Skip to content

ArvatiRecalls & reported problems

Celling Biosciences · Radiofrequency Temp Controlled Generator · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00817771029068), received between Nov 19, 2023 and Apr 11, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Manufacturing, Packaging or Shipping Problem1

Patient problems

  • Burn(s)2
  • Electric Shock1
  • Scar Tissue1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2024, DR. (B)(6) PLASTIC SURGERY (CUSTOMER ID: (B)(6)) REPORTED TO AN AREA SALE MANAGER THEY WERE EXPERIENCING ISSUES WITH BURNING WHEN PERFORMING THERMITIGHT PROCEDURE. THE PROVIDER WHO PERFORMED THE PROCEDURES IS NO LONGER AT THEIR PRACTICE. THERE WERE FOUR PATIENTS THAT WERE BURNED. OF THE FOUR PATIENTS, TWO HAVE PERMANENT SCARRING. PER THE CUSTOMER, ONE THEY HAVE BEEN TREATING FOR YEARS FREE OF CHARGE. THE TWO MOST SEVER CASES ARE FROM A SINGLE THERMITIGHT PROCEDURE TO THE NECK AND THIGH AREA. THESE BURNED OCCURRED ON THEIR ARVATI GENERATOR (SN: (B)(6). IN (B)(6) 2023 THE CUSTOMER SENT IN THEIR GENERATOR FOR REPAIR DUE TO A FAULT 150. A FAULT 150 IS A BATTERY ISSUE, AND THE UNIT WILL NOT FUNCTION. ALL THE BURNS HAPPENED PRIOR TO RETURNING THE GENERATOR TO OUR LOCATION FOR REPAIR. AT THE TIME OF THE FAULT 150, AND RETURN OF THE GENERATOR, THERE WAS NO MENTION OF THE ADVERSE EVENTS OR BURNS. A LOANER CERTIFIED THERMI RF+ (SN #: (B)(6) WAS SENT ON 03FEB2023. THE CUSTOMER ARVATI GENERATOR (SN #: (B)(6) WAS UNABLE TO BE EVALUATED DUE TO THE FAUL AND THE CUSTOMER DECIDED TO KEEP THE LOANER CERTIFIED THERMIRF+ GENERATOR (SN #: (B)(6). AT THIS TIME, THIS IS ALL THE INFORMATIO

    Patient problems: Burn(s); Scar Tissue

    MAUDE report 19087545 ↗
  • InjuryNov 19, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    PHYSICIAN REPORTED TO SALES REPRESENTATIVE ON (B)(6) 2021 THAT DURING THE THERMISMOOTH PROCEDURE THE PATIENT COMPLAINED OF A SPARKING AND BURNING SENSATION. PHYSICAN REPORTED THAT HE HAD PICTURES BUT THEY WERE NEVER OBTAINED DURING THE INITIAL CALL BACK ON (B)(6) 2021. PHYSICIAN DID REPORT THAT DURING SETUP HE HAD TO PUSH THE ELECTRODE INTO THE GENERATOR VERY HARD. PHYSICIAN REQUESTED THAT THE GENERATOR AND ELECTRODE BE EVALUATED. FOLLOW UP CALL ON (B)(6) 2021 BY THEN QUALITY MANAGER FOR SPINESMITH, COLLECTED THAT THERE WAS A MAXIMUM SKIN TEMP OF 45.5C FROM THE FLIR CAMERA, THE TREATMENT LASTED 10 MINUTES, PHYSICIAN REPORTED NO PROBLEMS DURING THE TREATMENT. CUSTOMER ACCOUNT NOTES THAT A 2ND TREATMENT WAS PERFORMED THE PATIENT OVER A BIGGER AREA (CROW'S FEET AND CHEEKBONES), PATIENT COMPLAINED AT 2ND TREATMENT AS WELL. ON (B)(6) 2023 ARVATI SN# (B)(6) AND ELECTRODES RFE-10-D-G2 SN# (B)(6) AND RFE-15-D-G2 SN# (B)(6) WERE SENT IN FOR INVESTIGATION. INSPECTION AND TESTING OF CUSTOMER OWNED GENERATOR SHOWED THAT ACTIVE RF PERFORMED FOR 1HR WITH SET TEMPERATURE OF 90C, IMPEDANCE LEVELS, TEMPERATURE RESPONSE AND SAFETY TESTING ALL TESTINGED WITHIN EXPECTED PARAMETERS. THERE WERE NOT ANOM

    Patient problems: Burn(s)

    MAUDE report 18167259 ↗
  • InjuryNov 19, 2023Manufacturing, Packaging or Shipping Problem; Patient Device Interaction Problem; Manufacturing, Packaging or Shipping Problem; Patient Device Interaction Problem

    PHYSICIAN REPORTED THAT HER LAST TWO PATIENTS REPORTED MULTIPLE SHOCK SENSATION DURING THEIR THERMIVA TREATMENTS. INVESTIGATION NOTED THAT THE PHYSICIAN WAS USING RFDE-VA-G2 LOT# L18E027 AND GROUNDING PAD CAT: #2130ACLOT# 20160720 EXPIRED AS OF 07/19/2018. AS A COURTESY, THE PHYSICIAN WAS SENT REPLACEMENT ELECTRODES AND GROUNDING PADS AND INFORMED THAT USE OF EXPIRED GROUNDING PADS IS NOT ACCEPTABLE. COMPLAINT WAS CLOSED 07/27/21. DURING 2023 FDA AUDIT, INSPECTOR INDICATED THAT THE COMPLAINT REQUIRED RE-ASSESSMENT. COMPLAINT WAS RE-OPENED ON 08/04/23. ATTEMPTS HAVE BEEN MADE ON 08/08/23 (VOICEMAIL LEFT), 08/18/23 (EMAIL) AND 08/22/23 (EMAIL). NO RESPONSE AS OF 11/19/23. REVIEW OF THE CUSTOMER WEBSITE NOTES THAT THE OFFICE IS STILL PROVIDING THERMIVA PROCEDURES. ADDITIONAL FOLLOW UP EMAIL REQUESTING STATUS OT THERMIVA PROCEDURE SENT TO PHYSICIAN ON 11/16/23 BY CURRENT SPINESMITH QUALITY MANAGER.

    Patient problems: Electric Shock

    MAUDE report 18167243 ↗

Other Electrosurgical, Cutting & Coagulation & Accessories devices

All Electrosurgical, Cutting & Coagulation & Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page