UroPass 61338BXRecalls & reported problems
Gyrus ACMI, LLC · UroPass Ureteral Access Sheath 13/15 Fr (4.3/5 mm) x 38 cm (5/PK) · Class 2
All models of UroPass (15)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00821925035447), received between Dec 22, 2021 and Dec 28, 2022 — classified as 2 injuries.
The most frequently reported specific device problem is “Material Separation” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K051593); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Sep 11, 2024Class IIOpen, Classified
Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures.
Z-0364-2025 · Root cause: Storage · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Separation2
Patient problems
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 28, 2022Material Separation; Material Separation
C22475559 MODEL# 61154BX / LOT # 09J1800111
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16061958 ↗InjuryDec 22, 2021Material Separation; Material Separation
THE CUSTOMER REPORTS DURING AN X-RAY FOLLOWING A TRANSURETHRAL URINARY CALCULUS REMOVAL PROCEDURE USING A UROPASS SHEATH, A FOREIGN BODY WAS FOUND TO REMAIN IN THE URINARY TRACT. A SECOND FLEXIBLE TRANSURETHRAL URETEROLITHOTRIPSY (F-TUL) WAS PERFORMED AND THE FOREIGN BODY WAS COLLECTED. THE FOREIGN BODY WAS ABOUT 10 MM IN LENGTH, WHITE, AND HAD CRACKS AROUND 4 MM FROM THE TIP. THE STUMP HAD BEEN TORN DIAGONALLY. THE CUSTOMER SPECULATES IT IS THE TIP OF THE UROPASS SHEATH THAT MAY HAVE BEEN BROKEN WHEN THE STONES WERE COLLECTED USING THE BASKET FORCEPS. THERE WERE NO ADDITIONAL CONSEQUENCES TO THE PATIENT REPORTED. ADDITIONAL DETAILS REGARDING THE PATIENT AND EVENT HAVE BEEN REQUESTED. AT THIS TIME, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 13065954 ↗
FDA clearance and classification
Cleared through 510(k) K051593 on Sep 2, 2005 (decision: Substantially Equivalent). Submitted by Acmi Corporation as “Uropass Ureteral Access Sheath, Model 61224”. The FDA review took 78 days. FDA record ↗
The FDA regulates product code KNY (accessories, catheter, g-u) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K051593. One submission often covers many models, so check the product and lot details in each record.
- Z-2401-2023 · May 19, 2023 · Open, ClassifiedUroPass Ureteral Access Sheaths, 5 pieces/box
Common questions about the UroPass 61338BX
Has the UroPass 61338BX been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0364-2025 (Class II), began on Sep 11, 2024: Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the UroPass 61338BX?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Dec 22, 2021 and Dec 28, 2022. In those reports, the most common specific device problem is “Material Separation” (2); the most common patient problem is “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the UroPass 61338BX?
None of the reports are classified as deaths, 2 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the UroPass 61338BX FDA approved?
It was cleared through 510(k) K051593 on Sep 2, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the UroPass 61338BX, and is it still sold?
The FDA's device database lists Gyrus ACMI, LLC as the company responsible for it (brand UroPass, model 61338BX). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ACCESSORIES, CATHETER, G-U devices
- UroPass 61238BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall10 reports
- UroPass 61138BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall5 reports
- UroPass 61024BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall5 reports
- UroPass 61224BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61146BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61254BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61046BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61054BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page