UroPass 61154BXRecalls & reported problems
Gyrus ACMI, LLC · UroPass Ureteral Access Sheath 11/13 Fr (3.7/4.3 mm) x 54 cm (5/PK) · Class 2
All models of UroPass (15)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00821925035386), received between Dec 8, 2022 and Dec 8, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “Material Separation” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K051593); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Sep 11, 2024Class IIOpen, Classified
Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures.
Z-0364-2025 · Root cause: Storage · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryDec 8, 2022Material Separation; Material Separation
THE CUSTOMER REPORTS DURING A CYSTO-URETEROSCOPY PROCEDURE USING A UROPASS URETERAL ACCESS SHEATH DILATOR, THE SHEATH TIP BROKE OFF INSIDE THE PATIENT¿S BLADDER WHEN THE PHYSICIAN WAS ATTEMPTING TO ADVANCE IT OVER THE WIRE INTO THE URETER. THE DEVICE FRAGMENTS WERE RETRIEVED. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS AS A RESULT OF THIS OCCURRENCE. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "STABLE." THERE WERE NO PROCEDURAL OR ANATOMICAL CHALLENGES THAT COULD HAVE CONTRIBUTED TO THE DEVICE BREAKING. THE CUSTOMER EXAMINED A LIKE DEVICE STILL SEALED IN THE PACKAGE. AFTER TRYING TO WIGGLE THE DEVICE INSIDE THE PACKAGE, THE DEVICE SNAPPED INSIDE THE PACKAGE. ADDITIONAL COMMENT BY PHYSICIAN "ADDITIONAL DEVICES WERE CHECKED WHILE IN THE PACKAGING/UNOPENED AND SHOWED SAME DEFECT." ONE IMAGE WAS PROVIDED SHOWING A DEVICE BROKEN IN THE PACKAGE. CLARIFYING INFORMATION HAS BEEN REQUESTED, AT THIS TIME, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED. THE CUSTOMER STORES THE DEVICES IN STERILE CORE HANGING UP THE WAY THEY ARE DESIGNED. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS THE PATIENT EVENT CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS THE DEVICE BROKEN IN THE PACKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15941597 ↗
FDA clearance and classification
Cleared through 510(k) K051593 on Sep 2, 2005 (decision: Substantially Equivalent). Submitted by Acmi Corporation as “Uropass Ureteral Access Sheath, Model 61224”. The FDA review took 78 days. FDA record ↗
The FDA regulates product code KNY (accessories, catheter, g-u) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K051593. One submission often covers many models, so check the product and lot details in each record.
- Z-2401-2023 · May 19, 2023 · Open, ClassifiedUroPass Ureteral Access Sheaths, 5 pieces/box
Common questions about the UroPass 61154BX
Has the UroPass 61154BX been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0364-2025 (Class II), began on Sep 11, 2024: Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the UroPass 61154BX?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Dec 8, 2022. That report lists device problem “Material Separation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the UroPass 61154BX?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the UroPass 61154BX FDA approved?
It was cleared through 510(k) K051593 on Sep 2, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the UroPass 61154BX, and is it still sold?
The FDA's device database lists Gyrus ACMI, LLC as the company responsible for it (brand UroPass, model 61154BX). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ACCESSORIES, CATHETER, G-U devices
- UroPass 61238BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall10 reports
- UroPass 61138BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall5 reports
- UroPass 61024BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall5 reports
- UroPass 61224BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61146BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61254BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61046BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61054BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page