UroPass 61346BXRecalls & reported problems
Gyrus ACMI, LLC · UroPass Ureteral Access Sheath 13/15 Fr (4.3/5 mm) x 46 cm (5/PK) · Class 2
All models of UroPass (15)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00821925035454), received between Nov 10, 2021 and Nov 10, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K051593); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Sep 11, 2024Class IIOpen, Classified
Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures.
Z-0364-2025 · Root cause: Storage · Recall details · FDA record ↗
Most reported problems
Device problems
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 10, 2021Break; Break
AS REPORTED, DURING TRANSURETHRAL LITHOTOMY PROCEDURE, THE ROOT OF THE SHEATH BROKE ABOUT 20 MINUTES AFTER THE START OF USE. THE COIL OF THE SHEATH DID NOT CAME OUT. THE DEVICE WAS REPLACED WITH SIMILAR DEVICE AND THE INTENDED PROCEDURE WAS COMPLETED WITH NO PATIENT HARM OR INJURY REPORTED DUE TO THE EVENT. NO USER INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12789924 ↗
FDA clearance and classification
Cleared through 510(k) K051593 on Sep 2, 2005 (decision: Substantially Equivalent). Submitted by Acmi Corporation as “Uropass Ureteral Access Sheath, Model 61224”. The FDA review took 78 days. FDA record ↗
The FDA regulates product code KNY (accessories, catheter, g-u) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K051593. One submission often covers many models, so check the product and lot details in each record.
- Z-2401-2023 · May 19, 2023 · Open, ClassifiedUroPass Ureteral Access Sheaths, 5 pieces/box
Common questions about the UroPass 61346BX
Has the UroPass 61346BX been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0364-2025 (Class II), began on Sep 11, 2024: Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the UroPass 61346BX?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Nov 10, 2021. That report lists device problem “Break” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the UroPass 61346BX?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the UroPass 61346BX FDA approved?
It was cleared through 510(k) K051593 on Sep 2, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the UroPass 61346BX, and is it still sold?
The FDA's device database lists Gyrus ACMI, LLC as the company responsible for it (brand UroPass, model 61346BX). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ACCESSORIES, CATHETER, G-U devices
- UroPass 61238BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall10 reports
- UroPass 61138BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall5 reports
- UroPass 61024BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall5 reports
- UroPass 61224BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61146BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61254BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61046BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61054BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page