DIEGO ELITE BB4040SSRecalls & reported problems
Gyrus ACMI, LLC · SHAVER BLADE, 4 MM, 40 DEGREES bipolar distal suction, SERRATED DIEGO ELITE · Class 2
All models of DIEGO ELITE (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00821925040564), received between Sep 18, 2022 and Sep 18, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (1 report).
12 other recalls cover devices cleared under the same FDA submission (K123429); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Split, Cut or Torn1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 18, 2022Mechanical Problem; Material Split, Cut or Torn; Mechanical Problem; Material Split, Cut or Torn
IT WAS REPORTED DURING REMOVAL OF POLYPS IN THE PARANASAL SINUS PROCEDURE , WHITE INSULATION ON SHAVER BLADE HAS COME LOOSE AND IS HALF TORN AND FOLDED OVER DURING SURGERY. REPORTER STATED THAT HE WAS AT THE SURGERY AND WAS THE FIRST TO NOTICE. ACCORDING TO THE REPORTER, IT COULD NOT BE SEEN THAT THE SURGEON HAD CAUSED IT WITH A SHARP INSTRUMENT. THE DEVICE WAS REPLACED AND THE INTENDED PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WAS NO PATIENT HARM, NO USER INJURY REPORTED DUE TO THE EVENT. THE SUBJECT DEVICE WAS DISPOSED OF BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15443471 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123429. One submission often covers many models, so check the product and lot details in each record.
- Z-1179-2019 · Mar 8, 2019 · TerminatedDiego Elite Turbinate Blades:Bipolar Blade, 2mm straight, standard, type A Part No BB2000SA
- Z-1181-2019 · Mar 8, 2019 · TerminatedDiego Elite Turbinate Blades Bipolar Blade, 4mm straight, Part No: BB4000SC
- Z-1183-2019 · Mar 8, 2019 · TerminatedDiego Elite Turbinate Blades Bipolar Blade, 4mm curved, Product Number: BB4040SC
- Z-1182-2019 · Mar 8, 2019 · TerminatedDiego Elite Turbinate Blades Bipolar Blade, 4mm curved, Product Number:BB4040SS
- Z-1187-2019 · Mar 8, 2019 · TerminatedDiego Elite Malleable, 4mm Standard Monopolar Blades Product Code:MM4000SS
- Z-1180-2019 · Mar 8, 2019 · TerminatedDiego Elite Turbinate Blades Bipolar Blade, 4mm straight, Part Number: BB4000SS
- Z-1178-2019 · Mar 8, 2019 · TerminatedOlympus Diego Elite Console MDCONS 100
- Z-1186-2019 · Mar 8, 2019 · TerminatedDiego Elite Blades Monopolar, 4mm Straight, TIA Product Code:MB4000SS
- Z-1184-2019 · Mar 8, 2019 · TerminatedDiego Elite Turbinate Blades Bipolar Blade, 4mm curved, Product Number: BB4040XS
- Z-1185-2019 · Mar 8, 2019 · TerminatedDiego Elite Turbinate Blades Bipolar Blade, 4mm curved, Product Number: BB4040XC
- Z-1753-2015 · May 8, 2015 · TerminatedGyrus ACMI Diego Elite Tubeset, intended for cutting, coagulation, drilling, debriding, and removal of bone, and soft and hard tissue in general ENT, Sinus/Rhinology, Nasopharyngea
- Z-1752-2015 · May 8, 2015 · TerminatedGyrus ACMI Diego Elite Tubeset; intended for cutting, coagulation, drilling, debriding, and removal of bone, and soft and hard tissue in general ENT, Sinus/Rhinology, Nasopharyngea
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page