Skip to content

PK TECHNOLOGY 952005PKRecalls & reported problems

Gyrus ACMI, LLC · PKS PLASMASPATULA 5mm/33cm (5/PK) · Class 2

All models of PK TECHNOLOGY (10)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00821925036253), received between Jun 6, 2022 and Apr 22, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Activation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Problem1
  • Break1
  • Sparking1

Patient problems

  • Burn(s)2
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2025Activation Problem; Activation Problem

    IT WAS REPORTED THAT THE SUBJECT DEVICE ENCOUNTERED ARCING WHEN ACTIVATED. THE ISSUE OCCURRED DURING A THERAPEUTIC HYSTERECTOMY PROCEDURE. THE PROCEDURE WAS COMPLETED WITH AN OLYMPUS BACKUP DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21882722 ↗
  • InjuryJun 30, 2022Break; Break

    UPON FOLLOW UP FOR PATIENT IDENTIFIER (B)(6) INVOLVING A PATIENT BURNED WHEN SUBJECT DEVICE SPARKED, THE CUSTOMER REPORTED A SECOND EVENT. DURING THE WEEK OF (B)(6) 2022, THE SUBJECT DEVICE FELL APART IN THE PATIENT AND ALL PIECES WERE RETRIEVED. THE CUSTOMER REFUSED TO PROVIDE ADDITIONAL INFORMATION AND STATED THE EVENT WAS SUBMITTED VIA MEDSUN REPORT.

    Patient problems: Burn(s); Foreign Body In Patient

    MAUDE report 14877708 ↗
  • InjuryJun 6, 2022Sparking; Sparking

    OLYMPUS RECEIVED THE FOLLOWING EVENT FROM AN OLYMPUS SALES SPECIALIST. DURING AN UNSPECIFIED PROCEDURE, THE DEVICE SPARKED AND CAUSED AN UNSPECIFIED PATIENT BURN WHEN THE USER HIT THE PLASTIC CODE DURING INSERTION WITH THE DEVICE. THE USER REPORTED THIS OCCURRED WHEN THE SPATULA WAS BEING PULLED OUT OF THE PATIENT AND THE BURN WENT THROUGH THE DRAPE TO THE PATIENT. THE CUSTOMER REFUSES TO PROVIDE ADDITIONAL INFORMATION AND STATED THIS WAS USER ERROR.

    Patient problems: Burn(s)

    MAUDE report 14604127 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page