Skip to content

PenEvac1 ICM-000-0572Recalls & reported problems

I.C. Medical, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00817688022527), received between Jul 7, 2026 and Jul 7, 2026 — classified as 1 malfunction.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 7, 2026

    DURING SURGICAL PROCEDURE IN OR, SURGEON PLACED SUCTION BOVIE BACK IN HOLSTER. THE BOVIE PENCIL TUBING RECOILED, CAUSING IT TO COME OUT. DRAPES REMOVED. PT ASSESSED. CODE RED CALLED. FIRE ALARM PULLED. AFTER EVENT CLEARED, SURGICAL PROCEDURE PROCEEDED. DRAPE & BOVIE SAVED. IT HAS BEEN IDENTIFIED THAT AFTER THE TUBING SURROUNDING THE HOT CAUTERY TIP WAS PLACED INSIDE THE SAFETY HOLSTER IS RECOILED BACK AND OUT OF THE HOLSTER LANDING ON THE DRAPE CAUSING A BURN. ALTHOUGH THIS WAS QUICKLY IDENTIFIED, THE PATIENT WAS NEGATIVELY IMPACTED AND PLACED AT INCREASED RISK FOR SSI [SURGICAL SITE INFECTION] AS THE DRAPES HAD TO BE REMOVED AND REPLACED OVER A SURGICALLY OPENED JOINT.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25734877 ↗

Other Electrosurgical, Cutting & Coagulation & Accessories devices

All Electrosurgical, Cutting & Coagulation & Accessories devices →