DIEGO ELITE MDFS100Recalls & reported problems
Gyrus ACMI, LLC · DIEGO ELITE FOOTSWITCH · Class 2
All models of DIEGO ELITE (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00821925027787), received between Jan 15, 2021 and Nov 20, 2023 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (2 reports).
12 other recalls cover devices cleared under the same FDA submission (K123429); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Problem2
- Blocked Connection1
- Break1
- Connection Problem1
- Defective Device1
- Electrical /Electronic Property Problem1
- Intermittent Continuity1
- Unexpected Shutdown1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Hemorrhage/Blood Loss/Bleeding1
- No Patient Involvement1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 20, 2023Connection Problem; Connection Problem
THE CUSTOMER REPORTED TO OLYMPUS, THE DIEGO ELITE MULTIDEBRIDER FOOTSWITCH WAS NOT RESPONDING. THE ISSUE WAS FOUND DURING PREPARATION FOR USE. THE CUSTOMER HAD EXPERIENCED AN ISSUE A FEW DAYS BEFOREHAND AND WAS WIGGLING THE WIRE BUT NOW THERE WAS NO RESPONSE AT ALL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18177601 ↗InjuryJul 26, 2023Blocked Connection; Blocked Connection
OLYMPUS FURTHER RECEIVED INFORMATION THAT THE INTENDED PROCEDURE WAS FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS). THE BLEEDING WAS RESOLVED BY CAUTERY. THE DEVICE WAS INSPECTED BEFORE USE WITH NO VISIBLE ISSUES. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 17396221 ↗MalfunctionMay 31, 2023Break; Mechanical Problem; Break; Mechanical Problem
IT WAS REPORTED THE FOOT PEDAL WENT ALL THE WAY DOWN. PER THE REPORT, THE SPRING FOR THE SHAFT SUSPECTED TO BE BROKEN. THE ISSUE FOUND AT PREPARATION FOR USE. THERE WAS NO PATIENT INVOLVEMENT ON THIS REPORTED EVENT. NO USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17028809 ↗InjuryFeb 23, 2023Intermittent Continuity; Unexpected Shutdown; Intermittent Continuity; Unexpected Shutdown
A USER FACILITY REPORTED TO OLYMPUS THAT WHILE USING A MULTIDEBRIDER FOOTSWITCH DURING AN UNSPECIFIED PROCEDURE WHILE REMOVING POLYPS, THE FOOT SWITCH WAS INTERMITTENT AND THEN LOST ALL POWER. THE PHYSICIAN WAS ABLE TO USE A MANUAL INSTRUMENT TO REMOVE THE POLYP AND THERE WAS NO IMPACT TO THE PATIENT'S PROCEDURE OUTCOME. HOWEVER, THE FOOT SWITCH ISSUE CAUSED A PROCEDURE DELAY OF AROUND 30 MINUTES WHILE ATTEMPTING TO REMEDY THE SITUATION. PATIENT IDENTIFIER (B)(6) IS FOR THE MULTIDEBRIDER POWER CONSOLE. PATIENT IDENTIFIER (B)(6) IS FOR THE MULTIDEBRIDER FOOTSWITCH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16432617 ↗MalfunctionAug 2, 2022Electrical /Electronic Property Problem; Mechanical Problem; Electrical /Electronic Property Problem; Mechanical Problem
IT WAS REPORTED THAT IN THE MIDDLE OF A FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS) PROCEDURE IN NOSE, CUSTOMER TRIED TO STOP THE ROTATION OF DIEGO'S TERMINAL BY RELEASING THE FOOTSWITCH. HOWEVER; THE BLADE KEPT ON ROTATING AT LOWER REVOLUTIONS EVEN THOUGH THE FOOTSWITCH WAS NOT PRESSED DOWN. IT DID NOT CAUSE ANY DAMAGE TO PATIENT AS IT WAS REMOVED IN TIME. THE DEVICE WAS REPLACED. THE INTENDED PROCEDURE WAS COMPLETED WITH A BACKUP FOOTSWITCH. THERE WAS NO PATIENT HARM, NO USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15143724 ↗MalfunctionJan 15, 2021Defective Device
IT WAS REPORTED THAT DURING AN ASSET RETURN INSPECTION THE DEVICE WAS FOUND WITH FAULTY SWITCH. THERE WAS NO PATIENT INVOLVEMENT ON THIS EVENT. NO USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 11181359 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123429. One submission often covers many models, so check the product and lot details in each record.
- Z-1186-2019 · Mar 8, 2019 · TerminatedDiego Elite Blades Monopolar, 4mm Straight, TIA Product Code:MB4000SS
- Z-1178-2019 · Mar 8, 2019 · TerminatedOlympus Diego Elite Console MDCONS 100
- Z-1182-2019 · Mar 8, 2019 · TerminatedDiego Elite Turbinate Blades Bipolar Blade, 4mm curved, Product Number:BB4040SS
- Z-1183-2019 · Mar 8, 2019 · TerminatedDiego Elite Turbinate Blades Bipolar Blade, 4mm curved, Product Number: BB4040SC
- Z-1184-2019 · Mar 8, 2019 · TerminatedDiego Elite Turbinate Blades Bipolar Blade, 4mm curved, Product Number: BB4040XS
- Z-1181-2019 · Mar 8, 2019 · TerminatedDiego Elite Turbinate Blades Bipolar Blade, 4mm straight, Part No: BB4000SC
- Z-1187-2019 · Mar 8, 2019 · TerminatedDiego Elite Malleable, 4mm Standard Monopolar Blades Product Code:MM4000SS
- Z-1185-2019 · Mar 8, 2019 · TerminatedDiego Elite Turbinate Blades Bipolar Blade, 4mm curved, Product Number: BB4040XC
- Z-1180-2019 · Mar 8, 2019 · TerminatedDiego Elite Turbinate Blades Bipolar Blade, 4mm straight, Part Number: BB4000SS
- Z-1179-2019 · Mar 8, 2019 · TerminatedDiego Elite Turbinate Blades:Bipolar Blade, 2mm straight, standard, type A Part No BB2000SA
- Z-1752-2015 · May 8, 2015 · TerminatedGyrus ACMI Diego Elite Tubeset; intended for cutting, coagulation, drilling, debriding, and removal of bone, and soft and hard tissue in general ENT, Sinus/Rhinology, Nasopharyngea
- Z-1753-2015 · May 8, 2015 · TerminatedGyrus ACMI Diego Elite Tubeset, intended for cutting, coagulation, drilling, debriding, and removal of bone, and soft and hard tissue in general ENT, Sinus/Rhinology, Nasopharyngea
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page