMED, AFTER A DELAY GREATER THAN 30 MINUTES, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24240568 ↗
  • MalfunctionFeb 9, 2023Arcing; Arcing

    IT WAS REPORTED THAT DURING A TONSILLECTOMY, THE EVAC 70 XTRA COBLATOR II WAND WAS ARCING WHILE USING THE SETTINGS 9-5. THE SETTINGS WERE MOVED TO 8-4 AND IT HELPED, BUT STILL NOTICED ARCING. THE PROCEDURE WAS COMPLETED WITH A SURGICAL DELAY OF 10 MINUTES USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16335121 ↗
  • InjuryJan 19, 2023No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT DURING A TONSILLECTOMY AND ADENOIDECTOMY PROCEDURE USING AN EVAC 70 XTRA COBLATOR II WAND, A BURN WAS NOTICED IN THE MOUTH IN AN AREA WHERE THE TIP WAS NOT ACTIVATED AT ANY POINT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A SURGICAL DELAY OF 1-2 MINUTES USING A BACK-UP DEVICE. THE BURN WAS EVALUATED BY A PLASTIC SURGEON, NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: Burn(s)

    MAUDE report 16206695 ↗
  • InjuryMar 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, DURING A TONSILLECTOMY/ADENOIDECTOMY USING A EVAC COBLATION WAND CONNECTED TO A COBLATOR CONTROLLER, THE WAND CONDUCTED THROUGH AN ALLIS CLAMP BURNT THE LIP OF THE PATIENT. IT WAS A SECOND DEGREE BURN WITH SHALLOW ULCERATION AT SITE WHICH WAS TREATED WITH ANTIBIOTIC OINTMENT AND IT IS NOW COMPLETELY HEALED UP. THE PROCEDURE WAS COMPLETED WITH A NON-SIGNIFICANT DELAY BY CHANGING THE SURGICAL TECHNIQUE TO ELECTROCAUTERY WITH SUCTION INSTEAD. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 13934813 ↗
  • MalfunctionMar 29, 2022Smoking; Sparking; Smoking; Sparking

    IT WAS REPORTED THAT, DURING A TONSILLECTOMY/ADENOIDECTOMY USING A EVAC COBLATION WAND CONNECTED TO A COBLATOR CONTROLLER, WHEN THE WAND WAS BEING USED, IT SPARKED AND PRODUCED SMOKE. THE PROCEDURE WAS COMPLETED WITH A NON-SIGNIFICANT DELAY BY CHANGING THE SURGICAL TECHNIQUE TO ELECTROCAUTERY WITH SUCTION INSTEAD. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13934802 ↗
  • InjuryAug 18, 2021Overheating of Device; Overheating of Device

    IT WAS REPORT THAT DURING TONSIL SURGERY, THE SIDE OF THE COBLATOR WAND WAS HOT AND LEFT A BURN MARK NEXT TO THE TONSILS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING A S+N BACK-UP DEVICE. NO FURTHER COMPLICATIONS REPORTED.

    Patient problems: Burn(s)

    MAUDE report 12337726 ↗
  • MalfunctionApr 15, 2021Failure to Deliver Energy; Failure to Deliver Energy

    IT WAS REPORTED THAT THE EVAC 70 WOULD NOT WORK AFTER PLUGGING IT IN. NO ARC NOTED PER PHYSICIAN. UNPLUGGED AND RESTARTED MACHINE AND STILL WAND WOULD NOT WORK, THERE WAS AN AUDIBLE TONE BUT THE STAFF AND DOCTOR SAID IT SOUNDED DIFFERENT THAN NORMAL. NO CASE INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11671368 ↗
  • MalfunctionFeb 11, 2021Arcing; Arcing

    IT WAS REPORTED THAT DURING A SURGERY, THE EVAC 70 WAND WAS ARCHING. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT A SIGNIFICANT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11319822 ↗
  • MalfunctionSep 6, 2018Sparking

    IT WAS REPORTED THAT THE DEVICE SPARKED.

    Patient problems: No Information

    MAUDE report 7851437 ↗
  • MalfunctionAug 17, 2018Defective Device

    IT WAS REPORTED THAT THE DEVICE DID NOT ABLATING FOR COAGING PROPERLY IT WAS WEAK. THERE WAS NO BACK UP DEVICE AVAILABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7794234 ↗
  • InjuryAug 12, 2018Decrease in Suction

    IT WAS REPORTED THAT THE WAND WAS NOT SUCKING NACI BACK INTO THE TIP OF THE WAND. PATIENT EXPERIENCED POST OPERATIVE BLEEDING.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 7775823 ↗
  • InjuryJan 30, 2018Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE COBLATOR II BURNED A PATIENT ON THE LIP.

    Patient problems: Burn, Thermal

    MAUDE report 7229210 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page