Coblation AC2823-01Recalls & reported problems
Smith & Nephew, Inc. · 2.3 MM INTEGRATED CABLE WAND ICW, SHORT BEVEL 35, 35 DEGREE · Class 2
All models of Coblation (62)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00817470004564), received between Nov 3, 2019 and Jul 8, 2025 — classified as 4 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K033584); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Insufficient Device Problem Information2
- Material Separation2
- Material Twisted/Bent2
- Connection Problem1
- Failure to Deliver Energy1
- Material Deformation1
- Peeled/Delaminated1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Known Impact Or Consequence To Patient3
- No Information1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 8, 2025Break; Material Twisted/Bent; Break; Material Twisted/Bent
IT WAS REPORTED THAT, DURING SURGERY, THE 2.3 MM SHORT BEVEL BLADE BROKE. IT IS UNKNOWN IF THERE WAS A BACK-UP DEVICE AVAILABLE OR IF THERE WAS A DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22447701 ↗MalfunctionFeb 21, 2025Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE INSULATION COATING OF THE SHAFT OF THE COBLATION CABLE WAND ICW PEELED OFF AT THE START OF USE. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED WITH AND IF THERE WAS A SMITH AND NEPHEW BACK UP DEVICE AVAILABLE. NO SURGICAL DELAY WAS REPORTED AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21433557 ↗MalfunctionMay 9, 2024Material Deformation; Insufficient Device Problem Information; Material Deformation; Insufficient Device Problem Information
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE SHORT BEVEL WAND WAS NOT WORKING WHEN CONNECTED TO THE CONSOLE. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS A SURGICAL DELAY GREATER THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19279939 ↗MalfunctionJan 18, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE SHORT BEVEL 35 DEGREE WAND TIP BROKE. IT IS IF THERE WAS A BACK-UP DEVICE AVAILABLE OR IF THERE WAS A DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18539504 ↗InjurySep 30, 2021Material Separation; Material Separation
IT WAS REPORTED THAT DURING ANKLE PROCEDURE, THE 2.3MM SHORT BEVEL 35 DEGREE ICW METAL TIPS SEEMED TO HAVE FALLEN OFF DURING SURGERY; (DISINTEGRATING) LEAVING METAL DEBRIS IN THE ANKLE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12558323 ↗InjuryFeb 19, 2021Failure to Deliver Energy; Failure to Deliver Energy
IT WAS REPORTED THAT DURING AN ACL PROCEDURE, THE WAND FAILED TO COAGULATE AND THE TIP DID NOT WORK AT THAT TIME, ALL THE TESTS WERE CARRIED OUT BUT IT DID NOT WORK. PROCEDURE WAS COMPLETED WITH A COMPETITOR DEVICE. NO SIGNIFICANT DELAY AND NO PATIENT COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11347878 ↗MalfunctionFeb 10, 2021Break; Break
IT WAS REPORTED THAT DURING SURGERY, THE TIP OF THE BEVEL WAND DISINTEGRATED CAUSING SMALL PARTS TO BREAK OFF IN THE PATIENT'S JOINT. THE DEVICE WORKED IN THE PRESET SETTING BUT IT WOULD NOT WORK WITH A CONTINUOUS ACTIVATION, ONLY PULSED. ALL FRAGMENTS WERE REMOVED FROM THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11306575 ↗InjurySep 10, 2020Connection Problem; Connection Problem
IT WAS REPORTED THAT, DURING A WRIST ARTHROSCOPY PROCEDURE, WHEN UNCOVERING THE "2.3MM SHORT BEVEL 35° IC WAND", IT WAS EVIDENT THAT THE CONNECTION OF THE RADIOFREQUENCY TIP DID NOT MAKE CONTACT WITH THE CONSOLE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING AN ALTERNATIVE S+N DEVICE, A "SHAVER 2.0". NO PATIENT INJURIES OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Information
MAUDE report 10516476 ↗InjuryJul 29, 2020Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT, DURING WAIST SURGERY, THE WAND WAS UNABLE TO PLUG INTO QUANTUM CONSOLES: IT WAS BROKEN. WHEN THE SURGEON TRIED TO PLUG IN, IT BENT THE PINS INSIDE AND DID NOT MATCH UP WITH CONSOLE. THE PROCEDURE WAS COMPLETED WITH A COMPETITOR DEVICE WITH NO SIGNIFICANT DELAY. NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10341866 ↗MalfunctionNov 18, 2019Material Separation; Material Separation
IT WAS REPORTED THAT DURING AN ANKLE PROCEDURE, THE DEVICE TIP STARTED TO MELT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO DELAY AND NO PATIENT INJURIES. RESULTS OF INVESTIGATION HAVE CONCLUDED THAT THERE WERE MISSING SECTIONS OF EPOXY INSULATION AROUND THE SPACER, AND POTENTIALLY THESE PIECES MIGHT FALL INTO THE PATIENT WHICH MAKES IT A REPORTABLE EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9334923 ↗MalfunctionNov 3, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT DURING THE COURSE OF THE SURGERY THE SPECIALIST ASKED FOR THE RADIOFREQUENCY TO FINISH A REMODELING AFTER SYNOVECTOMY, THE RADIOFREQUENCY WAS PASSED TO THE TABLE WHICH WAS PROPERLY CONNECTED TO THE CONSOLE. THE HANDPIECE WAS ENTERED INTO THE JOINT TO START PERFORMING THE PROCEDURE AND WHEN IT WAS ACTIVATED WITH THE PEDAL CUT THE COAGULATION DID NOT WORK. NO MOVEMENT IN THE TISSUE WAS EVIDENT. HENCE, THE SURGEON PROCEEDED TO DISCONNECT AND RECONNECT THE TIP TO SEE IF IT WORKED, BUT IT DID NOT WORK. THE PROCEDURE WAS FINISHED USING A J&J VPR. IT IS UNKNOWN IF THERE WAS A DELAY AND NO PATIENT INJURY WAS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9270654 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K033584. One submission often covers many models, so check the product and lot details in each record.
- Z-2309-2008 · Apr 9, 2008 · TerminatedArthroCare ArthroWand Covator with Integrated Cable Wand, Catalog number AC4340-01. The product is a medical device for human use in electrosurgical cutting and coagulation access
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page