Skip to content

ThermiRF+Recalls & reported problems

Celling Biosciences · Radiofrequency Temp Controlled Generator · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00817771029044), received between Feb 1, 2024 and Feb 14, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Off-Label Use” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Off-Label Use1
  • Use of Device Problem1

Patient problems

  • Burning Sensation2
  • Discomfort1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 14, 2024Off-Label Use; Use of Device Problem; Off-Label Use; Use of Device Problem

    ON 16JAN2024, CUSTOMER SUPPORT WAS NOTIFIED BY A SALES REPRESENTATIVE OF AN ADVERSE EVENT. CHETCO MEDICAL (ACCT #: (B)(4)) REPORTED A PATIENT THEY PERFORMED A THERMIVA TREATMENT HAD BURNS INSIDE HER VAGINA. THE THERMIVA TREATMENT WAS PERFORMED IN (B)(6) 2023. THE PATIENT REPORTED SHE FELT DISCOMFORT AFTER THE TREATMENT AND VISITED HER OBGYN IN (B)(6) 2023. THE PATIENT'S OBGYN INFORMED HER SHE HAD BURNS INSIDE HER VAGINA. THE CUSTOMER MENTIONED THEY WERE PERFORMING AGGRESSIVE TREATMENT DUE TO THE PATIENT'S URINARY INCONTINENCE. THE PATIENT ONLY HAD ONE THERMIVA TREATMENT IN (B)(6) WHEN IN (B)(6) HER OBGYN NOTIFIED HER OF THE BURN. THE CUSTOMER REPORTED IN (B)(6) THEY ARE PERFORMING PRP TREATMENTS FOR THE BURN ON THE PATIENT. THE CUSTOMER HAS NOT REPORTED ANY OTHER ISSUES WITH THE THERMIRF+ GENERATOR (SN #: (B)(6)). THE UNIT WAS REQUESTED FOR EVALUATION TO ENSURE IT IS FUNCTIONING AS NORMAL.

    Patient problems: Burning Sensation

    MAUDE report 18707983 ↗
  • InjuryFeb 1, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT NOTICED A BURN ON RIGHT THIGH WHERE GROUNDING PAD WAS PLACED DURING HER THERMIVA TREATMENT WHEN SHE ARRIVED HOME. OFFICE STAFF REPORTED THAT DURING THE TREATMENT PATIENT MENTIONED THAT THERE WAS A SENSATION OF HEAT WHERE THE GROUNDING PAD WAS PLACED, SO THE NURSE RE-APPLIED THE GROUNDING AND THE PATIENT REPORTED THAT THEY FELT FINE SO THE TREATMENT PROCEEDED.

    Patient problems: Burning Sensation; Discomfort

    MAUDE report 18627617 ↗

Other Electrosurgical, Cutting & Coagulation & Accessories devices

All Electrosurgical, Cutting & Coagulation & Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page