Skip to content

(Certified) ThermiRF+Recalls & reported problems

Celling Biosciences · Upgraded Arvati Radiofrequency Temp Controlled Generator · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00817771029181), received between Nov 19, 2023 and Nov 19, 2023 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Apparent Adverse Event1

Patient problems

  • Discomfort1
  • Hyperresponsive to stimuli1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 19, 2023No Apparent Adverse Event; No Apparent Adverse Event

    SPINESMITH REPRESENTATIVE REPORTED THAT THE ACCOUNT INFORMED THEM THAT THEIR RF+ THERMISMOOTH HANDPIECE IS HEATING UP TOO QUICKLY AND CAUSING (SPARKING AND ARCHING) WHEN IN CONTACT WITH THE PATIENT AND THEY ARE UNABLE TO DO PROCEDURES. SPINESMITH SENIOR CLINICAL SPECIALIST CONTACTED THE CUSTOMER FOR MORE INFORMATION. SPECIALIST REPORTED THAT AFTER SPEAKING WITH THE PRACTICE, SHE EXPLAINED THAT DURING THE PROCEDURE, THE PROBE TO SKIN CONTACT MAY BE DISRUPTED (FOR EXAMPLE IF DURING THE PROCEDURE THE PROBE WAS LIFTED FORM THE SKIN SURFACE OR IF THERE IS HAIR/DEBRIS IN THE WAY OF THE CONTACT SURFACE OF THE ELECTRODE), A SENSATION CAN OCCUR THAT FEELS LIKE PULLING/TIGHTENING ON THE SKIN OR A SHOCK. SPECIALIST DID REQUEST THE RETURN OF THE ELECTRODE USED FOR EVALUATION. CONTINUED CONVERSATION WITH OFFICE STAFF VERIFIED THAT WHILE PERFORMING THE PROCEDURE WITH THE THERMISMOOTH HANDPIECE, THERE WAS NO VISUAL "SPARKING" OR "ARCHING" OBSERVED RATHER THAT THE PATIENT DESCRIBED DISCOMFORT AS A "PULLING" OR "TIGHT" SENSATION WAS OCCURING. NO INJURIES WERE OBSERVED BY THE PHYSICIAN AND THEY HAVE SINCE HAD NO ISSUES. ELECTRODE WILL BE RETURNED FOR FINAL EVALUATION.

    Patient problems: Discomfort; Hyperresponsive to stimuli

    MAUDE report 18167239 ↗

Other Electrosurgical, Cutting & Coagulation & Accessories devices

All Electrosurgical, Cutting & Coagulation & Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page