Gyrus ACMI 777000Recalls & reported problems

Gyrus ACMI, LLC · G400 WORKSTATION · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Gyrus ACMI electrosurgical workstation is the generator that powers surgical cutting and coagulation tools. Many of its 103 reports describe low or missing output and circuit boards that broke or short-circuited. Most reports describe no patient involvement. The reports mention no deaths and 1 injury. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
103
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 103 adverse event reports that list this exact device (DI 00821925009004), received between Dec 17, 2020 and Aug 11, 2026 — classified as 1 injury, 102 malfunctions.

The most frequently reported specific device problem is “Output Problem” (38 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 5 reportsMar 2024: 2 reportsApr 2024: 1 reportsMay 2024: 6 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 3 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output Problem38
  • Output below Specifications27
  • Electrical /Electronic Property Problem8
  • Circuit Failure7
  • Defective Component7
  • No Device Output7
  • Break5
  • Power Problem5

Patient problems

  • No Clinical Signs, Symptoms or Conditions95
  • No Patient Involvement5
  • No Consequences Or Impact To Patient3
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 25 most recent of 103 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 11, 2026Output below Specifications

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50211498 ↗
  • MalfunctionJul 1, 2026Break

    THE CUSTOMER RETURNED THE ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25678140 ↗
  • MalfunctionJun 6, 2026Circuit Failure

    THE ELECTROSURGICAL, CUTTING AND COAGULATION AND ACCESSORIES WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF ASSET CHECK FAULTY. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, THE AUXILIARY CIRCUIT BOARD WAS FOUND TO HAVE LOW POWER OUTPUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25418921 ↗
  • MalfunctionApr 3, 2026Break; Degraded; Electrical /Electronic Property Problem

    IT WAS OBSERVED THAT DURING DEVICE EVALUATION, THE G400 GENERATOR EXHIBITED A BROKEN TRANSISTOR 5 COMPONENT ON THE PEAK RADIO FREQUENCY BOARD, AND METAL CAUSED A SHORT CIRCUIT TO THE CIRCUIT BOARD. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24778633 ↗
  • MalfunctionApr 2, 2026Degraded; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ELECTROSURGICAL CUTTING AND COAGULATION ACCESSORIES PRODUCED OUTPUT OUTSIDE THE ACCEPTABLE RANGE, AND A FAULT WAS DETECTED ON THE AUXILIARY BOARD. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24772582 ↗
  • MalfunctionMar 31, 2026Circuit Failure

    IT WAS OBSERVED THAT THE POWER BOARD AND THE PLASMA BLEND DESATURATION ARE LOWER THAN THE LIMIT AND CAN'T BE ADJUSTED DUE TO FAULTY PLASMA BLEND BOARD. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24749620 ↗
  • MalfunctionMar 30, 2026Power Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ELECTROSURGICAL GENERATOR HAD NO OUTPUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24728469 ↗
  • MalfunctionNov 25, 2025Output Problem; Output Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ELECTROSURGICAL, CUTTING AND COAGULATION AND ACCESSORIES EXHIBITED A FAILURE IN PLASMA BLEND DESICCATION OUTPUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23644503 ↗
  • MalfunctionJul 29, 2025Failure to Power Up; Failure to Power Up

    IT WAS REPORTED THAT THE GENERATOR EXPERIENCED A NO POWER OUTPUT PROBLEM DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22650194 ↗
  • MalfunctionApr 24, 2025Output below Specifications; Output below Specifications

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21905100 ↗
  • MalfunctionFeb 28, 2025Failure to Power Up; Failure to Power Up

    IT WAS REPORTED THAT THE GENERATOR HAD BIPOLAR NOT ENGAGING/POWERING ON. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21487597 ↗
  • MalfunctionFeb 24, 2025Power Problem; Power Problem

    IT WAS OBSERVED THAT DURING DEVICE EVALUATION, THAT THE GENERATOR HAD A FAULTY POWER SUPPLY UNIT. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21453955 ↗
  • MalfunctionFeb 20, 2025Insufficient Heating; Insufficient Heating

    THE CUSTOMER REPORTED THAT AN OLYMPUS G400 GENERATOR WAS NOT PRODUCING ENOUGH HEAT IN THE MIDDLE OF A LAPAROSCOPIC HYSTERECTOMY. ACCORDING TO THE INITIAL REPORTER, THE DEVICE WAS NOT GETTING HOT ENOUGH TO CAUTERIZING PROPERLY. REPORTEDLY, THE PROCEDURE WAS COMPLETED AFTER A 3-5 MINUTE DELAY BY USING ANOTHER DEVICE. THERE WAS NO REPORT OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21421754 ↗
  • MalfunctionNov 1, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE G400 GENERATOR EXHIBITED A FAILED OUTPUT DUE TO A FAULTY PLASMA BLEND BOARD. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20587046 ↗
  • MalfunctionSep 25, 2024Loss of Power; Loss of Power

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE G400 GENERATOR, GYRUS EXHIBITED A FAULTY POWER SUPPLY. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20301332 ↗
  • MalfunctionSep 23, 2024Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ELECTROSURGICAL GENERATOR EXHIBITED A FAULTY POWER SUPPLY UNIT (PSU) BOARD. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20287875 ↗
  • MalfunctionAug 14, 2024No Device Output; No Device Output

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GENERATOR HAD OUTPUT ISSUES. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19988115 ↗
  • MalfunctionJul 12, 2024Output Problem; Output Problem

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GENERATOR EXHIBITED OUTPUT ISSUES DUE TO A FAULTY PRINTED CIRCUIT BOARD. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19731495 ↗
  • MalfunctionJun 21, 2024Circuit Failure; Operating System Version or Upgrade Problem; Circuit Failure; Operating System Version or Upgrade Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ELECTROSURGICAL GENERATOR EXHIBITED A FAULTY POWER SUPPLY UNIT (PSU), ISSUE WITH AN OLD VERSION PLASMA BLEND BOARD, AND FAULTY AUXILIARY (AUX) BOARD. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19584225 ↗
  • MalfunctionJun 10, 2024No Device Output; No Device Output

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GENERATOR EXHIBITED A FAULTY AUX BOARD. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19504718 ↗
  • MalfunctionMay 21, 2024Break; Break

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GENERATOR EXHIBITED A FAULTY AUXILIARY BOARD FOUND WHEN THE NUMBER NINE SWITCH TEST FAILED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19366423 ↗
  • MalfunctionMay 21, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE G400 GENERATOR, GYRUS EXHIBITED ENERGY OUTPUT ISSUES. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19358089 ↗
  • MalfunctionMay 16, 2024Break; Break

    IT WAS REPORTED THE ELECTROSURGICAL KNIFE INSULATION TIP DROPPED; INTO THE PATIENT¿S STOMACH AND WAS NOT RETRIEVED. THE ISSUE OCCURRED DURING A THERAPEUTIC GASTRIC ESD. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19336703 ↗
  • MalfunctionMay 7, 2024Defective Component; Defective Component

    IT WAS REPORTED THAT THE ISSUE OCCURRED DURING A THERAPEUTIC LAPAROSCOPIC ASSISTED VAGINAL HYSTERECTOMY. THE PROCEDURE WAS DELAYED BY TEN MINUTES TO ALLOW THE DEVICE TO BE SWITCHED. THE PATIENT WAS UNDER GENERAL ANESTHESIA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19259120 ↗
  • MalfunctionMay 6, 2024Defective Component; Defective Component

    IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE GENERATOR HAD A FAULTY AUXILIARY BOARD. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19256463 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page