Coblation EICA7070-01Recalls & reported problems

Smith & Nephew, Inc. · PROCISE LW WAND · Class 2

All models of Coblation (62)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
16
Deaths reported
0
Injuries reported
12
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 00817470004243), received between Oct 30, 2019 and Feb 18, 2025 — classified as 12 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Material Separation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem9
  • Material Separation3
  • Break1
  • Component Missing1
  • Connection Problem1
  • Detachment of Device or Device Component1
  • Fire1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Restenosis4
  • Cyst(s)2
  • Adhesion(s)1
  • Burn(s)1
  • Hemorrhage/Blood Loss/Bleeding1
  • Insufficient Information1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 18, 2025Fire; Fire

    IT WAS REPORTED THAT, DURING A BILATERAL EAR TUBE REPLACEMENT, MICRODIRECT LARYNGOSCOPY, BRONCHOSCOPY & SUPRAGLOTTOPLASTY PROCEDURE, WHILE THE PATIENT HAD THE ET TUBE, SCOPE AND HAD USED A PROCISE COBLATOR II FOR 5 MINUTES IN THE AIRWAY, THE MOUTH CAUGHT INTO FLAMES. A "FIRE-BALL" WAS NOTED WITHIN THE ENDOTRACHEAL TUBE (POSITIONED IN THE OROPHARYNX). ALL TOOLS WERE IMMEDIATELY REMOVED FROM THE PATIENT'S MOUTH/AIRWAY AND SALINE WAS POURED ONTO SITE. WHEN THE PATIENT WAS STABILIZED, THE AREA INVOLVED WAS FLUSHED OUT WITH SALINE. THE EVENT TERMINATED THE PROCEDURE, NO FURTHER COBLATION WAS DONE. THERE WAS NO DELAY SINCE THE SURGEON WAS FINISHING THE CASE WHEN THE INCIDENT HAPPENED. THE PATIENT REQUIRED RE-INTUBATION WITH A NEW ET TUBE AND BRONCHOSCOPY WAS COMPLETED FOR AIRWAY ASSESSMENT. THE PATIENT DEVELOPED SIGNIFICANT AIRWAY BURNS AND REMAINED INTUBATED FOR MULTIPLE DAYS POST-INCIDENT.

    Patient problems: Burn(s)

    MAUDE report 21400312 ↗
  • MalfunctionMar 29, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT, DURING AN UNSPECIFIED PROCEDURE, A THE PROCISE'S ELECTRODE FELL OFF EXTERNAL TO THE PATIENT WHEN TRYING TO CLEAN IT. THE SURGERY WAS RESUMED, AFTER A 5 MINUTES DELAY, WITH A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16639910 ↗
  • MalfunctionDec 19, 2022Component Missing; Component Missing

    IT WAS REPORTED THAT, DURING A SUPRAGLOTTOPLASTY, THE TIP OF THE PROCISE COBLATOR II WAND WAS MISSING AN ELECTRODE. IT IS NOT CERTAIN WHEN THE ELECTRODE WENT MISSING, HOWEVER, IT DID NOT FALL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED AFTER A 1-MINUTE DELAY USING A BACK-UP DEVICE INSTEAD. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16003148 ↗
  • MalfunctionSep 28, 2021Break; Break

    IT WAS REPORTED THAT, THE PIECE OF THE TIP OF PROCISE LW COBLATOR II, BROKE OFF. IT IS UNKNOWN WHEN THE EVENT OCCURRED RELATIVE TO SURGERY, HOW THE PROCEDURE WAS FINISHED OR IF THERE WAS A SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12540927 ↗
  • InjuryJul 2, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON LITERATURE REVIEW ¿*MINIMALLY INVASIVE PROCEDURE FOR DIAGNOSIS AND TREATMENT OF VALLECULAR CYSTS IN CHILDREN: REVIEW OF 156 CASES.¿, AFTER SURGERY WITH A "PROCISE LW COBLATOR II WAND" A PATIENT HAD A RECURRENT CYST. THE COMPLICATION WAS TREATED WITH AN ADDITIONAL SURGERY. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Cyst(s)

    MAUDE report 12111542 ↗
  • InjuryJul 2, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON LITERATURE REVIEW ¿SURGICAL MANAGEMENT OF AIRWAY STENOSIS BY RADIOFREQUENCY COLLATION¿; AFTER SURGERY WITH A "PROCISE LW COLLATOR II WAND" A PATIENT HAD RECURRENT TRACHEAL STENOSIS. THE COMPLICATION WAS TREATED WITH AN ADDITIONAL OBLATION SURGERY. THE PATIENT REQUIRED FURTHER TREATMENT WITH A DILATION BALLOON. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Restenosis

    MAUDE report 12110299 ↗
  • InjuryJul 2, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON LITERATURE REVIEW ¿SURGICAL MANAGEMENT OF AIRWAY STENOSIS BY RADIOFREQUENCY COBLATION¿; AFTER SURGERY WITH A "PROCISE LW COBLATOR II WAND" A PATIENT HAD RECURRENT TRACHEAL STENOSIS. THE COMPLICATION HAS BEEN TREATED WITH ADDITIONAL COBLATION SURGERIES. THE PATIENT REQUIRED FURTHER TREATMENT WITH BALLOON DILATION, MITOMYCIN C INJECTION AND STENTING. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Restenosis

    MAUDE report 12110186 ↗
  • InjuryJul 2, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON LITERATURE REVIEW ¿SURGICAL MANAGEMENT OF AIRWAY STENOSIS BY RADIOFREQUENCY COBLATION¿; AFTER SURGERY WITH A "PROCISE LW COBLATOR II WAND" A PATIENT HAD RECURRENT TRACHEAL STENOSIS. THE COMPLICATION HAS BEEN TREATED WITH COBLATION SURGERIES. THE PATIENT REQUIRED FURTHER TREATMENT WITH A DILATION BALLOON. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Restenosis

    MAUDE report 12110125 ↗
  • InjuryJul 2, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON LITERATURE REVIEW ¿SURGICAL MANAGEMENT OF AIRWAY STENOSIS BY RADIOFREQUENCY COBLATION¿; AFTER SURGERY WITH A "PROCISE LW COBLATOR II WAND" A PATIENT HAD RECURRENT SUBGLOTTIC STENOSIS. THE COMPLICATION HAS BEEN TREATED WITH 3 COBLATION SURGERIES. THE PATIENT REQUIRED FURTHER TREATMENT WITH A DILATION BALLOON. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Restenosis

    MAUDE report 12110115 ↗
  • InjuryJul 2, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON LITERATURE REVIEW ¿*MINIMALLY INVASIVE PROCEDURE FOR DIAGNOSIS AND TREATMENT OF VALLECULAR CYSTS IN CHILDREN: REVIEW OF 156 CASES.¿, AFTER SURGERY WITH A "PROCISE LW COBLATOR II WAND" A PATIENT HAD A RECURRENT CYST. THE COMPLICATION WAS TREATED WITH AN ADDITIONAL SURGERY. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Cyst(s)

    MAUDE report 12109101 ↗
  • InjuryJul 1, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON LITERATURE REVIEW ¿LOW-TEMPERATURE RADIOFREQUENCY COBLATION REDUCES TREATMENT INTERVAL AND POST-OPERATIVE PAIN OF LARYNGOTRACHEAL RECURRENT RESPIRATORY PAPILLOMATOSIS¿, AN UNSPECIFIED NUMBER OF PATIENTS WERE RE-OPERATED AFTER THE PROCEDURE IN WHICH A PROCISE LW WAS USED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 12104941 ↗
  • InjuryJul 1, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON LITERATURE REVIEW ¿LOW-TEMPERATURE RADIOFREQUENCY COBLATION REDUCES TREATMENT INTERVAL AND POST-OPERATIVE PAIN OF LARYNGOTRACHEAL RECURRENT RESPIRATORY PAPILLOMATOSIS.¿, THREE PATIENTS DEVELOPED VOCAL CORD ADHESIONS AFTER PROCEDURES IN WHICH A PROCISE LW WAS USED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Adhesion(s)

    MAUDE report 12104869 ↗
  • InjuryJul 1, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON LITERATURE REVIEW ¿LOW-TEMPERATURE RADIOFREQUENCY COBLATION REDUCES TREATMENT INTERVAL AND POST-OPERATIVE PAIN OF LARYNGOTRACHEAL RECURRENT RESPIRATORY PAPILLOMATOSIS.¿, ONE PATIENT PRESENTED WITH ASPHYXIA AFTER THE PROCEDURE IN WHICH A PROCISE LW WAS USED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Respiratory Insufficiency

    MAUDE report 12104868 ↗
  • InjuryFeb 18, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING THE RESECTION PROCEDURE OF THE LARYNX LESION USING THE PROCISE LW WAND, AN LW RADIOFREQUENCY TIP WAS USED, AFTER 5 MIN OF USE THE ELECTRODE BROKE OFF AND THE TIP STOPPED WORKING, DID NOT COAGULATE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DELAY OF 30 MINUTES OR LESS USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11345650 ↗
  • InjuryDec 23, 2020Material Separation; Connection Problem; Material Separation; Connection Problem

    IT WAS REPORTED THAT DURING A PALATINE TONSIL RESECTION PROCEDURE, WHEN THE TISSUE RESECTION WAS STARTED THERE WAS SOME BLEEDING BUT THE PROCISE WAS NOT COAGULATING AND IT WAS FOUND THAT THE ELECTRODE WAS INCOMPLETE DUE TO THE INCREASED LEVEL OF BLEEDING IN THE PATIENT. THE ELECTRODE WAS WORN AND FELT INSIDE THE PATIENT'S ANATOMY. AT THE BEGINNING IT WAS A NORMAL LEVEL OF BLEEDING BUT DURING THE NOVELTY THE LEVEL OF BLEEDING INCREASED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DELAY GREATER THAN 30MIN, USING A BACK-UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 11068986 ↗
  • MalfunctionOct 30, 2019Material Separation; Material Separation

    IT WAS REPORTED THAT DURING THE SURGERY THE DEVICE BROKE, WHERE THE HEAD IS (THE END OF THE TIP). IT IS UNKNOWN IF THE BROKEN PIECES WERE REMOVED FROM THE PATIENT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO DELAY OR PATIENT INJURIES. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9257896 ↗